Orsiro 4.0/15/140Recalls & reported problems
Biotronik AG · Class 3 · Implant
All models of Orsiro (49)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07640130439016), received between Jun 28, 2019 and Mar 29, 2022 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated2
- Difficult to Remove1
- Entrapment of Device1
- Failure to Advance1
Patient problems
- No Consequences Or Impact To Patient4
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 29, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
AN ORSIRO DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY TO SEVERELY CALCIFIED LESION IN THE MODERATELY TORTUOUS DISTAL LAD. AFTER PREDILATATION THE LESION COULD NOT BE CROSSED WITH THE DEVICE. IN REMOVING THE ORSIRO, UPON ENTERING THE GUIDING CATHETER, THE STENT DISLODGED AND REMAINED IN THE PROXIMAL LAD. A DILATATION BALLOON WAS INSERTED AND INFLATED TO CRUSH THE STENT INTO THE VESSEL WALL. THE STENT DELIVERY BALLOON WAS REMOVED WITHOUT FURTHER INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13943883 ↗MalfunctionOct 30, 2019Difficult to Remove; Difficult to Remove
AN ORSIRO DRUG-ELUTING STENT SYSTEM WAS CHOSEN FOR TREATMENT. THE STENT WAS SUCCESSFULLY DEPLOYED. DURING REMOVAL OF THE DEVICE RESISTANCE WAS NOTED. THE LOT NUMBER OF THIS DEVICE IS UNKNOWN.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9257902 ↗MalfunctionSep 16, 2019Entrapment of Device; Entrapment of Device
(B)(6) YEAR OLD FEMALE. THE ORSIRO DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR USE. WHILE THE ORSIRO WAS ADVANCED INTO THE GUIDING CATHETER RESISTANCE WAS FELT AND IT WAS ATTEMPTED TO REMOVE THE ORSIRO. THEREBY THE STENT DISLODGED FROM THE BALLOON AND WAS SUBSEQUENTLY FOUND INSIDE THE CATHETER. EVENTUALLY IT WAS REMOVED FROM THE PATIENTS BODY WITH MINIMAL EFFORT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9071221 ↗MalfunctionJul 15, 2019Failure to Advance; Failure to Advance
AN ORSIRO DRUG-ELUTING STENT SYSTEM WAS CHOSEN TO TREAT A LESION (75 PERCENT STENOSIS DEGREE) IN A TORTUOUS LEFT MAIN. THE DEVICE COULD NOT CROSS THE LESION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8791648 ↗MalfunctionJun 28, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated
AN ORSIRO DRUG-ELUTING STENT SYSTEM WAS CHOSEN TO TREAT A SEVERELY CALCIFIED LESION (85% STENOSIS DEGREE) IN A TORTUOUS RCA. THE STENT CAME OFF THE BALLOON SOMETIME DURING PROCEDURE AND WAS ADAPTED TO THE VESSEL WALL. A GUIDEZILLA 6F EXTENSION CATHETER WAS INVOLVED DURING PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8746012 ↗
Other Coronary drug-eluting stent devices
- SYNERGY™ H7493926038300Boston Scientific Corporation · Coronary drug-eluting stent8 deaths124 reports
- SYNERGY™ H7493926016300Boston Scientific Corporation · Coronary drug-eluting stent7 deaths106 reports
- SYNERGY™ H7493926020300Boston Scientific Corporation · Coronary drug-eluting stent9 deaths90 reports
- XIENCE Sierra™ 1550350-33Abbott Vascular Inc. · Coronary drug-eluting stent1 recall32 reports
- SYNERGY™ H7493926016250Boston Scientific Corporation · Coronary drug-eluting stent6 deaths82 reports
- Resolute Onyx™ RONYX30022UXMedtronic, Inc. · Coronary drug-eluting stent6 deaths76 reports
- SYNERGY™ H7493926024300Boston Scientific Corporation · Coronary drug-eluting stent3 deaths75 reports
- Resolute Onyx™ RONYX30015UXMedtronic, Inc. · Coronary drug-eluting stent3 deaths73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page