M-Vizion Femoral Revision System 01.22.134Recalls & reported problems
Medacta International SA · Distal Stem Ø15mm L 180mm straight · Class 2 · Implant
All models of M-Vizion Femoral Revision System (30)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07630040728928), received between Sep 12, 2023 and May 30, 2025 — classified as 3 injuries.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- No Apparent Adverse Event1
Patient problems
- Bone Fracture(s)2
- Implant Pain1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryMay 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT CAME IN REPORTING PAIN AND THE CAUSE IS UNKNOWN. IT WAS NOTED THAT THE PROXIMAL BODY WAS RETROVERTED AND THE CAUSE IS UNKNOWN. AT ABOUT 2 MONTHS FROM PRIMARY ALL COMPONENTS WERE EXTRACTED INCLUDING THE NON-MEDACTA CUP/LINER. THE CASE WAS THEN CARRIED OUT WITH SUCCESS WITH A NEW M-VIZION STEM AND PROXIMAL BODY ALONG WITH A MEDACTA BIPOLAR HEAD CONSTRUCT. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Implant Pain
MAUDE report 22114094 ↗InjuryMay 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING THE REVISION SURGERY DUE TO PAIN, PERFORMED ABOUT 1 MONTH AND A HALF AFTER THE PRIMARY SURGERY, IT WAS NOTED THAT THE PROXIMAL BODY WAS RETROVERTED. THE PROXIMAL BODY WOULD NOT DISENGAGE FROM THE STEM. AN ATTEMPT AND MANUAL DISENGAGEMENT RESULTED IN THE ENTIRE CONSTRUCT EXTRACTION. THE CONSTRUCT WAS THEN REINSERTED WITH THE PROPER VERSION. REDUCTION WAS DIFFICULT AND RESULTED IN A NEW DISTAL FEMUR FRACTURE. ALL COMPONENTS WERE EXTRACTED INCLUDING THE NON-MEDACTA CUP/LINER. THE CASE WAS THEN CARRIED OUT WITH SUCCESS WITH A NEW M-VIZION STEM AND PROXIMAL BODY ALONG WITH A MEDACTA BIPOLAR HEAD CONSTRUCT. FRACTURE FIXED. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Bone Fracture(s)
MAUDE report 22043492 ↗InjurySep 12, 2023No Apparent Adverse Event; No Apparent Adverse Event
DURING THE PRIMARY HIP SURGERY AND DURING FINAL TORQUEING OF THE M-VIZION LOCKING SCREW, THE FEMUR FRACTURED DISTALLY. THE SURGEON WAS USING ANTIROTATIONAL HANDLE 01.22.10.0154. PLATE, SCREWS AND CABLES WERE PLACED TO FIX THE FRACTURE AND THE SURGERY WAS COMPLETED SUCCESSFULLY. POOR PATIENT'S BONE QUALITY.
Patient problems: Bone Fracture(s)
MAUDE report 17727779 ↗
Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices
- BIOLOX DELTA 1365-32-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented551 reports
- PINNACLE ALTRX 1221-36-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented481 reports
- BIOLOX DELTA 1365-32-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death314 reports
- BIOLOX DELTA 1365-36-720Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death189 reports
- DJO SURGICAL 400-03-362Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented187 reports
- Anthology 71365705Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented176 reports
- BIOLOX DELTA 1365-40-710Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented170 reports
- PINNACLE GRIPTION 1217-32-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented169 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page