Skip to content

Pulsar-18 T3 5/40/135Recalls & reported problems

Biotronik AG · Class 3 · Implant

All models of Pulsar-18 T3 (52)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07640130446878), received between Jul 20, 2023 and Dec 15, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2
  • Difficult to Remove1
  • Failure to Advance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 15, 2025Fracture; Fracture

    THE PULSAR-18 T3 PERIPHERAL SELF-EXPANDABLE STENT SYSTEM WAS SELECTED FOR TREATMENT. THE CATHETER WAS TORN (~ PROXIMAL 20CM) INSIDE PATIENT. EVERYTHING COULD BE RETRIEVED FROM THE PATIENT. FURTHER INFORMATION IS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23801858 ↗
  • MalfunctionJan 20, 2025Failure to Advance; Failure to Advance

    A DISSECTION WAS FOUND IN THE P2 SEGMENT OF THE POPLITEAL ARTERY, WHICH OCCURRED DURING PRE-DILATATION. THE PULSAR-18 T3 PERIPHERAL SELF-EXPANDABLE STENT SYSTEM WAS SELECTED FOR TREATMENT. THE ANGIO SHOWED THAT THE DISSECTION WAS LARGE AND THE VASCULAR MORPHOLOGY WAS RELATIVELY TWISTED, RESULTING IN A LARGE RESISTANCE TO THE STENT DELIVERY. EVENTUALLY, THE STENT COULD NOT REACH THE LESION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21193214 ↗
  • MalfunctionNov 2, 2023Difficult to Remove; Difficult to Remove

    A PULSAR-18 SELF-EXPANDABLE STENT SYSTEM WAS SELECTED FOR TREATMENT. AFTER SUCCESSFUL STENT IMPLANTATION, THE DELIVERY SYSTEM OF THE PULSAR-18 T3 GOT STUCK ON THE GUIDEWIRE. A COMPLETE RETRACTION WAS ONLY POSSIBLE WITH GREAT FORCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18057827 ↗
  • MalfunctionJul 20, 2023Fracture; Fracture

    A PULSAR-18 SELF-EXPANDABLE STENT SYSTEM WAS SELECTED FOR TREATMENT. AFTER SUCCESSFUL STENT IMPLANTATION, WHEN PULLING BACK THE DELIVERY SYSTEM, A PIECE BROKE OFF, BUT COULD BE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17363856 ↗

Other STENT, SUPERFICIAL FEMORAL ARTERY devices

All STENT, SUPERFICIAL FEMORAL ARTERY devices →

Sources and updates