Pantera LEO 3.5/8/145Recalls & reported problems
Biotronik AG · Class 2
All models of Pantera LEO (30)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07640130415553), received between Aug 13, 2019 and Aug 19, 2021 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Rupture3
- Difficult to Remove2
- Failure to Advance1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 19, 2021Material Rupture; Material Rupture
AFTER PRE-DILATATION OF A MODERATELY CALCIFIED LESION (90 PERCENT STENOSIS DEGREE) IN THE MID RCA WITH A SEMI-COMPLIANT BALLOON, A STENT WAS DEPLOYED. SINCE THE STENT WAS INSUFFICIENTLY EXPANDED, THE PANTERA LEO WAS SELECTED FOR POST-DILATATION BUT RUPTURED DURING INFLATION AT 20 ATM. THE DEVICE WAS RETRIEVED. SUBSEQUENTLY, ANOTHER NON-COMPLIANT BALLOON WAS USED FOR ADDITIONAL POST-DILATATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12342388 ↗MalfunctionJun 10, 2021Difficult to Remove; Difficult to Remove
A PANTERA LEO CORONARY BALLOON CATHETER WAS SELECTED FOR TREATMENT. DIFFICULTIES WERE EXPERIENCED IN WITHDRAWING THE PANTERA LEO OVER A GUIDE WIRE. BOTH DEVICES HAD TO BE REMOVED TOGETHER. ORIGINALLY REPORTED ON MDR-1028232-2021-00879, UNDER INCORRECT LOT NUMBER. CLOSED THAT REPORT AS BEING OPENED IN ERROR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11978100 ↗MalfunctionMar 31, 2021Material Rupture; Material Rupture
A PANTERA LEO CORONARY BALLOON CATHETER WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION (75 PERCENT STENOSIS DEGREE) IN THE RCA. THE BALLOON RUPTURED DURING INFLATION AT 18 ATM. ANOTHER BALLOON CATHETER WAS USED FOR THE INTERVENTION. IT WAS STATED THAT IT OCCURRED DUE TO CALCIFICATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11595549 ↗MalfunctionFeb 19, 2021Difficult to Remove; Difficult to Remove
A PANTERA LEO CORONARY BALLOON CATHETER WAS SELECTED FOR TREATMENT. DIFFICULTIES WERE EXPERIENCED IN WITHDRAWING THE PANTERA LEO OVER A GUIDE WIRE. BOTH DEVICES HAD TO BE REMOVED TOGETHER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11350890 ↗MalfunctionOct 23, 2019Material Rupture; Material Rupture
THE PANTERA LEO BALLOON CATHETER WAS SELECTED FOR USE IN THE RCA. DURING INFLATION THE BALLOON RUPTURED AT 14 ATM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9225325 ↗MalfunctionAug 13, 2019Failure to Advance; Failure to Advance
A PANTERA LEO BALLOON WAS CHOSEN TO TREAT A SEVERE CALCIFIED LESION (83 PERCENT STENOSIS DEGREE) IN A SEVERE TORTUOUS LAD. DESPITE REPEATED ATTEMPTS IT WAS NOT POSSIBLE TO CROSS THE LESION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8889861 ↗
Other Catheters, transluminal coronary angioplasty, percutaneous devices
- MINI TREK™ 1012270-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous223 reports
- Maverick²™ H7493892815200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous1 death203 reports
- MINI TREK™ 1012270-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous162 reports
- Maverick²™ H7493892820200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous161 reports
- NC TREK 1012454-08Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 recalls59 reports
- TREK™ 1012272-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 deaths158 reports
- NC TREK 1012453-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall84 reports
- NC TREK 1012453-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall1 death83 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page