Skip to content

Straumann® BLC and TLC Implants SLActive Roxolid TL (REF 035.7306S)Recalls & reported problems

Institut Straumann AG · TLC SP Implant, Ø 3.75mm RT, SLActive® 6mm, Roxolid® · Class 2 · Implant

All models of Straumann BLC and TLC Implants (153)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This is a Straumann SLActive Roxolid TL dental implant, a screw-shaped root-form implant used to anchor a replacement tooth in the jawbone. The 39 reports mostly describe the implant failing to fuse with the bone, or fusing and then losing that bond, with patients noting pain, swelling, scar tissue formation and a loose implant. No deaths were reported; 39 injuries were. There are no recalls naming this device. Reports rose from 10 in the earlier 12 months to 28 in the most recent one.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
45
Deaths reported
0
Injuries reported
45
Recalls naming it
0

Summary

The FDA's MAUDE database contains 45 adverse event reports that list this exact device (DI 07630031765222), received between Jun 17, 2024 and Sep 28, 2026 — classified as 45 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (31 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 1 reportsOct 2025: 3 reportsNov 2025: 5 reportsDec 2025: 6 reportsJan 2026: 4 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 2 reportsMay 2026: 4 reportsJun 2026: 4 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 6 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate31
  • Loss of Osseointegration11
  • Adverse Event Without Identified Device or Use Problem3

Patient problems

  • No Clinical Signs, Symptoms or Conditions21
  • Fibrosis20
  • Pain14
  • Inflammation5
  • Unspecified Infection3
  • Hypersensitivity/Allergic reaction2
  • Hyperresponsive to stimuli1

Recent adverse event reports

The 25 most recent of 45 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922011 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    Patient problems: Fibrosis; Inflammation; Pain

    MAUDE report 50921438 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    Patient problems: Pain

    MAUDE report 50919745 ↗
  • InjurySep 14, 2026Loss of Osseointegration

    Patient problems: Fibrosis; Inflammation; Pain; Unspecified Infection

    MAUDE report 50760322 ↗
  • InjurySep 14, 2026Loss of Osseointegration

    Patient problems: Fibrosis; Inflammation; Pain; Unspecified Infection

    MAUDE report 50760321 ↗
  • InjurySep 11, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in FDI 35. On (b)(6) 2026, non-osseointegration was verified. Patient presented with inadequate bone quality/quantity. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50738124 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2026 in FDI 15. Details of surgery: Ridge augmentation. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type IV and previous bone augmentation. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50282117 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 34. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50281833 ↗
  • InjuryJun 19, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25564666 ↗
  • InjuryJun 15, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25508843 ↗
  • InjuryJun 15, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN FDI 46. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 25507092 ↗
  • InjuryJun 4, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN FDI 37. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 25395877 ↗
  • InjuryMay 29, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25335630 ↗
  • InjuryMay 15, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 26. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain

    MAUDE report 25190153 ↗
  • InjuryMay 15, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 26. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain

    MAUDE report 25188690 ↗
  • InjuryMay 7, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25114716 ↗
  • InjuryApr 11, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24846560 ↗
  • InjuryApr 8, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2025 IN FDI 15. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE AND INADEQUATE BONE QUALITY/QUANTITY. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24819281 ↗
  • InjuryMar 19, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 35. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 24642133 ↗
  • InjuryJan 8, 2026Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN ADA 30, FAILURE OCCURRED UPON INSERTION. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND RIDGE AUGMENTATION. PATIENT PRESENTED WITH BONE TYPE III AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 24008911 ↗
  • InjuryJan 8, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 25. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN, MOBILITY AND HYPERSENSITIVITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hypersensitivity/Allergic reaction; Pain

    MAUDE report 24008330 ↗
  • InjuryJan 8, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24008328 ↗
  • InjuryJan 8, 2026Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN ADA 30, FAILURE OCCURRED UPON INSERTION. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND RIDGE AUGMENTATION. PATIENT PRESENTED WITH BONE TYPE III AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 24007917 ↗
  • InjuryDec 15, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 25. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain

    MAUDE report 23806943 ↗
  • InjuryDec 12, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23793024 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page