Shoulder System 04.01.0401Recalls & reported problems
Medacta International SA · GRS baseplate Full wedge 15 · Class 2 · Implant
All models of Shoulder System (63)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07630542714634), received between Aug 29, 2025 and Jul 14, 2026 — classified as 4 injuries.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Loosening of Implant Not Related to Bone-Ingrowth2
- Adverse Event Without Identified Device or Use Problem1
- Migration1
- Osseointegration Problem1
Patient problems
- Aseptic loosening3
- Implant Pain1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJul 14, 2026Loosening of Implant Not Related to Bone-Ingrowth
REVISION SURGERY AFTER 3 MONTHS FROM THE PRIMARY DUE TO A LOOSE GLENOID CAUSED BY A SUSPECTED INFECTION. THE SURGEON PERFORMED A WASHOUT AND REVISED THE PATIENT TO A HEMI. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Aseptic loosening
MAUDE report 25813731 ↗InjuryMay 28, 2026Adverse Event Without Identified Device or Use Problem
THE PATIENT CAME IN PRESENTING PAIN DUE TO A BASEPLATE THAT WAS IMPLANTED TOO SUPERIOR ON THE GLENOID FACE DURING THE PRIMARY SURGERY. ABOUT 5 MONTHS AND A HALF AFTER THE PRIMARY SURGERY, THE SURGEON REVISED THE BASEPLATE AND GLENOSPHERE TO A COMPETITOR BASEPLATE AND GLENOSPHERE, AND THE LINER TO A REV. CONSTRAINED E-CROSS LINER D 36/+0. THE MEDACTA STEM REMAINED AS IT WAS WELL FIXED. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Implant Pain
MAUDE report 25313932 ↗InjuryJan 23, 2026Loosening of Implant Not Related to Bone-Ingrowth; Migration
DURING A POST-OPERATION FOLLOW UP AT 1 YEAR FROM PRIMARY, THE SURGEON DISCOVERED THAT THE BASEPLATE HAD MIGRATED AFTER THE PATIENT`S PRIMARY LEFT SHOULDER SURGERY AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED THE BASEPLATE TO A THREADED BASEPLATE, THE GLENOSPHERE AND METAPHYSIS TO A SAME SIZE GLENOSPHERE AND METAPHYSIS, AND REVISED THE LINER TO A REVERSE CONSTRAINED LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Aseptic loosening
MAUDE report 24162769 ↗InjuryAug 29, 2025Osseointegration Problem; Osseointegration Problem
THE PATIENT UNDERWENT PRIMARY SURGERY ON (B)(6) 2025. ON (B)(6) 2025, THE PATIENT PRESENTED WITH PAIN DUE TO A FAILED SUBSCAPULARIS MUSCLE, THE CAUSE OF WHICH WAS UNKNOWN. THE SURGEON REVISED THE IMPLANT FROM AN ANATOMIC SHOULDER TO A REVERSE SHOULDER. THE SURGERY WAS COMPLETED SUCCESSFULLY. ON (B)(6) 2025, THE PATIENT AGAIN PRESENTED WITH PAIN, THIS TIME DUE TO A LOOSE BASEPLATE, WITH THE CAUSE REMAINING UNKNOWN. THE SURGEON REVISED THE REVERSE SHOULDER TO A HEMI SHOULDER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Aseptic loosening
MAUDE report 22922921 ↗
Other Shoulder Prosthesis, Reverse Configuration devices
- DJO SURGICAL 508-36-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration508 reports
- DJO SURGICAL 508-32-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration433 reports
- DJO SURGICAL 508-32-103Encore Medical, L.P. · shoulder prosthesis, reverse configuration421 reports
- DJO SURGICAL 509-00-036Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall259 reports
- DJO SURGICAL 508-32-104Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall203 reports
- DJO SURGICAL 509-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration233 reports
- DELTA XTEND 130760000Depuy (Ireland) · shoulder prosthesis, reverse configuration223 reports
- DJO SURGICAL 508-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration122 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page