Glenoid Reconstruction System 04.01.0304Recalls & reported problems

Medacta International SA · GRS baseplate - Ø24.5x30 +3mm · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07630345710338), received between Oct 15, 2025 and Oct 15, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1
  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Aseptic loosening1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 15, 2025Fracture; Loosening of Implant Not Related to Bone-Ingrowth; Fracture; Loosening of Implant Not Related to Bone-Ingrowth

    ON (B)(6) 2023, THE PATIENT UNDERWENT PRIMARY SURGERY WITH MY SHOULDER PLANNING AND GRS USING A BONE GRAFT TO CORRECT GLENOID RETROVERSION. THE CONSTRUCT WAS STABLE AND THE PATIENT WAS DOING WELL. DURING A FOLLOW-UP VISIT, IT WAS OBSERVED THAT THE GRAFT HAD RESORBED, LEADING TO GLENOID LOOSENING AND THE BREAKAGE OF TWO POLYAXIAL SCREWS. ON (B)(6) 2025, THE IMPLANTS WERE REMOVED AND REPLACED WITH A CEMENT SPACER TO ENABLE CT SCAN PLANNING FOR A CUSTOM-MADE GLENOID IMPLANT.

    Patient problems: Aseptic loosening

    MAUDE report 23290524 ↗