Skip to content

GMK Total Knee System 02.12.KA06RRecalls & reported problems

Medacta International SA · Femoral Component SPHERIKA Cemented S6R · Class 2 · Implant

All models of GMK Total Knee System (31)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07630345716149), received between Oct 17, 2025 and Mar 31, 2026 — classified as 2 injuries.

The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Device Appears to Trigger Rejection1

Patient problems

  • Implant Pain1
  • Medical device site infection1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 31, 2026Device Appears to Trigger Rejection

    THE PATIENT HAD A KNEE INFECTION AND THE PATHOGEN IS UNKNOWN. AT ABOUT 5 MONTHS POST PRIMARY THE SURGEON PERFORMED A WASHOUT AND REVISED THE GMK-SPHERIKA FEMORAL COMPONENT, 10 MM INSERT, AND S5 TIBIAL TRAY TO A GMK-REVISION FEMUR, TIBIAL TRAY AND 10 MM INSERT. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Medical device site infection

    MAUDE report 24738808 ↗
  • InjuryOct 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT UNDERWENT TKA ON (B)(6) 2024. ON (B)(6) 2024 AT FOLLOWUP THE PATIENT REPORTED ANTERIOR KNEE PAIN. TEST FOR INFECTION WERE NEGATIVE. SURGEON CONCLUSION WAS THAT THE PAIN WAS DUE TO PATELLA REMODELING AND THAT IT WOULD NATURALLY IMPROVED OVER TIME. ON (B)(6) 2024 PATIENT STILL REPORTED PAIN. SURGEON SCHEDULED PATELLA RESURFACING FOR (B)(6) 2025 BUT PATIENT CANCELLED THE PROCEDURE ON (B)(6) AND COULD NOT BE CONTACTED ANYMORE. IT IS UNKNOWN IF THE PATIENT UNDERWENT ADDITIONAL OR REVISION SURGERY OR ANY OTHER TREATMENT.

    Patient problems: Implant Pain

    MAUDE report 23321317 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page