Skip to content

Straumann BLX Line Extension - Implants, SRAs and Anatomic Abutments SLActive Roxolid BL (REF 061.4316)Recalls & reported problems

Institut Straumann AG · BLX Ø3.75 RB, SLActive® 16mm, RXD · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Straumann BLX dental implant is placed in bone. Most of the 302 reports describe the implant failing to bond with the bone, and some mention scar-like tissue around the implant, pain, infection, or bleeding. The reports include 302 injuries and no deaths, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
301
Deaths reported
0
Injuries reported
301
Recalls naming it
0

Summary

The FDA's MAUDE database contains 301 adverse event reports that list this exact device (DI 07630031736536), received between Feb 11, 2019 and Aug 27, 2026 — classified as 301 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (235 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 6 reportsDec 2023: 3 reportsJan 2024: 8 reportsFeb 2024: 3 reportsMar 2024: 4 reportsApr 2024: 6 reportsMay 2024: 1 reportsJun 2024: 5 reportsJul 2024: 4 reportsAug 2024: 2 reportsSep 2024: 3 reportsOct 2024: 10 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 5 reportsMar 2025: 2 reportsApr 2025: 3 reportsMay 2025: 0 reportsJun 2025: 14 reportsJul 2025: 9 reportsAug 2025: 14 reportsSep 2025: 2 reportsOct 2025: 2 reportsNov 2025: 7 reportsDec 2025: 6 reportsJan 2026: 4 reportsFeb 2026: 13 reportsMar 2026: 5 reportsApr 2026: 1 reportsMay 2026: 11 reportsJun 2026: 9 reportsJul 2026: 16 reportsAug 2026: 18 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202618
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate235
  • Loss of Osseointegration41
  • Adverse Event Without Identified Device or Use Problem27
  • Difficult to Insert3

Patient problems

  • No Clinical Signs, Symptoms or Conditions142
  • Fibrosis96
  • Pain61
  • Unspecified Infection49
  • Tissue Damage21
  • Hemorrhage/Blood Loss/Bleeding17
  • Abscess13
  • Inflammation13

Recent adverse event reports

The 25 most recent of 301 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 27, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2023 in FDI 35. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50480533 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: Fibrosis; Pain

    MAUDE report 50415624 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 45. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50284118 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50284049 ↗
  • InjuryAug 17, 2026Loss of Osseointegration

    The clinician reports the implant was inserted on (b)(6) 2024 in the patient's mouth. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50284043 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2025 in the patient's mouth. On (B)(6) 2026 non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50284040 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50284032 ↗
  • InjuryAug 14, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2026 in the patient's mouth. On (B)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Infection. No further patient complications were reported.

    Patient problems: Unspecified Infection

    MAUDE report 50258642 ↗
  • InjuryAug 13, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2026 in ADA 22. On (B)(6) 2026 non-osseointegration was verified. Patient presented with poor oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Infection, Pain, Mobility and Bleeding. No further patient complications were reported.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Pain; Unspecified Infection

    MAUDE report 50238731 ↗
  • InjuryAug 13, 2026Failure to Osseointegrate

    Patient problems: Pain; Unspecified Infection

    MAUDE report 50238721 ↗
  • InjuryAug 13, 2026Loss of Osseointegration

    Patient problems: Abscess; Fibrosis; Fistula; Hemorrhage/Blood Loss/Bleeding; Pain; Sinus Perforation; Unspecified Infection

    MAUDE report 50238512 ↗
  • InjuryAug 13, 2026Loss of Osseointegration

    Patient problems: Fibrosis; Inflammation; Pain; Swelling/ Edema; Unspecified Infection

    MAUDE report 50238509 ↗
  • InjuryAug 13, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026 (b)(6) in ADA 22. On 2026 (b)(6), non-osseointegration was verified. Patient presented with poor oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Infection, Pain, Mobility and Bleeding. No further patient complications were reported.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Pain; Unspecified Infection

    MAUDE report 50238129 ↗
  • InjuryAug 13, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 7. Details of surgery: Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III and poor oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis and Pain. No further patient complications were reported.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 50238121 ↗
  • InjuryAug 13, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (B)(6) 2025 in ADA 4. Details of surgery: Primary stability not achieved, Implant surface not completely covered with bone and Sinus perforation. On (B)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Infection, Peri-implantitis, Pain, Mobility, Bleeding, Abscess and Fistula. No further patient complications were reported.

    Patient problems: Abscess; Fibrosis; Fistula; Hemorrhage/Blood Loss/Bleeding; Pain; Sinus Perforation; Unspecified Infection

    MAUDE report 50238035 ↗
  • InjuryAug 13, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (B)(6) 2025 in ADA 1. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On (B)(6) 2026, loss of osseointegration was verified. Patient presented with bone type IV, poor oral hygiene, bruxism and inadequate bone quality/quantity. The device was forwarded to the manufacturer. At the event the patient experienced: Infection, Peri-implantitis, Pain, Mobility, Swelling and Inflammation. No further patient complications were reported.

    Patient problems: Fibrosis; Inflammation; Pain; Swelling/ Edema; Unspecified Infection

    MAUDE report 50238027 ↗
  • InjuryAug 10, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2026 in FDI 26. On (B)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50197795 ↗
  • InjuryAug 3, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50119644 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted on (b)(6)2026 in ADA 7. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50041413 ↗
  • InjuryJul 27, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2019 in the patient's mouth. Details of surgery: Ridge augmentation. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Pain, Mobility, Bleeding, Abscess and Inflammation. No further patient complications were reported.

    Patient problems: Abscess; Fibrosis; Hemorrhage/Blood Loss/Bleeding; Inflammation; Pain

    MAUDE report 50040434 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026 (b)(6) in ADA 7. On 2026 (b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50039588 ↗
  • InjuryJul 17, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE AND SINUS PERFORATION. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: NUMBNESS. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Numbness; Sinus Perforation

    MAUDE report 25859704 ↗
  • InjuryJul 17, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN ADA 11. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding

    MAUDE report 25859293 ↗
  • InjuryJul 17, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN ADA 11. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 25859239 ↗
  • InjuryJul 17, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN ADA 11. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding

    MAUDE report 25858590 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page