Straumann® BLC and TLC Implants SLA Roxolid BL (REF 035.9314S)Recalls & reported problems

Institut Straumann AG · BLC Implant, Ø 4.0mm RB, SLActive 14mm, Roxolid · Class 2 · Implant

All models of Straumann BLC and TLC Implants (153)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Straumann BLC dental implant is placed in the jaw to hold replacement teeth. The 23 reports mostly describe implants that did not bond with the bone and became loose, sometimes with pain or swelling, with no deaths and 23 injuries mentioned. No recalls name this device. Reports rose from 0 in the earlier period to 23 in the recent period.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
28
Deaths reported
0
Injuries reported
28
Recalls naming it
0

Summary

The FDA's MAUDE database contains 28 adverse event reports that list this exact device (DI 07630031780102), received between Jan 15, 2026 and Sep 28, 2026 — classified as 28 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (20 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 4 reportsApr 2026: 0 reportsMay 2026: 6 reportsJun 2026: 0 reportsJul 2026: 6 reportsAug 2026: 3 reportsSep 2026: 7 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate20
  • Adverse Event Without Identified Device or Use Problem6
  • Difficult to Insert2

Patient problems

  • No Clinical Signs, Symptoms or Conditions14
  • Fibrosis10
  • Pain6
  • Hyperresponsive to stimuli2
  • Swelling/ Edema2
  • Unspecified Infection2

Recent adverse event reports

The 25 most recent of 28 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50919961 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918490 ↗
  • InjurySep 14, 2026Difficult to Insert

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50760722 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50759776 ↗
  • InjurySep 14, 2026Difficult to Insert

    The clinician reports that the day the implant was placed in ADA 23, failure occurred upon insertion. Patient presented with bone type III. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50757921 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 26. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50757760 ↗
  • InjurySep 10, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50719622 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026 (b)(6) in FDI 11. Details of surgery: Implant surface not completely covered with bone, Ridge augmentation and Immediate extraction site. On 2026 (b)(6), non-osseointegration was verified. Patient presented with bone type III and bruxism. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50282512 ↗
  • InjuryAug 3, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2026 in ADA 12. Details of surgery: Ridge augmentation and Immediate extraction site. On (B)(6) 2026, non-osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Unspecified Infection

    MAUDE report 50119786 ↗
  • InjuryAug 3, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026-(b)(6) in ADA 12. Details of surgery: Ridge augmentation and Immediate extraction site. On 2026-(b)(6), non-osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Unspecified Infection

    MAUDE report 50118140 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 27. Details of surgery: Ridge augmentation. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50043151 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 27. Details of surgery: Primary stability not achieved. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50040224 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 27. Details of surgery: Ridge augmentation. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50039921 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 27. Details of surgery: Primary stability not achieved. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50039345 ↗
  • InjuryJul 17, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN ADA 23. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND SWELLING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 25858934 ↗
  • InjuryJul 17, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN ADA 23. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND SWELLING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 25858233 ↗
  • InjuryMay 28, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN ADA 12. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND INCREASED SENSITIVITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Hyperresponsive to stimuli; Pain

    MAUDE report 25320693 ↗
  • InjuryMay 28, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN ADA 12. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND INCREASED SENSITIVITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Hyperresponsive to stimuli; Pain

    MAUDE report 25319880 ↗
  • InjuryMay 26, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN FDI 23. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 25284880 ↗
  • InjuryMay 26, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25283284 ↗
  • InjuryMay 26, 2026Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN ADA 12, FAILURE OCCURRED UPON INSERTION. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. PATIENT PRESENTED WITH BONE TYPE IV AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 25282748 ↗
  • InjuryMay 26, 2026Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN ADA 12, FAILURE OCCURRED UPON INSERTION. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. PATIENT PRESENTED WITH BONE TYPE IV AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 25281508 ↗
  • InjuryMar 30, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN ADA 6. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE, RIDGE AUGMENTATION AND IMMEDIATE EXTRACTION SITE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 24733535 ↗
  • InjuryMar 30, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6)2025 IN ADA 6. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE, RIDGE AUGMENTATION AND IMMEDIATE EXTRACTION SITE. ON (B)(6)2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 24732457 ↗
  • InjuryMar 16, 2026Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN ADA 6, FAILURE OCCURRED UPON INSERTION. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED, RIDGE AUGMENTATION AND IMMEDIATE EXTRACTION SITE. PATIENT PRESENTED WITH BONE TYPE IV, FAIR ORAL HYGIENE AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24609353 ↗