Skip to content

Pulsar-18 T3 5/60/90Recalls & reported problems

Biotronik AG · Class 3 · Implant

All models of Pulsar-18 T3 (52)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07640130446489), received between Jun 15, 2020 and Nov 9, 2020 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Fracture1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 9, 2020Mechanical Problem; Mechanical Problem

    A PULSAR-18 T3 STENT SYSTEM WAS CHOSEN FOR TREATMENT OF A PRE-DILATED MODERATELY CALCIFIED LESION (90 PERCENT STENOSIS DEGREE) IN A MILDLY TORTUOUS PROXIMAL SFA. AFTER STENT DEPLOYMENT IT WAS NECESSARY TO TAKE OFF THE STENT SYSTEM TOGETHER WITH THE GUIDEWIRE. IT WAS NOTICED THAT THE SHAFT WAS COMPLETELY WRINKLED CATCHING THE GUIDEWIRE. AFTER STRETCHING THE SHAFT IT WAS POSSIBLE TO REMOVE THE GUIDEWIRE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10813204 ↗
  • MalfunctionJun 15, 2020Fracture; Fracture

    A PULSAR-18 T3 STENT SYSTEM WAS CHOSEN FOR TREATMENT OF A CALCIFIED LESION WITH 70 PERCENT STENOSIS DEGREE. AFTER SUCCESSFUL STENT DEPLOYMENT RESISTANCE WAS FELT WHILE IT WAS ATTEMPTED TO REMOVE THE SYSTEM. EVENTUALLY THE DEVICE FRACTURED DURING WITHDRAWAL. ALL DEVICE PARTS WERE REMOVED FROM THE PATIENTS BODY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10155366 ↗

Other STENT, SUPERFICIAL FEMORAL ARTERY devices

All STENT, SUPERFICIAL FEMORAL ARTERY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page