Shoulder System 04.01.0173Recalls & reported problems
Medacta International SA · Glenoidsphere 39xØ27 · Class 2 · Implant
All models of Shoulder System (63)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Medacta Glenosphere is the ball-and-socket component used in reverse shoulder replacement surgery. The 51 reports mostly describe the implant coming loose or dislocating, with patients needing repeat operations to fix it, and also mention infections and bone fractures around the implant site. There are 49 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 51
- Deaths reported
- 0
- Injuries reported
- 49
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 51 adverse event reports that list this exact device (DI 07630040706605), received between Sep 21, 2018 and Jun 9, 2026 — classified as 49 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (28 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated28
- Appropriate Device Problem Term/Code Not Available11
- Adverse Event Without Identified Device or Use Problem4
- Device Appears to Trigger Rejection4
- Migration4
- Detachment of Device or Device Component2
- Device Slipped2
- Loose or Intermittent Connection2
Patient problems
- Joint Dislocation20
- Insufficient Information7
- Bone Fracture(s)6
- Medical device site infection4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Failure of Implant2
- Implant Pain2
- Joint Laxity2
Recent adverse event reports
The 25 most recent of 51 reports, in the reporter's own words as published by the FDA.
InjuryJun 9, 2026Device Dislodged or Dislocated
REVISION SURGERY AT ABOUT 3 WEEKS FROM SURGERY FOR SHOULDER LUXATION. THE SURGEON REVISED SUCCESSFULLY GLENOSPHERE, LINER ANDMETAPHYSIS.
Patient problems: Joint Dislocation
MAUDE report 25442525 ↗InjuryMay 27, 2026Adverse Event Without Identified Device or Use Problem
ACROMION FRACTURE POSTOPERATIVELY, ABOUT 11 MONTHS POST PRIMARY. REASON UNKNOWN. THE SURGEON IS TREATED CONSERVATIVELY WITH PHYSIOTHERAPY AND SPLINT/IMMOBILIZATION.
Patient problems: Bone Fracture(s)
MAUDE report 25297690 ↗InjuryApr 20, 2026Device Dislodged or Dislocated
JOINT LUXATION OCCURRED ABOUT 1 WEEK AFTER THE PRIMARY SURGERY. ANOTHER LUXATION OCCURRED TWO WEEKS AFTER THE PRIMARY SURGERY. ABOUT 2 WEEKS AFTER THE PRIMARY SURGERY, THE PATIENT WAS REVISED. GLENOSPHERE SIZE INCREASED (DIAMETER 42), LINER +6MM AND METAPHYSIS+9MM/+20° R IMPLANTED. PATIENT HAS A BONE GRAFT.
Patient problems: Joint Dislocation
MAUDE report 24924657 ↗InjuryApr 15, 2026Adverse Event Without Identified Device or Use Problem
REVISION SURGERY AFTER 3 MONTHS AFTER THE PREVIOUS SURGERY (IT WAS DUE TO INFECTION) DUE TO SHOULDER INSTABILITY.
Patient problems: Joint Laxity
MAUDE report 24882480 ↗InjuryDec 18, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
REVISION SURGERY DUE TO INFECTION ABOUT 1 MONTH AFTER THE PRIMARY SURGERY. GLENOSPHERE, LINER, AND METAPHYSIS REVISED.
Patient problems: Medical device site infection
MAUDE report 23837285 ↗InjuryNov 5, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
PRIMARY SURGERY PERFORMED ON THE (B)(6) 2025. IMPLANTATION OF REVERSE SHOULDER IMPLANTS. REVISION SURGERY PERFORMED ON THE (B)(6) 2025 DUE TO SUSPECTED INFECTION. EXPLANTATION OF THE HUMERAL PE LINER, GLENOSPHERE AND METAPHYSIS. REIMPLANTATION OF NEW IMPLANTS WITH SAME SIZES.
Patient problems: Medical device site infection
MAUDE report 23480204 ↗InjuryJul 23, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
PRIMARY SURGERY: PERFORMED TO TREAT A FRACTURE WITH A LONG SHAFT. PRIMARY DATE AND DETAILS UNKNOWN. 1ST REVISION (DUE TO INFECTION): THE SURGEON REMOVED ONLY THE GLENOID, LEFT THE SHAFT IN PLACE, AND IMPLANTED A SPACER HEAD WITH CEMENT. SURGERY DATE AND DETAILS UNKNOWN. 2ND REVISION (CUSTOM GLENOID IMPLANTATION): PERFORMED ON (B)(6) 2025 TO IMPLANT CUSTOM GLENOID AFTER THE INFECTION WAS TREATED. 3RD REVISION (DUE TO SHOULDER LUXATION): PERFORMED FOLLOWING MULTIPLE LUXATIONS AFTER THE CUSTOM IMPLANT. IT WAS REPORTED THAT THE REVISION IS DUE TO POOR SOFT TISSUE TENSIONING. SURGERY DATE AND DETAILS UNKNOWN. 4TH REVISION (PLANNED FOR (B)(6) 2025): AFTER FURTHER DISLOCATIONS, DR. (B)(6) CONSULTED PROF. (B)(6) WHO WILL PERFORM A REVISION SURGERY IN (B)(6).
Patient problems: Joint Dislocation
MAUDE report 22574585 ↗InjuryJun 25, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
REVISION SURGERY DUE TO INFECTION AT ABOUT 2 YEARS AND 5 MONTHS FROM THE PRIMARY. NO FINAL IMPLANTS WERE IMPLANTED, ONLY A SPACER BECAUSE OF THE INFECTION. LATER ON, THE SURGEON WILL PUT IN NEW IMPLANTS.
Patient problems: Medical device site infection
MAUDE report 22326363 ↗InjuryMay 8, 2025Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem; Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem
AT ABOUT 1 MONTH AFTER PRIMARY, THE PATIENT HAS BEEN REVISED BECAUSE OF RECURRENT JOINT LUXATION THAT OCCURRED DURING HOSPITALIZATION AND AT HOME AFTER THE SURGERY. LINER, GLENOSPHERE AND METAPHYSIS REVISED SUCCESSFULLY.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 21984140 ↗InjuryApr 4, 2025Device Appears to Trigger Rejection; Device Dislodged or Dislocated; Device Appears to Trigger Rejection; Device Dislodged or Dislocated
REVISION SURGERY DUE TO INFECTION AND INSTABILITY, PATIENT REPORTED ALSO DISLOCATION. GLENOSPHERE, LINER AND THE METAPHYSIS REVISED. PRIMARY SURGERY PERFORMED BY DR (B)(6) ON THE (B)(6)2021. IMPLANTATION OF REVERSE SHOULDER COMPONENTS + BONE GRAFT AND WITH A CEMENTED STANDARD HUMERAL STEM. MYSHOULDER PLANNING ONLY (M_MIC_RTS_YB_04031948). 1ST REVISION SURGERY PERFORMED ON THE (B)(6) 2024. EXPLANTATION OF ALL PRIMARY COMPONENTS. REIMPLANTATION OF NEW REVERSE SHOULDER COMPONENTS + BONE GRAFT AND WITH A LONG HUMERAL STEM. 2ND REVISION SURGERY PERFORMED ON THE (B)(6) 2024. EXPLANTATION OF THE LONG HUMERAL STEM WITH THE METAPHYSIS AND THE LINER + EXPLANTATION OF THE GLENOSPHERE (TO TEST THE STABILITY OF THE BASEPLATE AND THE GRAFT). REIMPLANTATION OF A NEW LONG HUMERAL STEM (CEMENTED) WITH A NEW METAPHYSIS AND LINER +6MM (155 DEGREES) + IMPLANTATION OF A NEW GLENOSPHERE (LATERALIZED). 3RD REVISION SURGERY PERFORMED ON THE (B)(6) 2024. EXPLANTATION OF ALL COMPONENTS EXCEPT THE LONG HUMERAL STEM. IMPLANTATION OF A GLENOID CUSTOM IMPLANT ON THE (B)(6) 2024 WITH MY SHOULDER PLANNING (M_MOU_RTS_YB_04031948). 4TH REVISION SURGERY PERFORMED ON THE (B)(6) 2025. EXPLANTATION OF THE GLENOSPHERE,
Patient problems: Joint Dislocation; Joint Laxity; Medical device site infection
MAUDE report 21771301 ↗MalfunctionMar 25, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
A NOT FULLY SEATED GLENOSPHERE SCREW WAS NOTED FROM POST-OPERATIVE XRAYS PERFORMED ABOUT 3 MONTHS POST PRIMARY SURGERY. DURING THE SURGERY, THE GLENOSPHERE WAS INSERTED USING THE RETENTIVE GLENOSPHERE IMPACTOR OVER THE FLEXIBLE GLENOSPHERE GUIDE. THE TORQUE LIMITING SCREW DRIVER T15 3NM WAS USED TO SCREW THE GLENOSPHERE SCREW INTO THE BASEPLATE. AT THE MOMENT, THERE IS NO INTENTION TO PERFORM AN ADDITIONAL SURGERY.
Patient problems: Insufficient Information
MAUDE report 21687568 ↗MalfunctionMar 21, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
A NOT FULLY SEATED GLENOSPHERE SCREW WAS NOTED FROM POST-OPERATIVE XRAYS PERFORMED ABOUT 2 MONTHS POST PRIMARY SURGERY. DURING THE SURGERY, THE GLENOSPHERE WAS INSERTED USING THE RETENTIVE GLENOSPHERE IMPACTOR OVER THE FLEXIBLE GLENOSPHERE GUIDE. THE TORQUE LIMITING SCREW DRIVER T15 3NM WAS USED TO SCREW THE GLENOSPHERE SCREW INTO THE BASEPLATE. AT THE MOMENT, THERE IS NO INTENTION TO PERFORM AN ADDITIONAL SURGERY.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 21662620 ↗InjuryFeb 25, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
ACROMIAL STRESS FRACTURE OCCURRED APPROXIMATELY 2 MONTHS POST SURGERY.
Patient problems: Bone Fracture(s)
MAUDE report 21456686 ↗InjuryFeb 25, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
ACROMIAL STRESS FRACTURE OCCURRED APPROXIMATELY 2 MONTHS POST SURGERY.
Patient problems: Bone Fracture(s)
MAUDE report 21456685 ↗InjuryFeb 25, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
ACROMIAL STRESS FRACTURE OCCURRED APPROXIMATELY 3 MONTHS POST SURGERY.
Patient problems: Bone Fracture(s)
MAUDE report 21456684 ↗InjuryJan 29, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
REVISION SURGERY DUE TO JOINT LUXATION, AT ABOUT 2 MONTHS AFTER THE PRIMARY. THE SURGEON CHANGED THE HUMERAL LINER AND THE GLENOSPHERE. PRIMARY SURGERY PERFORMED IN (B)(6) 2024 ((B)(6) 2024)
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21262087 ↗InjuryJan 7, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
AT ABOUT 1 MONTH AFTER PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO JOINT LUXATION AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED THE GLENOSPHERE, METAPHYSIS AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Joint Dislocation
MAUDE report 21092769 ↗InjuryNov 23, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
REVISION SURGERY DUE TO JOINT LUXATION, AT ABOUT 7 MONTHS AFTER PRIMARY. THE SURGERY HAS BEEN COMPLETED SUCCESSFULLY REVISING THE LINER AND THE GLENOSPHERE.
Patient problems: Joint Dislocation
MAUDE report 18198434 ↗InjuryApr 26, 2023Migration; Migration
THE PATIENT HAD A PRIMARY SURGERY ON THE (B)(6) 2022. AT ABOUT 10 MONTHS AFTER THE PRIMARY, FROM THE X-RAYS HAS BEEN NOTICED THE UNSCREWING OF THE SCREW FROM THE GLENOSPHERE. THE SURGEON DO NOT PLAN TO REVISE THE IMPLANT AT THIS TIME AND CONTINUES WITH THE FOLLOW-UP OBSERVATION.
Patient problems: Implant Pain
MAUDE report 16813415 ↗InjuryOct 31, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
AT ABOUT 1 MONTH AND 3 WEEKS AFTER THE PRIMARY SURGERY, THE PATIENT CAME REPORTING PAIN DUE TO ANTERIOR DISLOCATION OF THE LINER FROM THE GLENOSPHERE AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED THE LINER AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Implant Pain
MAUDE report 15698469 ↗InjuryAug 18, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
REVISION SURGERY AT ABOUT 4 YEARS AND 3 MONTHS POST PRIMARY FOR GLENOSPHERE INTRAPROSTHETIC DISLOCATION FROM THE BASEPLATE. DURING THE PRIMARY SURGERY ONE OF THE GLENOSPHERE POLYAXIAL SCREW PRONG DEFORMED.
Patient problems: Joint Dislocation
MAUDE report 15247457 ↗InjuryFeb 18, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE SURGEON REVISED THE GLENOSPHERE AND THE LINER 2 YEARS AND 3 MONTHS AFTER PRIMARY FOLLOWING GLENOSPHERE SCREW LOOSENING. THE GLENOSPHERE WAS COMPLETELY MOBILE, BUT NOT DISPLACED. THE SCREW WAS COMPLETELY LOOSE ALSO, BUT STILL IN THE GLENOSPHERE.
Patient problems: Insufficient Information
MAUDE report 13557274 ↗InjuryFeb 15, 2022Device Dislodged or Dislocated; Migration; Device Dislodged or Dislocated; Migration
ABOUT 6 MONTHS AFTER THE PRIMARY SURGERY, THE SCREW OF THE GLENOSPHERE BECAME LOOSE AND THE GLENOSPHERE WAS NOT FIXED ANYMORE. REVISION SURGERY PERFORMED ON (B)(6) 2022.
Patient problems: Failure of Implant
MAUDE report 13532819 ↗InjuryFeb 9, 2022Loose or Intermittent Connection; Appropriate Device Problem Term/Code Not Available
REVISION SURGERY WILL BE PERFORMED ABOUT 1 YEARS AND 7 MONTHS AFTER THE PRIMARY SURGERY DUE TO GLENOSPHERE SCREW LOOSENING.
Patient problems: Insufficient Information
MAUDE report 13498302 ↗InjuryFeb 7, 2022Unintended Movement; Unintended Movement
REVISION SURGERY PERFORMED ABOUT 1 YEAR AND 10 MONTHS AFTER THE PRIMARY SURGERY BECAUSE GLENOSPHERE SCREW SLIPPED OUT. GLENOSPHERE, LINER AND ONE POLYAXIAL LOCKING SCREW (L34 TO L30) REVISED AND RETIGHTEN THE OTHER POLYAXIAL SCREWS. THE SURGEON COMMENTED THAT ONE POLYAXIAL LOCKING SCREW MAY NOT HAVE BEEN INSERTED FIRMLY AND THE SCREW HEAD MAY HAVE INTERFERED WITH THE FIXING OF THE GLENOSPHERE.
Patient problems: Insufficient Information
MAUDE report 13470949 ↗
FDA clearance and classification
Cleared through 510(k) K170452 on Oct 5, 2017 (decision: Substantially Equivalent). Submitted by Medacta International S.A. as “Medacta Shoulder System”. The FDA review took 232 days. FDA record ↗
The FDA regulates product code PHX (shoulder prosthesis, reverse configuration) under 21 CFR 888.3660 as a Class II device. 21 CFR 888.3660 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
FDA definition of this device type: Intended to be used to relieve pain and to replace the glenoid and humeral side of the shoulder joint in order to restore mobility in a grossly rotator cuff deficient joint with severe arthropathy or a previous joint replacement with a grossly rotator cuff deficient joint.
Common questions about the Shoulder System 04.01.0173
Has the Shoulder System 04.01.0173 been recalled?
No FDA recall record in our data names this exact device identifier (DI 07630040706605), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Shoulder System 04.01.0173?
The FDA's MAUDE database lists 51 adverse event reports for this device, received between Sep 21, 2018 and Jun 9, 2026. In those reports, the most common specific device problems are “Device Dislodged or Dislocated” (28), “Device Appears to Trigger Rejection” (4) and “Migration” (4); the most common patient problems are “Joint Dislocation” (20), “Bone Fracture(s)” (6) and “Medical device site infection” (4). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Shoulder System 04.01.0173?
None of the reports are classified as deaths, 49 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Shoulder System 04.01.0173 FDA approved?
It was cleared through 510(k) K170452 on Oct 5, 2017. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the Shoulder System 04.01.0173 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the Shoulder System 04.01.0173, and is it still sold?
The FDA's device database lists Medacta International SA as the company responsible for it (brand Shoulder System, model 04.01.0173). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Shoulder Prosthesis, Reverse Configuration devices
- DJO SURGICAL 508-36-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration513 reports
- DJO SURGICAL 508-32-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration435 reports
- DJO SURGICAL 508-32-103Encore Medical, L.P. · shoulder prosthesis, reverse configuration427 reports
- DJO SURGICAL 509-00-036Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall258 reports
- DJO SURGICAL 508-32-104Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall202 reports
- DJO SURGICAL 509-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration239 reports
- DELTA XTEND 130760000Depuy (Ireland) · shoulder prosthesis, reverse configuration221 reports
- DJO SURGICAL 508-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration124 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.