Straumann BLX Implant System SLA Roxolid BL (REF 061.9510)Recalls & reported problems

Institut Straumann AG · BLX Implant, Ø 6.5mm WB, SLA® 10mm, Roxolid® · Class 2 · Implant

All models of Straumann BLX Implant System (60)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Straumann BLX implant is a screw-shaped device placed into bone during dental surgery. Many of the 316 reports describe the implant failing to bond with the bone, and some mention pain or infection in the patient. All 316 reports are injury reports, with no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
323
Deaths reported
0
Injuries reported
323
Recalls naming it
0

Summary

The FDA's MAUDE database contains 323 adverse event reports that list this exact device (DI 07630031743916), received between Feb 15, 2021 and Sep 28, 2026 — classified as 323 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (175 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 7 reportsJan 2024: 8 reportsFeb 2024: 4 reportsMar 2024: 6 reportsApr 2024: 5 reportsMay 2024: 6 reportsJun 2024: 9 reportsJul 2024: 2 reportsAug 2024: 6 reportsSep 2024: 10 reportsOct 2024: 3 reportsNov 2024: 7 reportsDec 2024: 10 reportsJan 2025: 4 reportsFeb 2025: 7 reportsMar 2025: 6 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 17 reportsJul 2025: 7 reportsAug 2025: 8 reportsSep 2025: 2 reportsOct 2025: 9 reportsNov 2025: 12 reportsDec 2025: 14 reportsJan 2026: 6 reportsFeb 2026: 3 reportsMar 2026: 17 reportsApr 2026: 11 reportsMay 2026: 9 reportsJun 2026: 2 reportsJul 2026: 7 reportsAug 2026: 6 reportsSep 2026: 9 reportsOct 2026: 0 reports20242025202617
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate175
  • Adverse Event Without Identified Device or Use Problem90
  • Loss of Osseointegration49
  • Difficult to Insert9

Patient problems

  • No Clinical Signs, Symptoms or Conditions180
  • Fibrosis97
  • Pain42
  • Unspecified Infection33
  • Hemorrhage/Blood Loss/Bleeding22
  • Inflammation16
  • Swelling/ Edema11
  • Hyperresponsive to stimuli10

Recent adverse event reports

The 25 most recent of 323 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50921552 ↗
  • InjurySep 18, 2026Failure to Osseointegrate

    Patient problems: Swelling/ Edema

    MAUDE report 50825210 ↗
  • InjurySep 18, 2026Failure to Osseointegrate

    Patient problems: Swelling/ Edema

    MAUDE report 50824328 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50759461 ↗
  • InjurySep 11, 2026Loss of Osseointegration

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding

    MAUDE report 50738510 ↗
  • InjurySep 11, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026 (b)(6) in ADA 2. On 2026 (b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Pain, Mobility and Bleeding. No further patient complications were reported.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Pain

    MAUDE report 50737707 ↗
  • InjurySep 11, 2026Loss of Osseointegration

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding

    MAUDE report 50737589 ↗
  • InjurySep 7, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Unspecified Infection

    MAUDE report 50655338 ↗
  • InjurySep 7, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Unspecified Infection

    MAUDE report 50653209 ↗
  • InjuryAug 24, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50416494 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50416438 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50416423 ↗
  • InjuryAug 10, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 16. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50199081 ↗
  • InjuryAug 6, 2026Adverse Event Without Identified Device or Use Problem

    The clinician reports that the day the implant was placed in ADA 14, failure occurred upon insertion. Patient presented with inadequate bone quality/quantity. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50163591 ↗
  • InjuryAug 6, 2026Adverse Event Without Identified Device or Use Problem

    The clinician reports that the day the implant was placed in ADA 14, failure occurred upon insertion. Patient presented with inadequate bone quality/quantity. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50162267 ↗
  • InjuryJul 31, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6)2026 in ADA 13. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Pain, Mobility, Bleeding and Fistula. No further patient complications were reported.

    Patient problems: Fibrosis; Fistula; Hemorrhage/Blood Loss/Bleeding; Pain

    MAUDE report 50102918 ↗
  • InjuryJul 31, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 13. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Pain, Mobility, Bleeding and Fistula. No further patient complications were reported.

    Patient problems: Fibrosis; Fistula; Hemorrhage/Blood Loss/Bleeding; Pain

    MAUDE report 50102220 ↗
  • InjuryJul 27, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2025 in ADA 15. Details of surgery: Immediate extraction site. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility and Increased Sensitivity. No further patient complications were reported.

    Patient problems: Fibrosis; Hyperresponsive to stimuli

    MAUDE report 50023337 ↗
  • InjuryJul 27, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (B)(6) 2025 in ADA 15. Details of surgery: Immediate extraction site. On (B)(6) 2026 loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility and Increased Sensitivity. No further patient complications were reported.

    Patient problems: Fibrosis; Hyperresponsive to stimuli

    MAUDE report 50022782 ↗
  • InjuryJul 13, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN FDI 47. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain

    MAUDE report 25800564 ↗
  • InjuryJul 10, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN ADA 18. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 25783366 ↗
  • InjuryJul 2, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25694410 ↗
  • InjuryJun 22, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 25583724 ↗
  • InjuryJun 3, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN ADA 31. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE AND RIDGE AUGMENTATION. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV, INADEQUATE BONE QUALITY/QUANTITY AND PREVIOUS BONE AUGMENTATION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION, PERI-IMPLANTITIS, PAIN AND INFLAMMATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Inflammation; Pain; Unspecified Infection

    MAUDE report 25379267 ↗
  • InjuryMay 18, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN ADA 3. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMMEDIATE EXTRACTION SITE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25206507 ↗