Passeo-35 6/40/130Recalls & reported problems
Biotronik AG · Class 2
All models of Passeo-35 (24)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07640119558400), received between May 12, 2022 and May 24, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture2
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- Pain1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 24, 2024Material Rupture; Material Rupture
A PASSEO-35 PERIPHERAL BALLOON CATHETER WAS SELECTED FOR TREATMENT OF A SEVERELY CALCIFIED LESION (STENOSIS DEGREE: 90 PERCENT) IN THE MILDLY TORTUOUS MID LEFT UPPER LIMB VEIN. DURING EXPANSION AND WHEN THE PRESSURE REACHED 12 KPA, THE PASSEO-35 BALLOON RUPTURED BUT DID NOT REACH THE MAXIMUM PRESSURE FOR THAT CATHETER (14 KPA). THE PATIENT REPORTED SIGNIFICANT PAIN IN THE LEFT FOREARM, AND THE BALLOON WAS REPLACED TO CONTINUE THE PROCEDURE.
Patient problems: Pain
MAUDE report 19393845 ↗MalfunctionMay 12, 2022Material Rupture; Material Rupture
A PASSEO-35 BALLOON CATHETER WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION (55 PERCENT STENOSIS DEGREE) IN THE MODERATELY TORTUOUS SFA. WHEN PRESSURE WAS APPLIED UP TO 6 BAR THE BALLOON RUPTURED. ANOTHER PASSEO-35 WAS USED SUCCESSFULLY. THE PATIENT IS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14380040 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page