The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07630345710833), received between Nov 4, 2025 and Feb 16, 2026 — classified as 2 injuries.
The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Device Appears to Trigger Rejection1
- Osseointegration Problem1
Patient problems
- Aseptic loosening1
- Inadequate Osseointegration1
- Joint Laxity1
- Medical device site infection1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryFeb 16, 2026Adverse Event Without Identified Device or Use Problem; Osseointegration Problem
AT ABOUT 5 MONTHS FROM THE PRIMARY, DURING APOST-OPERATIVEAPPOINTMENTIT WAS OBSERVED THAT THE LEFT KNEE COMPONENTS HAD SHIFTED DUE TO LOOSENING OF THE TIBIAL TRAY. THE SURGEON CONFIRMED THE KNEE WAS NO LONGER BALANCED AND STABILITY WAS COMPROMISED. THE TIBIAL TRAY AND FEMORAL COMPONENT WERE REVISED FROM A CEMENTLESS TO A CEMENTED SYSTEM, AND THE LINER WAS UPSIZED (FROM 10 MM TO 13 MM) TO IMPROVE STABILITY. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Aseptic loosening; Inadequate Osseointegration; Joint Laxity
MAUDE report 24343890 ↗InjuryNov 4, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
AT ABOUT 1 MONTH AFTER THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON REVISED THE TOTAL KNEE (TRAY, INSERT, FEMUR) AND UTILIZED ANTIBIOTIC CEMENT WITH A GMK-SPHERIKA KNEE. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 23461672 ↗
Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗