Skip to content

Straumann® BLC and TLC Implants SLActive Roxolid BL (REF 035.9712S)Recalls & reported problems

Institut Straumann AG · BLC Implant, Ø 5.5mm WB, SLActive® 12mm, Roxolid® · Class 2 · Implant

All models of Straumann BLC and TLC Implants (153)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Straumann BLC implant is a dental implant placed in the jawbone to support a replacement tooth. Its 382 reports are all injury reports with no deaths, and mostly describe the implant failing to bond with the bone. Some describe fibrous scar tissue, pain or infection around the implant. Reports rose to 210 over the last year from 156 in the year before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
390
Deaths reported
0
Injuries reported
390
Recalls naming it
0

Summary

The FDA's MAUDE database contains 390 adverse event reports that list this exact device (DI 07630031764577), received between Jun 7, 2024 and Sep 28, 2026 — classified as 390 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (237 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 6 reportsJul 2024: 0 reportsAug 2024: 3 reportsSep 2024: 3 reportsOct 2024: 4 reportsNov 2024: 23 reportsDec 2024: 6 reportsJan 2025: 15 reportsFeb 2025: 6 reportsMar 2025: 6 reportsApr 2025: 4 reportsMay 2025: 0 reportsJun 2025: 37 reportsJul 2025: 15 reportsAug 2025: 15 reportsSep 2025: 17 reportsOct 2025: 12 reportsNov 2025: 23 reportsDec 2025: 30 reportsJan 2026: 16 reportsFeb 2026: 25 reportsMar 2026: 20 reportsApr 2026: 21 reportsMay 2026: 21 reportsJun 2026: 9 reportsJul 2026: 9 reportsAug 2026: 33 reportsSep 2026: 11 reportsOct 2026: 0 reports20242025202637
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate237
  • Adverse Event Without Identified Device or Use Problem113
  • Loss of Osseointegration25
  • Difficult to Insert9
  • Separation Failure5
  • Fracture3

Patient problems

  • No Clinical Signs, Symptoms or Conditions221
  • Fibrosis128
  • Pain65
  • Unspecified Infection35
  • Hyperresponsive to stimuli16
  • Swelling/ Edema15
  • Inflammation9
  • Hemorrhage/Blood Loss/Bleeding8

Recent adverse event reports

The 25 most recent of 390 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Fibrosis

    MAUDE report 50921592 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50920457 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Fibrosis

    MAUDE report 50918761 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918684 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50761030 ↗
  • InjurySep 14, 2026Separation Failure

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50759714 ↗
  • InjurySep 14, 2026Loss of Osseointegration

    Patient problems: Abscess; Fibrosis; Pain; Swelling/ Edema; Unspecified Infection

    MAUDE report 50759576 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 14. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50758034 ↗
  • InjurySep 14, 2026Separation Failure

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50757708 ↗
  • InjurySep 14, 2026Loss of Osseointegration

    Patient problems: Abscess; Fibrosis; Pain; Swelling/ Edema; Unspecified Infection

    MAUDE report 50757637 ↗
  • InjurySep 10, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50719525 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50415469 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50413397 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted on (b)(6)2026 in ADA 14. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On (b)(6)2026, non-osseointegration was verified. Patient presented with bone type IV. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50283728 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in ADA 3. Details of surgery: Implant surface not completely covered with bone and Sinus augmentation. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III and inadequate bone quality/quantity. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50283702 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 21. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50283698 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6)2026 in ADA 30. Details of surgery: Ridge augmentation and Immediate extraction site. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50283647 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 15. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50283399 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Peri-implantitis. No further patient complications were reported.

    Patient problems: Unspecified Infection

    MAUDE report 50283355 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 14. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type IV. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50281319 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in ADA 3. Details of surgery: Implant surface not completely covered with bone and Sinus augmentation. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III and inadequate bone quality/quantity. The device was forwarded to the manufacture. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50281299 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 21. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50281293 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 30. Details of surgery: Ridge augmentation and Immediate extraction site. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50281261 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2026 in ADA 15. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50281226 ↗
  • InjuryAug 14, 2026Failure to Osseointegrate

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Pain; Unspecified Infection

    MAUDE report 50259519 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page