Orsiro 3.5/13/140Recalls & reported problems
Biotronik AG · Class 3 · Implant
All models of Orsiro (49)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07640130438941), received between Apr 26, 2019 and Jun 1, 2021 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated5
- Failure to Advance4
- Patient Device Interaction Problem1
Patient problems
- No Consequences Or Impact To Patient6
- No Clinical Signs, Symptoms or Conditions2
- Hematoma1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 1, 2021Failure to Advance; Device Dislodged or Dislocated; Failure to Advance; Device Dislodged or Dislocated
AN ORSIRO DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A SEVERELY CALCIFIED LESION (90 PERCENT STENOSIS DEGREE) IN THE MILDLY TORTUOUS OSTIAL CX. THE DEVICE COULD NOT PASS THE HEAVILY CALCIFIED LESION EVEN WITH THE USE OF A GUIDELINER (TELEFLEX 6F). DURING ATTEMPT TO REMOVE THE DEVICE, THE STENT DISLODGED. THE STENT COULD NOT BE SNARED AND WAS ADAPTED TO THE VESSEL WALL WITH ANOTHER STENT. IT WAS STATED THAT THE STENT MIGHT GOT CAUGHT IN A PREVIOUS IMPLANTED STENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11917291 ↗InjuryFeb 19, 2021Patient Device Interaction Problem; Patient Device Interaction Problem
(B)(6) STUDY: AN ORSIRO DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT. AFTER SUCCESSFUL STENT DEPLOYMENT, THE PATIENT DEVELOPED AN INTRAMURA HEMATOMA PROXIMALLY. THIS REQUIRED AN ADDITIONAL STENT TO BE IMPLANTED PROXIMAL TO FIRST STENT.
Patient problems: Hematoma
MAUDE report 11348508 ↗MalfunctionOct 9, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
AN ORSIRO DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A SEVERELY CALCIFIED LESION IN A MODERATELY TORTUOUS MID CX. DURING DEVICE PREPARATION NEGATIVE PRESSURE WAS SUPPLIED. UPON DELIVERY, STENT BECAME DISLODGED ON CALCIUM SHELF AND MOVED DOWNSTREAM. STENT WAS SNARED OUT NEAR SFA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10658311 ↗MalfunctionFeb 24, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
AN ORSIRO DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF SEVERELY CALCIFIED LESION (99 PERCENT STENOSIS DEGREE) IN A MODERATELY TORTUOUS PROXIMAL LAD. DURING ATTEMPT OF LESION CROSSING, THE STENT WAS DISLODGED IN THE BODY. STENT WAS REMOVED ON (B)(6) 2020.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9747107 ↗MalfunctionSep 10, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE ORSIRO US DRUG ELUTING STENT SYSTEM WAS SELECTED FOR USE. UPON INSPECTION OF THE DELIVERY SYSTEM FOR REINSERTION IT WAS DETECTED THAT THE STENT WAS DISPLACED DISTALLY. THE STENT WAS INITIALLY DELIVERED THROUGH A GUIDELINER.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8982333 ↗MalfunctionAug 28, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated
AN ORSIRO DRUG-ELUTING STENT SYSTEM WAS CHOSEN FOR TREATMENT. DURING PREPARATION WHILE REMOVING THE PROTECTION CAP THE STENT DISLODGED AND REMAINED IN THE CAP.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8941823 ↗MalfunctionAug 21, 2019Failure to Advance; Failure to Advance
THE ORSIRO DRUG-ELUTING STENT SYSTEM WAS CHOSEN FOR THE TREATMENT OF A CALCIFIED AND TORTUOUS LESION IN THE PROXIMAL RCA. AFTER PRE-DILATATION THE TARGET LESION COULD NOT BE CROSSED AND THEREFORE THE ORSIRO WAS REMOVED FROM THE PATIENTS BODY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8915888 ↗MalfunctionMay 16, 2019Failure to Advance; Failure to Advance
THE ORSIRO DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR USE. DURING LESION CROSSING THE STENT COULD NOT PASS THE LESION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8617829 ↗MalfunctionApr 26, 2019Failure to Advance; Failure to Advance
AN ORSIRO CORONARY DRUG-ELUTING STENT SYSTEM WAS SELECTED TO TREAT A MODERATE CALCIFIED LESION (80% STENOSIS DEGREE) IN A CORONARY BYPASS GRAFT. DESPITE PREDILATATION THE DEVICE COULD NOT CROSS THE LESION. AFTER FURTHER DILATATION THE DEVICE COULD BE DELIVERED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8558378 ↗
Other Coronary drug-eluting stent devices
- SYNERGY™ H7493926038300Boston Scientific Corporation · Coronary drug-eluting stent8 deaths124 reports
- SYNERGY™ H7493926016300Boston Scientific Corporation · Coronary drug-eluting stent7 deaths106 reports
- SYNERGY™ H7493926020300Boston Scientific Corporation · Coronary drug-eluting stent9 deaths90 reports
- SYNERGY™ H7493926016250Boston Scientific Corporation · Coronary drug-eluting stent6 deaths82 reports
- XIENCE Sierra™ 1550350-33Abbott Vascular Inc. · Coronary drug-eluting stent1 recall32 reports
- Resolute Onyx™ RONYX30022UXMedtronic, Inc. · Coronary drug-eluting stent6 deaths76 reports
- SYNERGY™ H7493926024300Boston Scientific Corporation · Coronary drug-eluting stent3 deaths75 reports
- Resolute Onyx™ RONYX30015UXMedtronic, Inc. · Coronary drug-eluting stent3 deaths73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page