Skip to content

GMK Sphere Total Knee System 02.12.E0413FLRecalls & reported problems

Medacta International SA · Tibial Insert Fixed SPHERE FLEX #4/13 mm L E-Cross · Class 2 · Implant

All models of GMK Sphere Total Knee System (209)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07630971261877), received between Feb 2, 2022 and Jul 15, 2026 — classified as 9 injuries.

The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Appears to Trigger Rejection6
  • Adverse Event Without Identified Device or Use Problem1
  • Device Dislodged or Dislocated1
  • Unstable1

Patient problems

  • Medical device site infection4
  • Unspecified Infection2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Joint Dislocation1
  • Joint Laxity1
  • Loss of Range of Motion1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 15, 2026Device Appears to Trigger Rejection

    AT 1 MONTH FROM THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE GMK-SPHERE TIBIAL INSERT E-CROSS - FLEX 4L - 13MM TO A SAME SIZE INSERT. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Medical device site infection

    MAUDE report 25822783 ↗
  • InjuryMay 12, 2026Device Dislodged or Dislocated

    AT 3 YEARS 1 MONTH FROM PRIMARY, THE PATIENT CAME IN PRESENTING PAIN DUE TO A DISENGAGEMENT OF THE LINER FROM THE TIBIAL TRAY. THERE WAS NO TRAUMA INDICATED AND THE SURGEON DID NOT HAVE ANY ISSUES SNAPPING THE POLY IN DURING THE PRIMARY SURGERY. THE SURGEON PERFORMED A WASHOUT AND REVISED THE INSERT WITH ONE OF THE SAME SIZE. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Dislocation

    MAUDE report 25148977 ↗
  • InjuryApr 14, 2026Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY DUE TO KNEE STIFFNESS PATIENT CAME IN PRESENTING STIFFNESS ABOUT 1 YEAR AND 4 MONTHS AFTER PRIMARY SURGERY. THE SURGEON REVISED 13MM INSERT TO A 20MM INSERT, REVISED THE FEMORAL COMPONENT TO A TINBN COATED FEMORAL COMPONENT, AND REVISED THE TIBIAL TRAY TO A TINBN COATED TIBIAL TRAY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Loss of Range of Motion

    MAUDE report 24871666 ↗
  • InjuryNov 10, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    ON (B)(6) 2025, THE PATIENT HAD PRIMARY LEFT KNEE SURGERY. ON (B)(6) 2025, THE PATIENT CAME IN PRESENTING AN INFECTION AND THE PATHOGEN WAS UNKNOWN. THE SURGEON REVISED THE TRAY, INSERT, AND FEMUR. THE SURGERY WAS SUCCESSFULLY COMPLETED (B)(4). PRESENTLY, ON (B)(6) 2025, THE PATIENT CAME IN PRESENTING AN INFECTION AND THE CAUSE IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND POLY SWAP. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Medical device site infection

    MAUDE report 23510670 ↗
  • InjuryNov 5, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    AT ABOUT 2 WEEKS AFTER THE PRIMARY, THE PATIENT CAME IN DUE TO AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND POLY SWAP. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Medical device site infection

    MAUDE report 23477200 ↗
  • InjurySep 12, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    AT ABOUT 1 YEAR AND 10 MONTHS THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT, REMOVED ALL IMPLANTS, AND IMPLANTED AN ANTIBIOTIC SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Medical device site infection

    MAUDE report 23036822 ↗
  • InjuryFeb 23, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    THE PATIENT HAD A PRIMARY KNEE SURGERY ON (B)(6) 2023. ON 23 OCTOBER 2023, THE PATIENT CAME IN REPORTING INSTABILITY AND THE CAUSE OF THE INSTABILITY IS UNKNOWN. THE SURGEON REVISED THE INSERT. THE SURGERY WAS COMPLETED SUCCESSFULLY. PRESENTLY, ON (B)(6) 2024, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Unspecified Infection

    MAUDE report 18765735 ↗
  • InjuryDec 20, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    AT ABOUT 2 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Unspecified Infection

    MAUDE report 18364385 ↗
  • InjuryFeb 2, 2022Unstable

    AT 20 DAYS AFTER THE PRIMARY, THE PATIENT CAME IN REPORTING INSTABILITY AND THE CAUSE OF THE INSTABILITY IS UNKNOWN. THE SURGEON REVISED THE INSERT (13 MM TO 14 MM) AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Laxity

    MAUDE report 13423111 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page