Skip to content

AMIStem-P 01.18.413Recalls & reported problems

Medacta International SA · AMIStem-P Lat. #3 · Class 2 · Implant

All models of AMIStem-P (18)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
10
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 07630040720120), received between Feb 24, 2020 and Feb 6, 2026 — classified as 10 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Appropriate Device Problem Term/Code Not Available3
  • Device Appears to Trigger Rejection3
  • Loss of Osseointegration1
  • Malposition of Device1
  • Osseointegration Problem1

Patient problems

  • Inadequate Osseointegration3
  • Unspecified Infection3
  • Pain2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Bone Fracture(s)1
  • Insufficient Information1
  • Medical device site infection1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 6, 2026Adverse Event Without Identified Device or Use Problem

    AT ABOUT 2 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN PRESENTING PAIN DUE TO A PERIPROSTHETIC FRACTURE AT THE LESSER TROCHANTER AND THE CAUSE IS UNKNOWN. THE SURGEON CABLED THE FRACTURE AND REVISED THE AMISTEM-P SIZE 3 STEM TO AN AMISTEM-C SIZE 5 STEM, AND REVISED THE HEAD TO A SAME SIZE HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Bone Fracture(s)

    MAUDE report 24276570 ↗
  • InjuryAug 26, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    AFTER 4 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN DUE TO AN INFECTION WITH AN UNKNOWN CAUSE. THE IMPLANTS WERE REMOVED AND REPLACED WITH A COMPETITOR ANTIBIOTIC CEMENT SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Medical device site infection

    MAUDE report 22894259 ↗
  • InjuryOct 4, 2024Osseointegration Problem; Osseointegration Problem

    REVISION SURGERY PERFORMED ABOUT 3 YEARS AND 10 MONTHS AFTER THE PRIMARY SURGERY DUE TO STEM LOOSENING. STEM AND HEAD REVISED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20366840 ↗
  • InjuryJul 22, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    AT ABOUT 1 YEAR AND 3 MONTHS AFTER PRIMARY, THE PATIENT HAS BEEN REVISED BECAUSE OF INFECTION WHICH CAUSED STEM LOOSENING. NO MOBILIZATION OF THE ACETABULAR SHELL HAS BEEN DETECTED. THE SURGEON REVISED SUCCESSFULLY ALL COMPONENTS AND IMPLANTED A M-VIZION STEM WITH VERSAFITCUP DM COMPONENTS AND CERAMIC BALL HEAD.

    Patient problems: Unspecified Infection

    MAUDE report 19790171 ↗
  • InjuryJun 7, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    AT ABOUT 4 DAYS AFTER THE PRIMARY SURGERY, THE SURGEON WANTED TO REVISE THE PATIENT BECAUSE HE WAS NOT SATISFIED BY THE WAY HE INITIALLY IMPLANTED IMPLANTS DURING THE PRIMARY. THE SURGEON REVISED ALL COMPONENTS AND THE SURGERY WAS COMPLETED SUCCESSFULLY. NO ISSUES WERE DETECTED WITH THE IMPLANTS.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 17077538 ↗
  • InjuryFeb 9, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    REVISION SURGERY AT ABOUT 3 YEARS AND 3 MONTHS FROM PRIMARY FOR INFECTION. ALL DEVICES REVISED SUCCESSFULLY AND SPACER IMPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 16332043 ↗
  • InjuryDec 16, 2022Loss of Osseointegration; Adverse Event Without Identified Device or Use Problem; Loss of Osseointegration; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REPORTING PAIN. FROM THE XRAYS, AN EMPTY SPACE BETWEEN THE PROXIMAL EDGE OF THE PROSTHESIS AND THE BONE WAS NOTED (STEM WAS NOT MOBILIZED). THE STEM WAS REVISED ABOUT 1 YEAR AND 1 MONTH AFTER THE PRIMARY SURGERY.

    Patient problems: Inadequate Osseointegration; Pain

    MAUDE report 15996275 ↗
  • MalfunctionDec 23, 2021Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY FOR STEM LOOSENING PERFORMED 1 YEAR AFTER PRIMARY. THE SURGEON REMOVED SUCCESSFULLY THE STEM, THE HEAD, AND THE SLEEVE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 13076103 ↗
  • InjuryNov 7, 2021Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY FOR STEM LOOSENING PERFORMED 1 YEAR AND 3 MONTHS AFTER PRIMARY.

    Patient problems: Insufficient Information

    MAUDE report 12758102 ↗
  • InjuryAug 5, 2020Malposition of Device; Malposition of Device

    THE PATIENT CAME IN 9 DAYS AFTER THE PRIMARY SURGERY REPORTING PAIN DUE TO A FEMORAL FRACTURE DUE TO THE STEM BEING IN VARUS. THE STEM WAS OVERSIZED AND MALPOSITIONED. THE SURGEON REVISED THE STEM AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Pain

    MAUDE report 10366324 ↗
  • InjuryFeb 24, 2020Appropriate Device Problem Term/Code Not Available

    THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON REMOVED ALL IMPLANTS AND IMPLANTED AN ANTIBIOTIC SPACER MORE THAN 2 MONTHS AFTER PRIMARY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Unspecified Infection

    MAUDE report 9744696 ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page