Skip to content

Straumann BLX Implant System SLActive Roxolid BL (REF 061.9306)Recalls & reported problems

Institut Straumann AG · Bone Level X Implant, Ø 6.5mm WB, SLActive® 6mm, Roxolid® · Class 2 · Implant

All models of Straumann BLX Implant System (60)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Straumann BLX SLActive dental implant has 772 FDA reports, all of them injury reports, and no deaths. Most reports describe the implant failing to bond with the jawbone or losing that bond, and many mention scar tissue, pain or infection. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
776
Deaths reported
0
Injuries reported
776
Recalls naming it
0

Summary

The FDA's MAUDE database contains 776 adverse event reports that list this exact device (DI 07630031736666), received between May 24, 2019 and Sep 28, 2026 — classified as 776 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (402 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 14 reportsDec 2023: 4 reportsJan 2024: 4 reportsFeb 2024: 11 reportsMar 2024: 13 reportsApr 2024: 14 reportsMay 2024: 19 reportsJun 2024: 6 reportsJul 2024: 8 reportsAug 2024: 10 reportsSep 2024: 14 reportsOct 2024: 12 reportsNov 2024: 18 reportsDec 2024: 23 reportsJan 2025: 21 reportsFeb 2025: 10 reportsMar 2025: 15 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 27 reportsJul 2025: 14 reportsAug 2025: 16 reportsSep 2025: 8 reportsOct 2025: 11 reportsNov 2025: 4 reportsDec 2025: 12 reportsJan 2026: 9 reportsFeb 2026: 10 reportsMar 2026: 15 reportsApr 2026: 21 reportsMay 2026: 21 reportsJun 2026: 9 reportsJul 2026: 7 reportsAug 2026: 11 reportsSep 2026: 7 reportsOct 2026: 0 reports20242025202627
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate402
  • Loss of Osseointegration219
  • Adverse Event Without Identified Device or Use Problem123
  • Difficult to Insert25
  • Separation Failure7
  • Fracture5

Patient problems

  • No Clinical Signs, Symptoms or Conditions355
  • Fibrosis303
  • Pain117
  • Unspecified Infection95
  • Hemorrhage/Blood Loss/Bleeding34
  • Inflammation34
  • Tissue Damage32
  • Swelling/ Edema18

Recent adverse event reports

The 25 most recent of 776 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50921433 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50920490 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50920366 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918690 ↗
  • InjurySep 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50807416 ↗
  • InjurySep 17, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6)2020 in FDI 27. On (b)(6)2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50807049 ↗
  • InjurySep 17, 2026Fracture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50806696 ↗
  • InjuryAug 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 47. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type IV, fair oral hygiene and inadequate bone quality/quantity. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50480547 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted on (b)(6)2025 in the patient's mouth. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50284009 ↗
  • InjuryAug 17, 2026Adverse Event Without Identified Device or Use Problem

    The clinician reports that the day the implant was placed in FDI 27, failure occurred upon insertion. Details of surgery: Primary stability not achieved. Patient presented with bone type III. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50283874 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 14. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50282559 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 14. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50281042 ↗
  • InjuryAug 14, 2026Loss of Osseointegration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50258729 ↗
  • InjuryAug 10, 2026Loss of Osseointegration

    The clinician reports the implant was inserted 2022 (b)(6) in ADA 13. Details of surgery: Immediate extraction site. On 2026 (b)(6), loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50199183 ↗
  • InjuryAug 10, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in FDI 46. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50198828 ↗
  • InjuryAug 10, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2026 in FDI 36. On (B)(6) 2026 non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50197780 ↗
  • InjuryAug 10, 2026Loss of Osseointegration

    The clinician reports the implant was inserted 2022 (b)(6) in ADA 13. Details of surgery: Immediate extraction site. On 2026 (b)(6), loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50197340 ↗
  • InjuryAug 3, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50120970 ↗
  • InjuryJul 31, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in FDI 36. On(b)(6) 2026, on-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50102544 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2025-(b)(6) in FDI 26. On 2026-(b)(6), non-osseointegration was verified. Patient presented with bone type III. No product was returned to the manufacturer. At the event the patient experienced: Peri-implantitis and Pain. No further patient complications were reported.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 50041490 ↗
  • InjuryJul 13, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN FDI 16. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25803666 ↗
  • InjuryJul 10, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN FDI 15. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25786261 ↗
  • InjuryJul 9, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN FDI 34. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25768691 ↗
  • InjuryJul 3, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2026 IN ADA 18. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND SWELLING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 25705918 ↗
  • InjuryJul 3, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN ADA 18. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND SWELLING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 25705368 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page