This is a Medacta AMIStem-P hip implant stem used in hip replacement surgery. It has 24 adverse event reports, all describing injuries and no deaths. Reports mostly describe the stem loosening or sinking within months or years of surgery, along with bone fractures, poor bone bonding and infections, often requiring revision surgery. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
24
Deaths reported
0
Injuries reported
24
Recalls naming it
0

Summary

The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 07630040720014), received between Dec 19, 2018 and Nov 28, 2025 — classified as 24 injuries.

The most frequently reported specific device problem is “Fracture” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available8
  • Fracture6
  • Loose or Intermittent Connection5
  • Osseointegration Problem3
  • Adverse Event Without Identified Device or Use Problem2
  • Device Appears to Trigger Rejection2
  • Inadequacy of Device Shape and/or Size2
  • Difficult to Remove1

Patient problems

  • Bone Fracture(s)8
  • Inadequate Osseointegration7
  • Medical device site infection2
  • Pain2
  • Failure of Implant1
  • Hip Fracture1
  • Joint Disorder1
  • Patient Problem/Medical Problem1

Recent adverse event reports

The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 28, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    AT ABOUT 5 YEARS AND 9 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN DUE TO AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON REMOVED ALL MEDACTA IMPLANTS AND PUT IN COMPETITOR SPACERS WITH ANTIBIOTIC CEMENT. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Medical device site infection

    MAUDE report 23669494 ↗
  • InjuryJun 18, 2025Implant subsidence; Implant subsidence

    THE PATIENT HAD PAIN DUE TO A SUBSIDED STEM AND THE CAUSE IS UNKNOWN. AT ABOUT 3 MONTHS FROM PRIMARY THE SURGEON REVISED THE STEM AND HEAD AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Pain

    MAUDE report 22269520 ↗
  • InjuryApr 4, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    AT 5 YEARS 8 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON REMOVED ALL MEDACTA HARDWARE AND IMPLANTED A ANTIBIOTIC SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Medical device site infection

    MAUDE report 21771298 ↗
  • InjuryJan 10, 2025Osseointegration Problem; Osseointegration Problem

    THE PATIENT HAD PAIN DUE TO A LOOSE STEM AND THE CAUSE OF THE LOOSE STEM IS UNKNOWN. AT ABOUT 4 YEARS AND 1 MONTH POST PRIMARY THE SURGEON REVISED THE STEM AND HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 21122214 ↗
  • InjuryJul 25, 2024Osseointegration Problem; Osseointegration Problem

    AT ABOUT 1 YEAR 3 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE STEM AND THE CAUSE OF THE LOOSE STEM IS UNKNOWN. THE SURGEON REVISED THE STEM AND HEAD.. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 19827031 ↗
  • InjuryApr 28, 2023Osseointegration Problem; Osseointegration Problem

    AT ABOUT 3 YEARS 5 MONTHS AFTER THE PRIMARY, REVISION SURGERY FOLLOWING ASEPTIC STEM LOOSENING.

    Patient problems: Inadequate Osseointegration

    MAUDE report 16831309 ↗
  • InjuryJul 21, 2022Loose or Intermittent Connection; Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY PERFORMED DUE TO STEM AND CUP LOOSENING 1 YEAR AND 1 MONTH AFTER THE PRIMARY SURGERY. NO INFECTION DETECTED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 15074525 ↗
  • InjuryJun 17, 2022Loose or Intermittent Connection; Appropriate Device Problem Term/Code Not Available

    THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE STEM AND THE CAUSE OF THE LOOSE STEM IS UNKNOWN. THE SURGEON REVISED THE STEM AND HEAD 3 YEARS AND 4 MONTHS AFTER PRIMARY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 14732902 ↗
  • InjurySep 8, 2021Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY PERFORMED AFTER 9 MONTHS FROM THE PRIMARY SURGERY DUE TO AN INFECTION. ALL IMPLANTS WERE REVISED.

    Patient problems: Unspecified Infection

    MAUDE report 12438441 ↗
  • InjuryJul 22, 2021Loose or Intermittent Connection; Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY FOR STEM LOOSENING 3 MONTHS AFTER PRIMARY THR. THE SURGERY WAS COMPLETED SUCCESSFULLY, STEM AND HEAD REVISED SUCCESSFULLY.

    Patient problems: Failure of Implant

    MAUDE report 12210365 ↗
  • InjuryJun 11, 2021Appropriate Device Problem Term/Code Not Available

    2 MONTHS AFTER THE PRIMARY, PERIPROSTHETIC FEMORAL FRACTURE. STEM AND HEAD SUCCESSFULLY REVISED.

    Patient problems: Bone Fracture(s)

    MAUDE report 11981288 ↗
  • InjuryMay 21, 2021Inadequacy of Device Shape and/or Size; Appropriate Device Problem Term/Code Not Available

    THE PATIENT CAME IN REPORTING PAIN AND INSTABILITY DUE TO A LEG LENGTH DISCREPANCY. THE PATIENT'S OPERATED LEG WAS LONGER THAN THEN NON-OPERATED LEG. THE PATIENT HAD PELVIC OBLIQUITY ISSUES AND ON X-RAYS THE LEGS LOOKED EQUAL, BUT CLINICALLY THE PATIENT WAS LONG ON THE OPERATED SIDE. 1 YEAR AND 8 MONTHS AFTER PRIMARY THE SURGEON REVISED SUCCESSFULLY THE PATIENT STEM, HEAD AND LINER.

    Patient problems: Pain

    MAUDE report 11862763 ↗
  • InjuryApr 21, 2021Loose or Intermittent Connection; Unintended Movement

    REVISION SURGERY 4 MONTHS AFTER THE PRIMARY SURGERY FOR STEM MOBILIZATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 11700205 ↗
  • InjuryMar 31, 2021Fracture; Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY 1 WEEK AFTER PRIMARY SURGERY. THE REASON IS ABOUT 2CM FEMUR SPIRAL FRACTURE ON THE LESSER TROCHANTER AND A SMALL FRACTURE ON THE GREAT TROCHANTER CAUSING STEM SUBSIDENCE. THE FEMUR DEVICES HAVE BEEN REPLACED AND THE FRACTURES FIXED WITH COMPETITOR DEVICES.

    Patient problems: Bone Fracture(s)

    MAUDE report 11595555 ↗
  • InjuryDec 14, 2020Fracture

    THE PATIENT CAME IN, 6 DAYS AFTER PRIMARY SURGERY, REPORTING PAIN DUE TO A PERIPROSTHETIC FRACTURE OF THE FEMUR AND THE CAUSE OF THE FRACTURE IS UNKNOWN. THE SURGEON REVISED THE STEM AND HEAD AND CABLED THE FRACTURE. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Bone Fracture(s)

    MAUDE report 11003731 ↗
  • InjuryOct 13, 2020Loose or Intermittent Connection

    REVISION SURGERY PERFORMED DUE TO STEM LOOSENING, PROXIMALLY, 1 YEAR AND 9 MONTHS AFTER THE PRIMARY SURGERY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 10669526 ↗
  • InjuryAug 11, 2020Fracture

    THE PATIENT CAME IN REPORTING PAIN DUE TO A PERIPROSTHETIC FRACTURE. THE CAUSE OF IT IS UNKNOWN. PRIMARY ON THE (B)(6) 2020. THE SURGEON PLANS TO REVISE THE STEM. NO REVISION HAS BEEN SCHEDULED AT THIS TIME.

    Patient problems: Bone Fracture(s)

    MAUDE report 10389666 ↗
  • InjuryJun 25, 2020Inadequacy of Device Shape and/or Size; Appropriate Device Problem Term/Code Not Available

    THE PATIENT CAME IN REPORTING A LEG LENGTH DISCREPANCY. PRIMARY CASE WAS AMIS AND THERE WERE NO TROUBLE DURING STEM IMPLANTATION. THE SURGEON CHECKED THE LEG LENGTH USING A OVERLAY ON THE C-ARM IMAGE OF A SCOUT FILM. HOWEVER, SOMETIMES HE HAD TO ADJUST THE BED HEIGHT AND THE IMAGES ARE OFF. THE STEM SET PROUD (HIGHER THAN IT SHOULD HAVE BEEN IN RELATION TO THE FEMUR) AFTER THE PRIMARY SURGERY. THE SURGEON REVISED THE STEM AND HEAD 3 MONTHS AFTER PRIMARY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Disorder

    MAUDE report 10193922 ↗
  • InjuryMar 25, 2020Fracture; Appropriate Device Problem Term/Code Not Available

    THE PATIENT HAD HIP PAIN DUE TO FEMUR CALCAR FRACTURE. THE CAUSE OF THE CALCAR FRACTURE IS UNKNOWN. 1 MONTH AFTER PRIMARY SURGERY THE SURGEON REVISED SUCCESSFULLY THE STEM AND THE CERAMIC BALL HEAD.

    Patient problems: Bone Fracture(s)

    MAUDE report 9880867 ↗
  • InjuryMar 6, 2020Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY PERFORMED ONE WEEK AFTER THE PRIMARY SURGERY, DUE TO FEMORAL FRACTURE. THE SURGEON DECIDED TO USE A STAPPING TO CONTAIN THE FRACTURE, NO IMPLANT REVISED. THE SURGERY WAS COMPLETED SUCCESSFULLY. THE REASON FOR THE FRACTURE IS UNKNOWN.

    Patient problems: Bone Fracture(s)

    MAUDE report 9799437 ↗
  • InjuryNov 14, 2019Fracture

    DURING THE FINAL STEM INSERTION THE GREAT TROCHANTER FRACTURED. THE SURGEON DECIDED TO REPLACE THE STEM AND TO USE CERCLAGE. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Bone Fracture(s)

    MAUDE report 9320274 ↗
  • InjuryJun 19, 2019Fracture

    THE PATIENT CAME IN COMPLAINING OF PAIN DUE TO A FRACTURED FEMUR 7 DAYS AFTER PRIMARY. THE FRACTURE OCCURRED WHILE THE PATIENT WAS STANDING ON ONE LEG AND PUTTING THEIR PANTS ON. NO IMPLANTS WERE REVISED. THE SURGEON CABLED THE FRACTURE AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Hip Fracture

    MAUDE report 8711165 ↗
  • InjuryJan 18, 2019Insufficient Device Problem Information

    9 DAYS AFTER THE PRIMARY THE PATIENT PRESENTED A FRACTURE AT THE LEVEL OF THE LESSER TROCHANTER. THE SURGEON REVISED THE STEM AND THE CERAMIC HEAD SUCCESSFULLY.

    Patient problems: Bone Fracture(s)

    MAUDE report 8260595 ↗
  • InjuryDec 19, 2018Difficult to Remove; Positioning Problem

    REVISION SURGERY 1 MONTH AND HALF AFTER THE PRIMARY SURGERY BECAUSE THE STEM WAS TOO DEEP (1 CM), SURGEON WANTED TO REMOVE THE STEM AND PUT A CEMENTED AMISTEM. THE STEM WAS IMPOSSIBLE TO BE REMOVED, SO HE DECIDED TO KEEP THE STEM AND REPLACE THE OTHERS IMPLANTS.

    Patient problems: Patient Problem/Medical Problem

    MAUDE report 8178701 ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices