Skip to content

Straumann BLX Implant System SLActive Roxolid BL (REF 061.6314)Recalls & reported problems

Institut Straumann AG · Bone Level X Implant, Ø 4.5mm RB, SLActive® 14mm, Roxolid® · Class 2 · Implant

All models of Straumann BLX Implant System (60)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Straumann BLX implant is a dental implant placed into the jawbone. Most of its 975 reports describe the implant not fusing with the bone, and some describe pain, infection and swelling. The reports mention no deaths and 975 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
985
Deaths reported
0
Injuries reported
985
Recalls naming it
0

Summary

The FDA's MAUDE database contains 985 adverse event reports that list this exact device (DI 07630031736598), received between May 13, 2019 and Sep 28, 2026 — classified as 985 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (578 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 11 reportsDec 2023: 8 reportsJan 2024: 8 reportsFeb 2024: 12 reportsMar 2024: 9 reportsApr 2024: 8 reportsMay 2024: 14 reportsJun 2024: 7 reportsJul 2024: 21 reportsAug 2024: 17 reportsSep 2024: 17 reportsOct 2024: 22 reportsNov 2024: 14 reportsDec 2024: 11 reportsJan 2025: 13 reportsFeb 2025: 12 reportsMar 2025: 16 reportsApr 2025: 3 reportsMay 2025: 0 reportsJun 2025: 42 reportsJul 2025: 15 reportsAug 2025: 24 reportsSep 2025: 12 reportsOct 2025: 20 reportsNov 2025: 12 reportsDec 2025: 26 reportsJan 2026: 9 reportsFeb 2026: 16 reportsMar 2026: 25 reportsApr 2026: 17 reportsMay 2026: 9 reportsJun 2026: 11 reportsJul 2026: 18 reportsAug 2026: 17 reportsSep 2026: 15 reportsOct 2026: 0 reports20242025202642
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate578
  • Adverse Event Without Identified Device or Use Problem229
  • Loss of Osseointegration147
  • Difficult to Insert20
  • Fracture6
  • Separation Failure6
  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions478
  • Fibrosis300
  • Pain216
  • Unspecified Infection132
  • Inflammation70
  • Swelling/ Edema53
  • Tissue Damage50
  • Abscess38

Recent adverse event reports

The 25 most recent of 985 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50921999 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Fibrosis

    MAUDE report 50921482 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50919943 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50919612 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918482 ↗
  • InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem

    The clinician reports that the day the implant was placed in ADA 10, failure occurred upon insertion. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50807834 ↗
  • InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem

    The clinician reports that the day the implant was placed in ADA 10, failure occurred upon insertion. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50806535 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50759594 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50759303 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 30. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50757658 ↗
  • InjurySep 10, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50720201 ↗
  • InjurySep 10, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50719046 ↗
  • InjurySep 7, 2026Failure to Osseointegrate

    Patient problems: Fibrosis; Fistula

    MAUDE report 50655402 ↗
  • InjurySep 7, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50655392 ↗
  • InjurySep 7, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50655382 ↗
  • InjuryAug 31, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2022 in ADA 29. Details of surgery: Implant surface not completely covered with bone. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Infection, Peri-implantitis, Mobility, Swelling, Bleeding and Fistula. No further patient complications were reported.

    Patient problems: Fibrosis; Fistula; Hemorrhage/Blood Loss/Bleeding; Swelling/ Edema; Unspecified Infection

    MAUDE report 50529295 ↗
  • InjuryAug 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2026 in FDI 13. On (B)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50481163 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2025(b)(6) in FDI 36. On 2026(b)(6), non-osseointegration was verified. Patient presented with bone type III. No product was returned to the manufacturer. At the event the patient experienced: Peri-implantitis and Pain. No further patient complications were reported.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 50415743 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50414516 ↗
  • InjuryAug 14, 2026Adverse Event Without Identified Device or Use Problem

    The clinician reports that the day the implant was placed in ADA 4, failure occurred upon insertion. Details of surgery: Primary stability not achieved, Ridge augmentation and Immediate extraction site. Patient presented with bone type III and inadequate bone quality/quantity. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50258814 ↗
  • InjuryAug 14, 2026Adverse Event Without Identified Device or Use Problem

    The clinician reports that the day the implant was placed in ADA 4, failure occurred upon insertion. Details of surgery: Primary stability not achieved, Ridge augmentation and Immediate extraction site. Patient presented with bone type III and inadequate bone quality/quantity. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50258189 ↗
  • InjuryAug 13, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50238717 ↗
  • InjuryAug 13, 2026Loss of Osseointegration

    The clinician reports the implant was inserted 2025 (b)(6) in ADA 12. On 2025 (b)(6), loss of osseointegration was verified. Patient presented with bone type IV, poor oral hygiene and inadequate bone quality/quantity. The device was forwarded to the manufacturer. At the event the patient experienced: Pain. No further patient complications were reported.

    Patient problems: Pain

    MAUDE report 50238708 ↗
  • InjuryAug 13, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2024 (b)(6) in ADA 7. On 2026 (b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50238116 ↗
  • InjuryAug 13, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (B)(6) 2025 in ADA 12. On (B)(6) 2025, loss of osseointegration was verified. Patient presented with bone type IV, poor oral hygiene and inadequate bone quality/quantity. The device was forwarded to the manufacturer. At the event the patient experienced: Pain. No further patient complications were reported.

    Patient problems: Pain

    MAUDE report 50238098 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page