Skip to content

DM Converter – TiN coated 01.32.3641CFRecalls & reported problems

Medacta International SA · DM Converter D/DMB - TiN coated · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07630971256347), received between Jun 27, 2024 and Jun 27, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 27, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE PATIENT HAD A PRIMARY WITH MPACT NO HOLE CUP, MPACT DM CONVERTER AND DOUBLE MOBILITY LINER IMPLANTED ON THE (B)(6) 2024. THE CUP WAS PLACED AS USUAL, 40 INCLINATION AND ROUGHLY 20 ANTEVERSION. THE PATIENT SHOWED SIGNS OF A SPINAL DEFORMITY. THE PATIENT DISLOCATED POSTERIORLY 3 TIMES OVER IN 2 WEEKS. IT WAS FOUND THAT THE PATIENT HAD AN EXTREMELY HYPER MOBILE HIP (EXTREMELY LAX IN HER LIGAMENTS) AND HER HIP POSTERIOR CAPSULE WAS DETACHED. THE SURGEON DECIDED TO REMOVE THE MPACT NO HOLE CUP WITH DM CONVERTER LINER, THE DOUBLE MOBILITY LINER AND PUT IN A HIGHLY CONSTRAINED CUP (OF COMPETITOR). THIS WAS DONE ON THE (B)(6) 2024. THE SURGEON REPORTED THAT HE DOES NOT BELIEVE THIS WAS AN IMPLANT FAULT, THE PATIENT JUST REQUIRED A MORE CONSTRAINT HIP CONFIGURATION.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 19623329 ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices →

Sources and updates