Skip to content

Pantera LEO 3.0/12/145Recalls & reported problems

Biotronik AG · Class 2

All models of Pantera LEO (30)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07640130415645), received between Apr 23, 2019 and May 13, 2020 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Rupture4
  • Leak/Splash1

Patient problems

  • No Consequences Or Impact To Patient4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 13, 2020Leak/Splash; Material Rupture; Leak/Splash; Material Rupture

    A PANTERA LEO BALLOON SYSTEM WAS CHOSEN FOR TREATMENT OF A SEVERELY CALCIFIED LESION (98 PERCENT STENOSIS DEGREE) IN THE LAD. DURING INFLATION THE BALLOON BURST AT 14 ATM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10051167 ↗
  • MalfunctionApr 2, 2020Material Rupture; Material Rupture

    THE PANTERA LEO BALLOON CATHETER WAS SELECTED FOR PRE-DILATATION OF A CALCIFIED LESION IN THE LAD. DURING THE SECOND INFLATION THE BALLOON RUPTURED AT TWENTY ATM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9915511 ↗
  • MalfunctionSep 10, 2019Material Rupture; Material Rupture

    THE PANTERA LEO BALLOON CATHETER WAS SELECTED FOR PRE-DILATATION. DURING THE THIRD INFLATION, THE BALLOON RUPTURED AT 20 BAR. THEREFORE THE DEVICE WAS REMOVED AND ANOTHER BALLOON CATHETER WAS USED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8982316 ↗
  • MalfunctionApr 23, 2019Material Rupture; Material Rupture

    OUS MDR - THE PANTERA LEO BALLOON CATHETER WAS SELECTED FOR PRE-DILATATION. DURING INFLATION THE BALLOON RUPTURED AT 14 ATM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8540414 ↗