M-Vizion Femoral Revision System 01.22.408Recalls & reported problems
Medacta International SA · Proximal Body Ø20mm L 50mm LAT with holes · Class 2 · Implant
All models of M-Vizion Femoral Revision System (30)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07630971263574), received between Mar 27, 2025 and Jul 31, 2025 — classified as 3 injuries.
The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Appears to Trigger Rejection1
- Inadequacy of Device Shape and/or Size1
- Positioning Problem1
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Medical device site infection1
- Pain1
- Unequal Limb Length1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJul 31, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
AT ABOUT 2 YEARS AND 10 MONTHS FROM PRIMARY SURGERY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION, AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT, REMOVED ALL IMPLANTS, AND IMPLANTED AN ANTIBIOTIC SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 22674986 ↗InjuryJul 29, 2025Positioning Problem; Positioning Problem
AT ABOUT 1 WEEK FROM PRIMARY THE PATIENT UNDERWENT REVISION SURGERY DUE TO A MALPOSITIONING OF THE DISTAL STEM (FALSA VIA). THE ENTIRE CONSTRUCT HAS BEEN REVISED. THE TWO PARTS COULD NOT BE SEPARATED AS THE TIP OF THE SCREWDRIVER USING DURING PRIMARY BROKE INSIDE THE LOCKING SCREW. DURING THE TIGHTENING OF THE LOCKING SCREW THE SCREW DRIVER TIP BROKE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 22643026 ↗InjuryMar 27, 2025Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
M-VIZION STEM WAS IMPLANTED ON (B)(6) 2025. THE PATIENT NEEDED TO BE REVISED BECAUSE WAS EXPERIENCING PAIN, INSTABILITY AND LEG LENGTH DISCREPANCY, DUE TO A TOO SHORT PROXIMAL BODY IMPLANTED AT THE PRIMARY SURGERY. THE REVISION SURGERY TOOK PLACE 1 MONTH AFTER THE PRIMARY SURGERY AND THE PLAN WAS TO REVISE ONLY THE PROXIMAL BODY. HOWEVER, THE DISASSEMBLY ROD BENT DURING PROXIMAL STEM EXTRACTION FROM THE DISTAL BODY. DESPITE THE ATTEMPTS, THE PROXIMAL BODY COULD NOT BE REMOVED FROM THE DISTAL STEM, THEREFORE THE SURGEON DECIDED TO REMOVE THE ENTIRE STEM (PROXIMAL + DISTAL). THE REVISION SURGERY WAS ALSO TREATED WITH AN M-VIZION. HERE, THE SURGEON CHOSE A 16MM 180MM DISTAL STEM WITH A LONGER PROXIMAL BODY 20MM 70MM TO COMPENSATE FOR THE INSTABILITY.
Patient problems: Pain; Unequal Limb Length
MAUDE report 21708063 ↗
Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices
- BIOLOX DELTA 1365-32-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented551 reports
- PINNACLE ALTRX 1221-36-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented481 reports
- BIOLOX DELTA 1365-32-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death314 reports
- BIOLOX DELTA 1365-36-720Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death189 reports
- DJO SURGICAL 400-03-362Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented187 reports
- Anthology 71365705Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented176 reports
- BIOLOX DELTA 1365-40-710Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented170 reports
- PINNACLE GRIPTION 1217-32-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented169 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗