Shoulder System 04.01.0172Recalls & reported problems

Medacta International SA · Glenoidsphere 36xØ27 · Class 2 · Implant

All models of Shoulder System (63)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 07630040706599), received between Jan 10, 2020 and Mar 26, 2025 — classified as 8 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Appropriate Device Problem Term/Code Not Available2
  • Device Dislodged or Dislocated2
  • Fracture1
  • Migration1

Patient problems

  • Bone Fracture(s)3
  • Joint Dislocation2
  • Failure of Implant1
  • Loss of Range of Motion1
  • Medical device site infection1
  • Toxicity1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACROMIAL STRESS FRACTURE OCCURRED APPROXIMATELY 3 MONTHS POST-PRIMARY SURGERY. THE PATIENT HAS NOT BEEN REVISED, THE SURGEON OPTED FOR A CONSERVATIVE TREATMENT. THE PATIENT HAS SIGNIFICANT RHEUMATOID AND OSTEOARTHRITIS, AND ALMOST COMPLETE OBLITERATION OF THE GLENOHUMERAL JOINT.

    Patient problems: Bone Fracture(s)

    MAUDE report 21695694 ↗
  • InjuryMar 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACROMIAL STRESS FRACTURE OCCURRED APPROXIMATELY 3 MONTHS POST SURGERY. THE PATIENT HAS NOT BEEN REVISED, THE SURGEON OPTED FOR A CONSERVATIVE TREATMENT.

    Patient problems: Bone Fracture(s)

    MAUDE report 21547427 ↗
  • InjuryDec 7, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY DUE TO INFECTION PERFORMED ABOUT 1 MONTH AFTER THE PRIMARY SURGERY. GLENOSPHERE AND LINER REVISED SUCCESSFULLY.

    Patient problems: Medical device site infection

    MAUDE report 18281373 ↗
  • InjuryDec 6, 2023Migration; Migration

    AT 3 YEARS AND 5 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN AND A LIMITED RANGE OF MOTION DUE TO DISSOCIATION OF THE GLENOSPHERE FROM THE BASEPLATE. THE CAUSE OF THE DISSOCIATION IS UNKNOWN. THE SURGEON REVISED THE LINER AND GLENOSPHERE AND THE SURGERY WAS COMPLETED SUCCESSFULLY. THERE ARE SCREW RING MARKS ON THE BACK SIDE OF THE GLENOSPHERE, AND TWO POLYAXIAL SCREW HEADS WERE ALSO FOUND TO BE DAMAGED. A MILD DEGREE OF METALLOSIS WAS ALSO OBSERVED.

    Patient problems: Loss of Range of Motion

    MAUDE report 18272871 ↗
  • InjuryJun 28, 2022Device Dislodged or Dislocated; Appropriate Device Problem Term/Code Not Available

    AT ABOUT 4 MONTHS AFTER THE PRIMARY, REVISION OF METHAPHYSIS, GLENOSPHERE, AND LINER FOLLOWING SHOULDER LUXATION . THE DEFICIT OF THE MUSCLE-TENDON COMPLEX, LINKED TO INACTIVITY AND MASS, MAY HAVE REDUCED JOINT STABILITY.

    Patient problems: Joint Dislocation

    MAUDE report 14839514 ↗
  • InjuryAug 3, 2021Device Dislodged or Dislocated

    REVISION SURGERY PERFORMED 5 MONTHS AFTER PRIMARY FOLLOWING INSTABILITY AND LUXATION (GLENOSPHERE DISLOCATED FROM THE LINER). THE GLENOSPHERE WAS CHANGED FROM A 36 TO A 39 TO PROVIDE MORE STABILITY. THE PATIENT IS IN A WHEELCHAIR AND IS AFFECTED BY MULTIPLE SCLEROSIS.

    Patient problems: Failure of Implant; Joint Dislocation

    MAUDE report 12266810 ↗
  • InjuryJun 25, 2021Fracture

    PRIMARY SURGERY PERFORMED IN (B)(6) 2021. FRACTURE OF THE ACROMION DETECTED 1 MONTH AFTER THE SURGERY. NO REVISION SURGERY SCHEDULED AT THIS TIME.

    Patient problems: Bone Fracture(s)

    MAUDE report 12067299 ↗
  • InjuryJan 10, 2020Appropriate Device Problem Term/Code Not Available

    ABOUT 7 MONTHS AFTER PRIMARY REVISION SURGERY PERFORMED DUE TO DISSOCIATION OF THE GLENOSPHERE FROM THE GLENOID BASEPLATE, DURING THE REVISION SURGERY METALLOSIS HAS BEEN DISCOVERED. THE GLENOSPHERE, AND THE HUMERAL REVERSE HC LINER HAVE BEEN REPLACED WITH NEW IMPLANTS, THE GLENOID BASEPLATE WAS LEFT IN PLACE.

    Patient problems: Toxicity

    MAUDE report 9574062 ↗

Other Shoulder Prosthesis, Reverse Configuration devices

All Shoulder Prosthesis, Reverse Configuration devices →