Skip to content

M-Vizion Femoral Revision System 01.22.134Recalls & reported problems

Medacta International SA · Distal Stem Ø15mm L 180mm straight · Class 2 · Implant

All models of M-Vizion Femoral Revision System (30)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07630040728928), received between Sep 12, 2023 and May 30, 2025 — classified as 3 injuries.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • No Apparent Adverse Event1

Patient problems

  • Bone Fracture(s)2
  • Implant Pain1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT CAME IN REPORTING PAIN AND THE CAUSE IS UNKNOWN. IT WAS NOTED THAT THE PROXIMAL BODY WAS RETROVERTED AND THE CAUSE IS UNKNOWN. AT ABOUT 2 MONTHS FROM PRIMARY ALL COMPONENTS WERE EXTRACTED INCLUDING THE NON-MEDACTA CUP/LINER. THE CASE WAS THEN CARRIED OUT WITH SUCCESS WITH A NEW M-VIZION STEM AND PROXIMAL BODY ALONG WITH A MEDACTA BIPOLAR HEAD CONSTRUCT. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Implant Pain

    MAUDE report 22114094 ↗
  • InjuryMay 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING THE REVISION SURGERY DUE TO PAIN, PERFORMED ABOUT 1 MONTH AND A HALF AFTER THE PRIMARY SURGERY, IT WAS NOTED THAT THE PROXIMAL BODY WAS RETROVERTED. THE PROXIMAL BODY WOULD NOT DISENGAGE FROM THE STEM. AN ATTEMPT AND MANUAL DISENGAGEMENT RESULTED IN THE ENTIRE CONSTRUCT EXTRACTION. THE CONSTRUCT WAS THEN REINSERTED WITH THE PROPER VERSION. REDUCTION WAS DIFFICULT AND RESULTED IN A NEW DISTAL FEMUR FRACTURE. ALL COMPONENTS WERE EXTRACTED INCLUDING THE NON-MEDACTA CUP/LINER. THE CASE WAS THEN CARRIED OUT WITH SUCCESS WITH A NEW M-VIZION STEM AND PROXIMAL BODY ALONG WITH A MEDACTA BIPOLAR HEAD CONSTRUCT. FRACTURE FIXED. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Bone Fracture(s)

    MAUDE report 22043492 ↗
  • InjurySep 12, 2023No Apparent Adverse Event; No Apparent Adverse Event

    DURING THE PRIMARY HIP SURGERY AND DURING FINAL TORQUEING OF THE M-VIZION LOCKING SCREW, THE FEMUR FRACTURED DISTALLY. THE SURGEON WAS USING ANTIROTATIONAL HANDLE 01.22.10.0154. PLATE, SCREWS AND CABLES WERE PLACED TO FIX THE FRACTURE AND THE SURGERY WAS COMPLETED SUCCESSFULLY. POOR PATIENT'S BONE QUALITY.

    Patient problems: Bone Fracture(s)

    MAUDE report 17727779 ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices →