Shoulder System 04.01.0159Recalls & reported problems

Medacta International SA · Glenoid polyaxial locking screw - L22 · Class 2 · Implant

All models of Shoulder System (63)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07630040706469), received between Jul 31, 2020 and Nov 18, 2025 — classified as 4 injuries, 1 malfunction.

The most frequently reported specific device problem is “Fracture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture3
  • Appropriate Device Problem Term/Code Not Available1
  • Device Appears to Trigger Rejection1
  • Loose or Intermittent Connection1

Patient problems

  • Bone Fracture(s)2
  • Medical device site infection1
  • Metal Related Pathology1
  • Nerve Damage1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 18, 2025Fracture; Fracture

    INTRAOPERATIVELY, THE SURGEON INITIALLY JUDGED THAT THE HEAD HAD MERELY BEEN STRIPPED. HOWEVER, A POSTOPERATIVE X-RAY REVEALED THAT THE BROKEN SCREW HEAD HAD MIGRATED AND ENTERED THE HUMERUS. ON THE SAME DAY, ADDITIONAL SURGICAL INTERVENTION WAS PERFORMED TORE MOVE THE SCREW FRAGMENT THROUGH AN INCISION. THE CAUSE IS UNKNOWN, BUT IT IS POSSIBLE THAT THE SCREW HEAD BROKE DURING THE FINAL TIGHTENING PHASE. THE SURGERY WAS COMPLETED SUCCESSFULLY. THE TOTAL SURGERY TIME WAS 2 HOURS, AND THE DELAY TIME WAS 1 HOUR. THE SCREW INVOLVED WAS LOCATED IN THE SUPERIOR PART OF THE BASEPLATE. A TOTAL OF FOUR POLYAXIAL SCREWS, INCLUDING THE BROKEN ONE, WERE PLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23583100 ↗
  • InjuryDec 27, 2024Fracture; Device Appears to Trigger Rejection; Fracture; Device Appears to Trigger Rejection

    REVISION SURGERY PERFORMED 1 YEAR AND 9 MONTHS AFTER THE PRIMARY SURGERY DUE TO SUSPECTED SEPTIC LOOSENING AND METALLOSIS ON THE HUMERAL AND GLENOID PART. BONE GRAFT USED IN PRIMARY SURGERY. ONE POLYAXIAL SCREW WAS FOUND BROKEN. ALL IMPLANTS WERE REMOVED AND A SPACER WAS PUT IN.

    Patient problems: Medical device site infection; Metal Related Pathology

    MAUDE report 21026680 ↗
  • MalfunctionAug 11, 2022Fracture; Fracture

    2 GLENOID POLYAXIAL LOCKING SCREW PRONG HEADS BROKE WHILE THE SURGEON WAS TRYING TO INSERT THEM IN THE GLENOID BASEPLATE (1X 22MM, 1X 26MM). FINALLY A GLENOID BONE FRACTURE OCCURRED AND THE SURGEON DECIDED TO SWITCH TO AN ANATOMICAL SYSTEM. THE BONE WAS VERY HARD.

    Patient problems: Bone Fracture(s)

    MAUDE report 15208238 ↗
  • InjuryJun 24, 2021Loose or Intermittent Connection

    THE PATIENT CAME IN REPORTING PAIN DUE TO THE SCREW PLACED IN THE GLENOID IN THE PRIMARY HITTING A NERVE. 8 MONTHS AFTER THE PRIMARY, THE SURGEON REVISED THE SCREW WITH A SHORTER SCREW. THE SURGEON ALSO REVISED THE GLENOSPHERE, HUMERAL REVERSE METAPHYSIS, AND HUMERAL REVERSE LINER SINCE THE JOINT LOOSENED UP INTRAOPERATIVELY UNEXPECTEDLY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Nerve Damage

    MAUDE report 12057829 ↗
  • InjuryJul 31, 2020Appropriate Device Problem Term/Code Not Available

    DURING THE FIRST SURGERY (2018) THERE WAS AN INTRA-OP CAUDAL GLENOID FRACTURE DURING THE IMPLANTATION OF AN INVERSE SHOULDER IMPLANT. THE SURGEON REPORTED TO AGENT THAT HE HAS IN SUCH CASES 2 OPTIONS: 1) GO FOR A HEMI PROTHESIS OR 2) GO FOR AN INVERSE PROTHESIS BUT IMPLANT IT TOO CRANIAL AND THEN LATER REVISE THE GLENOID SIDE FOR A ANATOMIC POSITION AFTER FRACTURE HEALING AND IN CASE OF PROBLEM. THE SURGEON HAS CHOSEN OPTION 2 TO GO ON. THIS REVISION OF TODAY WAS THE REVISION OF THE OPTION 2 - A "PLANNED REVISION". THE SURGEON TOLD ME THAT THE REVISION IS NOT IMPLANT RELATED. DIAPHYSIS WAS NOT REMOVED. ALL OTHER COMPONENTS WERE REMOVED AND SENT TO SONIFICATION (WHY NOT AVAILABLE).

    Patient problems: Bone Fracture(s)

    MAUDE report 10349979 ↗