Skip to content

NextAR Knee Platform 15.8010Recalls & reported problems

Medacta International SA · MyKnee NextAR TibPinPositioner_Ass-CT-GMK-LM · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07630345731647), received between Aug 7, 2025 and Aug 7, 2025 — classified as 1 malfunction.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 7, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    DURING TKA, THE TIBIAL INSERT TRIAL COULD NOT BE PROPERLY PLACED INTO THE INTENDED POSITION. UPON REVIEW BY THE SURGEON, IT WAS DETERMINED THAT THE CAUSE WAS MISALIGNMENT BETWEEN THE KEEL TRIAL AND THE TIBIAL TRAY TRIAL; THE KEEL TRIAL WAS NOT SEATED CORRECTLY. SEVERAL ATTEMPTS WERE MADE TO IMPACT THE KEEL TRIAL AND INSERT THE TIBIAL INSERT TRIAL, BUT THE ISSUE COULD NOT BE RESOLVED. AFTER REMOVING THE KEEL TRIAL AND VISUALLY INSPECTING THE INTRAMEDULLARY CANAL, IT WAS FOUND THAT THE NEXTAR TARGET HOLDER FIXATION PIN WAS INTERFERING WITH THE KEEL TRIAL. THE DEPTH OF THE FIXATION PIN WAS MANUALLY ADJUSTED UNTIL THE INTERFERENCE WAS ELIMINATED. THE SURGEYR WAS COMPLETED SUCCESSFULLY. THE TOTAL SURGERY TIME WAS 1 HOUR AND 45 MINUTES, AND THE DELAY TIME WAS 30 MINUTES.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 22737154 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page