Straumann® BLC and TLC Implants SLA Roxolid BL (REF 035.8412S)Recalls & reported problems
Institut Straumann AG · BLC Implant, Ø 4.5mm WB, SLA® 12mm, Roxolid® · Class 2 · Implant
All models of Straumann BLC and TLC Implants (153)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
These are Straumann dental implants used as root-form anchors in the jawbone. The 64 reports filed between January and August 2026 mostly describe implants that fail to fuse with the bone, or that lose that fusion later, along with pain, inflammation, swelling and scarring at the site; a few mention implants breaking or being hard to place. All 64 reports are injuries, with no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 62
- Deaths reported
- 0
- Injuries reported
- 62
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 62 adverse event reports that list this exact device (DI 07630031764379), received between Jan 27, 2026 and Sep 11, 2026 — classified as 62 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (47 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Osseointegrate47
- Adverse Event Without Identified Device or Use Problem8
- Loss of Osseointegration3
- Difficult to Insert2
- Separation Failure2
Patient problems
- Fibrosis30
- Pain21
- No Clinical Signs, Symptoms or Conditions18
- Unspecified Infection14
- Swelling/ Edema8
- Hyperresponsive to stimuli6
- Inflammation4
- Abscess2
Recent adverse event reports
The 25 most recent of 62 reports, in the reporter's own words as published by the FDA.
InjurySep 11, 2026Failure to Osseointegrate
Patient problems: Pain; Unspecified Infection
MAUDE report 50737860 ↗InjurySep 7, 2026Failure to Osseointegrate
Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Pain; Swelling/ Edema; Unspecified Infection
MAUDE report 50654033 ↗InjuryAug 24, 2026Failure to Osseointegrate
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50415488 ↗InjuryAug 24, 2026Failure to Osseointegrate
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50415474 ↗InjuryAug 24, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6)2026 in FDI 16. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.
Patient problems: Fibrosis; Pain
MAUDE report 50415314 ↗InjuryAug 24, 2026Failure to Osseointegrate
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50413404 ↗InjuryAug 24, 2026Failure to Osseointegrate
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50413378 ↗InjuryAug 17, 2026Failure to Osseointegrate
The clinician reports the implant was inserted ( b)(6) 2026 in the patient's mouth. Details of surgery: Implant surface not completely covered with bone. On (b)(6) 2026 , non-osseointegration was verified. Patient presented with inadequate bone quality/quantity. No product was returned to the manufacturer. At the event the patient experienced: Infection and Pain. No further patient complications were reported.
Patient problems: Pain; Unspecified Infection
MAUDE report 50283493 ↗InjuryAug 17, 2026Difficult to Insert
The clinician reports that the day the implant was placed in ADA 13, failure occurred upon insertion. Details of surgery: Primary stability not achieved. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50282920 ↗InjuryAug 17, 2026Separation Failure
The clinician reports that the day the implant was placed in ADA 29, failure occurred upon insertion. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50282912 ↗InjuryAug 17, 2026Failure to Osseointegrate
The clinician reports the implant was inserted(b)(6) 2026 in ADA 6. Details of surgery: Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Infection, Pain and Mobility. No further patient complications were reported.
Patient problems: Fibrosis; Pain; Unspecified Infection
MAUDE report 50282869 ↗InjuryAug 17, 2026Separation Failure
The clinician reports that the day the implant was placed in ADA 29, failure occurred upon insertion. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50281190 ↗InjuryAug 17, 2026Difficult to Insert
The clinician reports that the day the implant was placed in ADA 13, failure occurred upon insertion. Details of surgery: Primary stability not achieved. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50281186 ↗InjuryAug 17, 2026Failure to Osseointegrate
The clinician reports the implant was inserted 2026 (b)(6) in ADA 6. Details of surgery: Implant surface not completely covered with bone. On 2026 (b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Infection, Pain and Mobility. No further patient complications were reported.
Patient problems: Fibrosis; Pain; Unspecified Infection
MAUDE report 50281152 ↗InjuryAug 14, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Fibrosis
MAUDE report 50258789 ↗InjuryAug 14, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Fibrosis
MAUDE report 50258117 ↗InjuryAug 10, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in FDI 14. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50199575 ↗InjuryAug 10, 2026Failure to Osseointegrate
The clinician reports the implant was inserted 2026 (b)(6) in FDI 46. On 2026 (b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.
Patient problems: Fibrosis
MAUDE report 50198654 ↗InjuryAug 3, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in FDI 15. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.
Patient problems: Fibrosis; Pain
MAUDE report 50119304 ↗InjuryJul 31, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Peri-implantitis. No further patient complications were reported.
Patient problems: Unspecified Infection
MAUDE report 50102589 ↗InjuryJul 27, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.
Patient problems: Fibrosis
MAUDE report 50043195 ↗InjuryJul 27, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in FDI 15. Details of surgery: Primary stability not achieved. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.
Patient problems: Fibrosis
MAUDE report 50040651 ↗InjuryJul 3, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN ADA 6. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INCREASED SENSITIVITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Hyperresponsive to stimuli
MAUDE report 25706024 ↗InjuryJul 3, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN ADA 6. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INCREASED SENSITIVITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Hyperresponsive to stimuli
MAUDE report 25705398 ↗InjuryJun 19, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN FDI 23. DETAILS OF SURGERY: RIDGE AUGMENTATION. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25564864 ↗
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗