Skip to content

Straumann® BLC and TLC Implants SLActive Roxolid TL (REF 035.7610S)Recalls & reported problems

Institut Straumann AG · TLC SP Implant, Ø 4.5mm WT, SLActive® 10mm, Roxolid® · Class 2 · Implant

All models of Straumann BLC and TLC Implants (153)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Straumann TLC implant is a dental implant placed in the jawbone to support a replacement tooth. All 281 reports are injury reports, mostly describing implants that did not bond with the bone, with some becoming loose and patients reporting pain, swelling and infection. Reports rose from 96 in the previous 12 months to 163 in the last 12 months. There were no deaths, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
293
Deaths reported
0
Injuries reported
293
Recalls naming it
0

Summary

The FDA's MAUDE database contains 293 adverse event reports that list this exact device (DI 07630031766779), received between Mar 22, 2024 and Sep 28, 2026 — classified as 293 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (225 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 3 reportsJul 2024: 9 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 6 reportsNov 2024: 16 reportsDec 2024: 0 reportsJan 2025: 5 reportsFeb 2025: 3 reportsMar 2025: 12 reportsApr 2025: 3 reportsMay 2025: 0 reportsJun 2025: 23 reportsJul 2025: 7 reportsAug 2025: 10 reportsSep 2025: 7 reportsOct 2025: 10 reportsNov 2025: 10 reportsDec 2025: 17 reportsJan 2026: 15 reportsFeb 2026: 13 reportsMar 2026: 18 reportsApr 2026: 19 reportsMay 2026: 25 reportsJun 2026: 18 reportsJul 2026: 24 reportsAug 2026: 8 reportsSep 2026: 8 reportsOct 2026: 0 reports20242025202625
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate225
  • Adverse Event Without Identified Device or Use Problem36
  • Loss of Osseointegration22
  • Difficult to Insert6
  • Separation Failure4

Patient problems

  • No Clinical Signs, Symptoms or Conditions171
  • Fibrosis92
  • Pain47
  • Unspecified Infection21
  • Inflammation13
  • Swelling/ Edema12
  • Hemorrhage/Blood Loss/Bleeding7
  • Hyperresponsive to stimuli6

Recent adverse event reports

The 25 most recent of 293 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 46. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50921609 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50921182 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: Hyperresponsive to stimuli; Unspecified Infection

    MAUDE report 50921122 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50760423 ↗
  • InjurySep 11, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50738188 ↗
  • InjurySep 11, 2026Loss of Osseointegration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50738177 ↗
  • InjurySep 10, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50719627 ↗
  • InjurySep 10, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50719626 ↗
  • InjuryAug 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. Patient presented with inadequate bone quality/quantity. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50481868 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50415684 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50415678 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in FDI 16. On 2026(b)(6), non-osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50415231 ↗
  • InjuryAug 14, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6)2026 in the patient's mouth. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50258951 ↗
  • InjuryAug 13, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50238396 ↗
  • InjuryAug 10, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. Details of surgery: Sinus augmentation and Ridge augmentation. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50199389 ↗
  • InjuryAug 6, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in FDI 36. On 2026(b)(6), non-osseointegration was verified. Patient presented with inadequate bone quality/quantity. No product was returned to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50163470 ↗
  • InjuryJul 31, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted on (b)(6)2026 in FDI 46. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50103408 ↗
  • InjuryJul 31, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 36. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50103406 ↗
  • InjuryJul 31, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (B)(6) 2026 in FDI 37. On (B)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50103405 ↗
  • InjuryJul 31, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in the patient's mouth. On 2026(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50102771 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 46. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50042865 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6)2026 in the patient's mouth. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50042181 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in ADA 18. Details of surgery: Primary stability not achieved, Ridge augmentation and Immediate extraction site. On (b)(6) 2025, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50041843 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in ADA 3. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50041762 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50041756 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page