Orsiro 3.0/35/140Recalls & reported problems
Biotronik AG · Class 3 · Implant
All models of Orsiro (49)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07640130439382), received between Aug 21, 2019 and Nov 10, 2021 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Failure to Advance” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Failure to Advance1
- Inflation Problem1
- Material Deformation1
- Patient Device Interaction Problem1
- Use of Device Problem1
Patient problems
- No Consequences Or Impact To Patient2
- Cardiac Arrest1
- No Clinical Signs, Symptoms or Conditions1
- Thrombosis/Thrombus1
- Vascular Dissection1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryNov 10, 2021Patient Device Interaction Problem; Patient Device Interaction Problem
7 DAYS AFTER THE ORSIRO DRUG-ELUTING STENT IMPLANTATION IN THE PROX. LAD, THE PATIENT RETURNED TO THE HOSPITAL WITH A CARDIAC ARREST. A 100 PERCENT SUBACUTE STENT THROMBOSIS WAS DETECTED AND TREATED WITH ASPIRATION THROMBECTOMY AND A BALLOON DILATION INSIDE THE STENT.
Patient problems: Cardiac Arrest; Thrombosis/Thrombus
MAUDE report 12787400 ↗MalfunctionMar 29, 2021Inflation Problem; Inflation Problem
AN ORSIRO DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT. IT WAS NOT POSSIBLE TO DEPLOY THE STENT TO TARGET LESION. THE BALLOON WAS INFLATED BUT THE STENT DID NOT DEPLOY. THE INFLATION PRESSURE IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11581910 ↗InjuryDec 12, 2020Use of Device Problem; Use of Device Problem
(B)(6) STUDY: THE ORSIRO DRUG ELUTING STENT SYSTEM WAS CHOSEN FOR THE TREATMENT. AFTER DEPLOYMENT A DISSECTION OCCURRED AND THEN IT WAS DECIDED TO USE ANOTHER ORSIRO TO COVER THE DISSECTION SUCCESSFULLY.
Patient problems: Vascular Dissection
MAUDE report 10996502 ↗MalfunctionMar 31, 2020Material Deformation; Material Deformation
AN ORSIRO DRUG-ELUTING STENT SYSTEM WAS CHOSEN FOR TREATMENT OF A SVG. THE DEVICE COULD NOT BE ADVANCED THROUGH A 6F GUIDELINER AND WAS REMOVED. AFTER REMOVAL IT WAS DETECTED THAT THE STENT WAS DEFORMED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9908661 ↗MalfunctionAug 21, 2019Failure to Advance; Failure to Advance
THE ORSIRO DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR USE. AFTER PRE-DILATATION THE CALCIFIED AND TORTUOUS TARGET LESION COULD NOT BE CROSSED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8915697 ↗
Other Coronary drug-eluting stent devices
- SYNERGY™ H7493926038300Boston Scientific Corporation · Coronary drug-eluting stent8 deaths124 reports
- SYNERGY™ H7493926016300Boston Scientific Corporation · Coronary drug-eluting stent7 deaths106 reports
- SYNERGY™ H7493926020300Boston Scientific Corporation · Coronary drug-eluting stent9 deaths90 reports
- XIENCE Sierra™ 1550350-33Abbott Vascular Inc. · Coronary drug-eluting stent1 recall32 reports
- SYNERGY™ H7493926016250Boston Scientific Corporation · Coronary drug-eluting stent6 deaths82 reports
- Resolute Onyx™ RONYX30022UXMedtronic, Inc. · Coronary drug-eluting stent6 deaths76 reports
- SYNERGY™ H7493926024300Boston Scientific Corporation · Coronary drug-eluting stent3 deaths75 reports
- Resolute Onyx™ RONYX30015UXMedtronic, Inc. · Coronary drug-eluting stent3 deaths73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page