Skip to content

Straumann TLX Implant System SLActive Roxolid TL (REF 035.3116S)Recalls & reported problems

Institut Straumann AG · TLX SP Implant, Ø 3.75mm RT, SLActive® 16mm, Roxolid® · Class 2 · Implant

All models of Straumann TLX Implant System (132)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
11
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 07630031742810), received between Dec 15, 2022 and Jul 27, 2026 — classified as 11 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 5 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 3 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate7
  • Loss of Osseointegration4

Patient problems

  • Fibrosis8
  • Pain8
  • Hemorrhage/Blood Loss/Bleeding4
  • Inflammation4
  • Swelling/ Edema4
  • Unspecified Infection4
  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 42. Details of surgery: Primary stability not achieved. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50010103 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 32. Details of surgery: Primary stability not achieved. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50010102 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 16. Details of surgery: Primary stability not achieved. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50010101 ↗
  • InjuryMar 10, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21565797 ↗
  • InjuryMar 7, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN ADA 14. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION, PAIN, MOBILITY, SWELLING, BLEEDING AND INFLAMMATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Inflammation; Pain; Swelling/ Edema; Unspecified Infection

    MAUDE report 21550729 ↗
  • InjuryMar 7, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN ADA 12. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION, PAIN, MOBILITY, SWELLING, BLEEDING AND INFLAMMATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Inflammation; Pain; Swelling/ Edema; Unspecified Infection

    MAUDE report 21550689 ↗
  • InjuryMar 7, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN ADA 14. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION, PAIN, MOBILITY, SWELLING, BLEEDING AND INFLAMMATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Inflammation; Pain; Swelling/ Edema; Unspecified Infection

    MAUDE report 21550016 ↗
  • InjuryMar 7, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN ADA 12. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION, PAIN, MOBILITY, SWELLING, BLEEDING AND INFLAMMATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Inflammation; Pain; Swelling/ Edema; Unspecified Infection

    MAUDE report 21550000 ↗
  • InjuryDec 15, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022 IN FDI 25. ON (B)(6) 2022, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15989476 ↗
  • InjuryDec 15, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022 IN FDI 25. ON (B)(6) 2022, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain

    MAUDE report 15989438 ↗
  • InjuryDec 15, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022 IN FDI 15. ON (B)(6) 2022, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15989432 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page