AMIStem-P 01.18.417Recalls & reported problems
Medacta International SA · AMIStem-P Lat. #7 · Class 2 · Implant
All models of AMIStem-P (18)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07630040720168), received between Apr 30, 2019 and Dec 20, 2025 — classified as 4 injuries.
The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Appropriate Device Problem Term/Code Not Available1
- Device Appears to Trigger Rejection1
- Device Dislodged or Dislocated1
Patient problems
- Bone Fracture(s)2
- Bacterial Infection1
- Pain1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryDec 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AT ABOUT 1 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO A BONE BROKEN FEMUR DISTAL TO THE MEDACTA STEM. THE PATIENT DID NOT REPORT ANY TRAUMA. THE SURGEON CABLED THE FRACTURE AND REVISED THE STEM AND HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Bone Fracture(s)
MAUDE report 23858468 ↗InjuryMay 31, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
THE PATIENT HAD A PRIMARY HIP SURGERY ON (B)(6) 2021. ON THE (B)(6) 2021, THE PATIENT CAME WITH SIGNS OF INFECTION AND A WASHOUT WAS PERFORMED. ON THE (B)(6) 2023, THE PATIENT CAME DUE TO SIGNS OF INFECTION AND THE PATHOGEN IS STAPHYLOCOCCUS. ALL THE IMPLANTS HAVE BEEN REVISED AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Bacterial Infection
MAUDE report 17029867 ↗InjuryOct 13, 2020Device Dislodged or Dislocated; Appropriate Device Problem Term/Code Not Available
THE PATIENT CAME IN REPORTING PAIN DUE TO A SUBSIDED STEM AFTER 1 YEAR AND 5 MONTHS FROM THE PRIMARY SURGERY. THE SURGEON REVISED THE STEM, HEAD, AND LINER. DURING THE PRIMARY HIP SURGERY, WHILE THE SURGEON WAS INSERTING THE STEM, THE PATIENT'S GREATER TROCHANTER FRACTURED. THE SURGEON CABLED THE FRACTURE.
Patient problems: Pain
MAUDE report 10668668 ↗InjuryApr 30, 2019Adverse Event Without Identified Device or Use Problem
DURING THE PRIMARY HIP SURGERY AND WHILE INSERTING THE STEM, THE PATIENT'S GREATER TROCHANTER FRACTURED. THE SURGEON CABLED THE FRACTURE AND THERE WAS A 5-MINUTE DELAY IN THE CASE. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Bone Fracture(s)
MAUDE report 8564991 ↗
Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices
- PINNACLE 1217-25-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED735 reports
- NA 1365-52-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED683 reports
- APEX 1246-03-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 deaths613 reports
- PINNACLE 1217-20-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 deaths590 reports
- PINNACLE 1217-30-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 death484 reports
- NA 1365-53-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED404 reports
- PINNACLE ALTRX 1221-36-054Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED388 reports
- PINNACLE ALTRX 1221-32-050Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED386 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page