Straumann BLX Line Extension - Implants, SRAs and Anatomic Abutments SLActive Roxolid BL (REF 061.4316)Recalls & reported problems
Institut Straumann AG · BLX Ø3.75 RB, SLActive® 16mm, RXD · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Straumann BLX dental implant is placed in bone. Most of the 302 reports describe the implant failing to bond with the bone, and some mention scar-like tissue around the implant, pain, infection, or bleeding. The reports include 302 injuries and no deaths, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 301
- Deaths reported
- 0
- Injuries reported
- 301
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 301 adverse event reports that list this exact device (DI 07630031736536), received between Feb 11, 2019 and Aug 27, 2026 — classified as 301 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (235 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Osseointegrate235
- Loss of Osseointegration41
- Adverse Event Without Identified Device or Use Problem27
- Difficult to Insert3
Patient problems
- No Clinical Signs, Symptoms or Conditions142
- Fibrosis96
- Pain61
- Unspecified Infection49
- Tissue Damage21
- Hemorrhage/Blood Loss/Bleeding17
- Abscess13
- Inflammation13
Recent adverse event reports
The 25 most recent of 301 reports, in the reporter's own words as published by the FDA.
InjuryAug 27, 2026Loss of Osseointegration
The clinician reports the implant was inserted (b)(6) 2023 in FDI 35. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.
Patient problems: Fibrosis
MAUDE report 50480533 ↗InjuryAug 24, 2026Failure to Osseointegrate
Patient problems: Fibrosis; Pain
MAUDE report 50415624 ↗InjuryAug 17, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in FDI 45. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50284118 ↗InjuryAug 17, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2025 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50284049 ↗InjuryAug 17, 2026Loss of Osseointegration
The clinician reports the implant was inserted on (b)(6) 2024 in the patient's mouth. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50284043 ↗InjuryAug 17, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (B)(6) 2025 in the patient's mouth. On (B)(6) 2026 non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50284040 ↗InjuryAug 17, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2025 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50284032 ↗InjuryAug 14, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (B)(6) 2026 in the patient's mouth. On (B)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Infection. No further patient complications were reported.
Patient problems: Unspecified Infection
MAUDE report 50258642 ↗InjuryAug 13, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (B)(6) 2026 in ADA 22. On (B)(6) 2026 non-osseointegration was verified. Patient presented with poor oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Infection, Pain, Mobility and Bleeding. No further patient complications were reported.
Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Pain; Unspecified Infection
MAUDE report 50238731 ↗InjuryAug 13, 2026Failure to Osseointegrate
Patient problems: Pain; Unspecified Infection
MAUDE report 50238721 ↗InjuryAug 13, 2026Loss of Osseointegration
Patient problems: Abscess; Fibrosis; Fistula; Hemorrhage/Blood Loss/Bleeding; Pain; Sinus Perforation; Unspecified Infection
MAUDE report 50238512 ↗InjuryAug 13, 2026Loss of Osseointegration
Patient problems: Fibrosis; Inflammation; Pain; Swelling/ Edema; Unspecified Infection
MAUDE report 50238509 ↗InjuryAug 13, 2026Failure to Osseointegrate
The clinician reports the implant was inserted 2026 (b)(6) in ADA 22. On 2026 (b)(6), non-osseointegration was verified. Patient presented with poor oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Infection, Pain, Mobility and Bleeding. No further patient complications were reported.
Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Pain; Unspecified Infection
MAUDE report 50238129 ↗InjuryAug 13, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in ADA 7. Details of surgery: Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III and poor oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis and Pain. No further patient complications were reported.
Patient problems: Pain; Unspecified Infection
MAUDE report 50238121 ↗InjuryAug 13, 2026Loss of Osseointegration
The clinician reports the implant was inserted (B)(6) 2025 in ADA 4. Details of surgery: Primary stability not achieved, Implant surface not completely covered with bone and Sinus perforation. On (B)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Infection, Peri-implantitis, Pain, Mobility, Bleeding, Abscess and Fistula. No further patient complications were reported.
Patient problems: Abscess; Fibrosis; Fistula; Hemorrhage/Blood Loss/Bleeding; Pain; Sinus Perforation; Unspecified Infection
MAUDE report 50238035 ↗InjuryAug 13, 2026Loss of Osseointegration
The clinician reports the implant was inserted (B)(6) 2025 in ADA 1. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On (B)(6) 2026, loss of osseointegration was verified. Patient presented with bone type IV, poor oral hygiene, bruxism and inadequate bone quality/quantity. The device was forwarded to the manufacturer. At the event the patient experienced: Infection, Peri-implantitis, Pain, Mobility, Swelling and Inflammation. No further patient complications were reported.
Patient problems: Fibrosis; Inflammation; Pain; Swelling/ Edema; Unspecified Infection
MAUDE report 50238027 ↗InjuryAug 10, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (B)(6) 2026 in FDI 26. On (B)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.
Patient problems: Fibrosis
MAUDE report 50197795 ↗InjuryAug 3, 2026Failure to Osseointegrate
Patient problems: Fibrosis
MAUDE report 50119644 ↗InjuryJul 27, 2026Failure to Osseointegrate
The clinician reports the implant was inserted on (b)(6)2026 in ADA 7. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50041413 ↗InjuryJul 27, 2026Loss of Osseointegration
The clinician reports the implant was inserted (b)(6) 2019 in the patient's mouth. Details of surgery: Ridge augmentation. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Pain, Mobility, Bleeding, Abscess and Inflammation. No further patient complications were reported.
Patient problems: Abscess; Fibrosis; Hemorrhage/Blood Loss/Bleeding; Inflammation; Pain
MAUDE report 50040434 ↗InjuryJul 27, 2026Failure to Osseointegrate
The clinician reports the implant was inserted 2026 (b)(6) in ADA 7. On 2026 (b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50039588 ↗InjuryJul 17, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE AND SINUS PERFORATION. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: NUMBNESS. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Numbness; Sinus Perforation
MAUDE report 25859704 ↗InjuryJul 17, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN ADA 11. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding
MAUDE report 25859293 ↗InjuryJul 17, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN ADA 11. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 25859239 ↗InjuryJul 17, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN ADA 11. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding
MAUDE report 25858590 ↗
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.