Skip to content

GMK Total Knee System 02.12.0722RRecalls & reported problems

Medacta International SA · Femoral Component Sphere #2+ R TiNbN coated · Class 2 · Implant

All models of GMK Total Knee System (31)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07630040711531), received between May 27, 2022 and Aug 7, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Malposition of Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Malposition of Device1
  • Positioning Problem1
  • Unstable1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 7, 2025Positioning Problem; Positioning Problem

    DURING TKA THE EFFICIENCY TRIAL FEMORAL COMPONENT WAS USED TO TIRAL THE FEMUR AND DRILL THE HOLES FOR THE FEMORAL IMPLANT PEGS. AFTERWARD, WHEN TRYING TO INSERT THE FINAL IMPLANT THE SURGEON NOTICED THAT THE PEGS OF THE IMPLANTS WERE NOT ALIGNED WITH THE HOLES. THE MISMATCH WAS IN THE ANTERIOR DIRECTION OF APPROXIMATIVELY 5MM. SURGEON REMOVED THE IMPLANT, REDRILLED THE HOLES AND REFINED THE BONE CUTS. THE PATIENT BONE HAD TO BE RECEMENTED AGAIN AND FINALLY IMPLANT WAS POSITIONED. A TOTAL OF 20 MINUTES OF DELAY OVER 45 MINUTES OF SURGERY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22737764 ↗
  • InjuryMay 27, 2022Unstable; Malposition of Device; Malposition of Device

    THE PATIENT CAME IN REPORTING PATELLA INSTABILITY AND THE CAUSE IS UNKNOWN. THE SURGEON PLANS TO REVISE THE PATELLA IN (B)(6) 2022 (PRIMARY WAS ON THE (B)(6) 2021). NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. MORE DETAILS TO FOLLOW ONCE REVISION TAKES PLACE.

    Patient problems: Insufficient Information

    MAUDE report 14522259 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page