Straumann® BLC and TLC Implants SLA Roxolid BL (REF 035.9312S)Recalls & reported problems

Institut Straumann AG · BLC Implant, Ø 4.0mm RB, SLActive 12mm, Roxolid · Class 2 · Implant

All models of Straumann BLC and TLC Implants (153)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Straumann BLC dental implant is a screw-shaped post placed in the jawbone. The 82 reports mostly describe the implant failing to bond with the jawbone, and some mention infection and loose implants. Reports rose from 0 in the previous 12 months to 82 in the last 12 months, and there are no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
94
Deaths reported
0
Injuries reported
94
Recalls naming it
0

Summary

The FDA's MAUDE database contains 94 adverse event reports that list this exact device (DI 07630031780096), received between Dec 15, 2025 and Sep 28, 2026 — classified as 94 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (62 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 4 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 4 reportsApr 2026: 12 reportsMay 2026: 13 reportsJun 2026: 11 reportsJul 2026: 14 reportsAug 2026: 21 reportsSep 2026: 15 reportsOct 2026: 0 reports20242025202621
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate62
  • Adverse Event Without Identified Device or Use Problem15
  • Loss of Osseointegration11
  • Difficult to Insert4
  • Separation Failure2

Patient problems

  • No Clinical Signs, Symptoms or Conditions59
  • Fibrosis24
  • Pain14
  • Unspecified Infection12
  • Swelling/ Edema6
  • Hyperresponsive to stimuli3
  • Abscess2
  • Hemorrhage/Blood Loss/Bleeding2

Recent adverse event reports

The 25 most recent of 94 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Loss of Osseointegration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50920393 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918623 ↗
  • InjurySep 17, 2026Loss of Osseointegration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50807733 ↗
  • InjurySep 17, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 29. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50806516 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50759796 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 10. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50757774 ↗
  • InjurySep 11, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50738533 ↗
  • InjurySep 11, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 29. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III and inadequate bone quality/quantity. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50737617 ↗
  • InjurySep 10, 2026Failure to Osseointegrate

    Patient problems: Fibrosis; Pain

    MAUDE report 50719582 ↗
  • InjurySep 10, 2026Separation Failure

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50719477 ↗
  • InjurySep 10, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50719466 ↗
  • InjurySep 10, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50719447 ↗
  • InjurySep 10, 2026Separation Failure

    The clinician reports that the day the implant was placed in ADA 11, failure occurred upon insertion. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. Patient presented with bone type III. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50718744 ↗
  • InjurySep 10, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50718710 ↗
  • InjurySep 10, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026 (b)(6) in ADA 11. On 2026 (b)(6), non-osseointegration was verified. Patient presented with previous bone augmentation. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50718687 ↗
  • InjuryAug 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 12. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III and previous bone augmentation. The device was forwarded to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50481417 ↗
  • InjuryAug 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 12. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III and previous bone augmentation. The device was forwarded to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50480376 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in ADA 10. Details of surgery: Implant surface not completely covered with bone, Ridge augmentation and Immediate extraction site. On 2026(b)(6), non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Pain. No further patient complications were reported.

    Patient problems: Pain

    MAUDE report 50416188 ↗
  • InjuryAug 24, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 5. Details of surgery: Implant surface not completely covered with bone. On 2026-07-22, loss of osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50416151 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6)2026 in FDI 12. Details of surgery: Implant surface not completely covered with bone. On (b)(6)2026, non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Pain, Mobility, Swelling, Bleeding and Fistula. No further patient complications were reported.

    Patient problems: Fibrosis; Fistula; Hemorrhage/Blood Loss/Bleeding; Pain; Swelling/ Edema

    MAUDE report 50415459 ↗
  • InjuryAug 24, 2026Difficult to Insert

    The clinician reports that the day the implant was placed in ADA 29, failure occurred upon insertion. Details of surgery: Primary stability not achieved, Implant surface not completely covered with bone, Ridge augmentation and Immediate extraction site. Patient presented with bone type III and inadequate bone quality/quantity. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50415122 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: Pain

    MAUDE report 50413547 ↗
  • InjuryAug 24, 2026Loss of Osseointegration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50413538 ↗
  • InjuryAug 21, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 24. Details of surgery: Immediate extraction site. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bruxism. The device was forwarded to the manufacturer. At the event the patient experienced: Infection, Mobility and Increased Sensitivity. No further patient complications were reported.

    Patient problems: Fibrosis; Hyperresponsive to stimuli; Unspecified Infection

    MAUDE report 50378206 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 9. Details of surgery: Ridge augmentation. On (b)(6) 2026, non-osseointegration was verified. Patient presented with previous bone augmentation. The device was forwarded to the manufacturer. At the event the patient experienced: Pain, Mobility and Increased Sensitivity. No further patient complications were reported.

    Patient problems: Fibrosis; Hyperresponsive to stimuli; Pain

    MAUDE report 50283609 ↗

Other Implant, endosseous, root-form devices

All Implant, endosseous, root-form devices →