Straumann® BLC and TLC Implants SLActive Roxolid BL (REF 035.9706S)Recalls & reported problems

Institut Straumann AG · BLC Implant, Ø 5.5mm WB, SLActive® 6mm, Roxolid® · Class 2 · Implant

All models of Straumann BLC and TLC Implants (153)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Straumann BLC implant has 194 reports, all of them injuries, and no deaths. Most describe the implant failing to bond with the bone, and many mention fibrosis, meaning scar-like tissue around the implant, and pain. Reports rose from 65 in the prior 12 months to 118 in the most recent 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
201
Deaths reported
0
Injuries reported
201
Recalls naming it
0

Summary

The FDA's MAUDE database contains 201 adverse event reports that list this exact device (DI 07630031764546), received between Jul 4, 2024 and Sep 28, 2026 — classified as 201 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (129 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 2 reportsSep 2024: 4 reportsOct 2024: 3 reportsNov 2024: 3 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 19 reportsJul 2025: 8 reportsAug 2025: 7 reportsSep 2025: 9 reportsOct 2025: 14 reportsNov 2025: 8 reportsDec 2025: 16 reportsJan 2026: 7 reportsFeb 2026: 16 reportsMar 2026: 20 reportsApr 2026: 12 reportsMay 2026: 11 reportsJun 2026: 10 reportsJul 2026: 10 reportsAug 2026: 7 reportsSep 2026: 8 reportsOct 2026: 0 reports20242025202620
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate129
  • Adverse Event Without Identified Device or Use Problem42
  • Loss of Osseointegration28
  • Difficult to Insert2

Patient problems

  • No Clinical Signs, Symptoms or Conditions97
  • Fibrosis85
  • Pain28
  • Unspecified Infection24
  • Inflammation12
  • Hemorrhage/Blood Loss/Bleeding9
  • Abscess4
  • Numbness2

Recent adverse event reports

The 25 most recent of 201 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50919956 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in ADA 14. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50918500 ↗
  • InjurySep 18, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50825900 ↗
  • InjurySep 18, 2026Loss of Osseointegration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50825303 ↗
  • InjurySep 18, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2026 in ADA 19. Details of surgery: Implant surface not completely covered with bone. On (B)(6) 2026, non-osseointegration was verified. Patient presented with bone type III and fair oral hygiene. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50824477 ↗
  • InjurySep 18, 2026Loss of Osseointegration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50824386 ↗
  • InjurySep 10, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50719599 ↗
  • InjurySep 7, 2026Loss of Osseointegration

    Patient problems: Fibrosis; Pain; Unspecified Infection

    MAUDE report 50653541 ↗
  • InjuryAug 27, 2026Loss of Osseointegration

    The clinician reports the implant was inserted on (b)(6)2025 in ADA 14. On (b)(6)2026, loss of osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50481367 ↗
  • InjuryAug 27, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2025 in ADA 14. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50480352 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 14. Details of surgery: Ridge augmentation. On (b)(6) 2026, non-osseointegration was verified. Patient presented with previous bone augmentation. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50282843 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 14. Details of surgery: Ridge augmentation. On (b)(6) 2026, non-osseointegration was verified. Patient presented with previous bone augmentation. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50281103 ↗
  • InjuryAug 14, 2026Loss of Osseointegration

    The clinician reports the implant was inserted 2025(b)(6) in ADA 15. Details of surgery: Implant surface not completely covered with bone. On 2025(b)(6), loss of osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis, Mobility, Bleeding and Inflammation. No further patient complications were reported.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Inflammation; Unspecified Infection

    MAUDE report 50259349 ↗
  • InjuryAug 14, 2026Loss of Osseointegration

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Inflammation; Unspecified Infection

    MAUDE report 50258330 ↗
  • InjuryAug 10, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 26. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Pain and Swelling. No further patient complications were reported.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 50199569 ↗
  • InjuryJul 30, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in ADA 18. On 2026(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50088236 ↗
  • InjuryJul 30, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026-(b)(6) in ADA 18. On 2026-(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50087568 ↗
  • InjuryJul 27, 2026Adverse Event Without Identified Device or Use Problem

    The clinician reports that the day the implant was placed in ADA 14, failure occurred upon insertion. Details of surgery: Primary stability not achieved. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50041711 ↗
  • InjuryJul 27, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (B)(6) 2025 in ADA 3. On (B)(6) 2026, loss of osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50023324 ↗
  • InjuryJul 27, 2026Loss of Osseointegration

    The clinician reports the implant was inserted 2025 (b)(6) in ADA 3. On 2026 (b)(6), loss of osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50022770 ↗
  • InjuryJul 6, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN ADA 20. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH PREVIOUS BONE AUGMENTATION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain

    MAUDE report 25722223 ↗
  • InjuryJul 6, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN ADA 20. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH PREVIOUS BONE AUGMENTATION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain

    MAUDE report 25720402 ↗
  • InjuryJul 3, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN ADA 14. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 25705742 ↗
  • InjuryJul 3, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN ADA 14. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 25705210 ↗
  • InjuryJul 2, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 27. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25694004 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices