AMIStem-P Collared 01.18.442Recalls & reported problems
Medacta International SA · AMIStem-P Collared Lat. #2 · Class 2 · Implant
All models of AMIStem-P Collared (18)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07630040720311), received between Nov 16, 2021 and Mar 30, 2026 — classified as 6 injuries.
The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device Appears to Trigger Rejection2
- Osseointegration Problem2
- Loosening of Implant Not Related to Bone-Ingrowth1
- Loss of Osseointegration1
Patient problems
- Inadequate Osseointegration2
- Aseptic loosening1
- Implant Pain1
- Insufficient Information1
- Medical device site infection1
- Unspecified Infection1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryMar 30, 2026Osseointegration Problem
AT ABOUT 1 YEAR AND 4 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN PRESENTING PAIN DUE TO A LOOSE STEM AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED THE AMISTEM-P SIZE 2 STEM TO A SIZE 3 AND REVISED THE 40 MM BIOLOX HEAD TO A SAME SIZE BIOLOX HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY. THERE WAS NO REPORTED TRAUMA.
Patient problems: Aseptic loosening; Inadequate Osseointegration
MAUDE report 24710798 ↗InjuryNov 22, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
ABOUT TWO WEEKS AFTER PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON REVISED ALL MEDACTA HARDWARE AND REIMPLANTED PERMANENT PRODUCT. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 20758450 ↗InjuryOct 11, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
AT 1 MONTH FROM THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON REVISED ALL MEDACTA HARDWARE AND REIMPLANTED PERMANENT PRODUCT. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Unspecified Infection
MAUDE report 20426054 ↗InjuryMay 4, 2023Loss of Osseointegration; Loss of Osseointegration
AT ABOUT 3 YEARS 4 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO A POTTED STEM AND THE CAUSE OF THE POTTED STEM IS UNKNOWN. THE SURGEON REVISED THE STEM, HEAD, AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Inadequate Osseointegration
MAUDE report 16866688 ↗InjuryFeb 1, 2023Osseointegration Problem; Osseointegration Problem
AT ABOUT 2 YEARS 11 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN AND INSTABILITY. X-RAYS SHOWED A PEDESTAL FORMATION AT THE DISTAL TIP OF THE STEM. IN ADDITION, RADIOLUCENCY WAS NOTED AROUND THE PROXIMAL PORTION OF THE STEM. THE SURGEON REVISED THE STEM, HEAD, AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Implant Pain
MAUDE report 16276131 ↗InjuryNov 16, 2021Loosening of Implant Not Related to Bone-Ingrowth
AT 4 MONTHS AFTER PRIMARY SURGERY THE SURGEON REVISED THE PATIENT STEM, HEAD AND LINER FOR STEM LOOSENING.
Patient problems: Insufficient Information
MAUDE report 12820240 ↗
Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices
- PINNACLE 1217-25-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED735 reports
- NA 1365-52-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED683 reports
- APEX 1246-03-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 deaths613 reports
- PINNACLE 1217-20-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 deaths590 reports
- PINNACLE 1217-30-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 death484 reports
- NA 1365-53-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED404 reports
- PINNACLE ALTRX 1221-36-054Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED388 reports
- PINNACLE ALTRX 1221-32-050Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED386 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page