Skip to content

Pulsar-18 T3 5/80/135Recalls & reported problems

Biotronik AG · Class 3 · Implant

All models of Pulsar-18 T3 (52)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07640130446892), received between Feb 26, 2020 and Apr 22, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation Failure2
  • Lack of Effect1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Device Embedded In Tissue or Plaque1
  • Hematoma1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 22, 2025Activation Failure; Activation Failure

    THE PULSAR-18 T3 PERIPHERAL SELF-EXPANDABLE STENT SYSTEM WAS SELECTED FOR TREATMENT OF A SEVERELY CALCIFIED LESION (60 PERCENT STENOSIS DEGREE) IN A MILDLY TORTUOUS PART OF THE MID POPLITEAL ARTERY. THE LESION WAS PRE-DILATED AS REQUIRED, AND THE DEVICE WAS POSITIONED, BUT THE STENT DID NOT EXPAND. FINALLY, WHEN THE SYSTEM WAS REMOVED, THE STENT CAME OUT AS WELL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21884160 ↗
  • MalfunctionMar 7, 2023Lack of Effect; Lack of Effect

    A PK PAPYRUS COVERED STENT SYSTEM WAS SELECTED FOR TREATMENT OF A TYPE III PERFORATION IN THE MID LAD, WHICH OCCURRED DURING STENTING WITH A DES. THE PK PAPYRUS STENT DID NOT SEAL THE PERFORATION PROPERLY. ANOTHER PK PAPYRUS WAS IMPLANTED OVERLAPPING, AND THE PERFORATION WAS SUCCESSFULLY SEALED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16497496 ↗
  • MalfunctionFeb 26, 2020Activation Failure; Activation Failure

    A PULSAR-18 T3 STENT SYSTEM WAS CHOSEN FOR TREATMENT OF A MODERATELY CALCIFIED LESION (100 PERCENT STENOSIS DEGREE) IN A MILDLY TORTUOUS POPLITEAL ARTERY. THE SHEATH COULD NOT BE CROSSED WITH THE DEVICE. DURING WITHDRAWAL THE STENT OPENED UP INSIDE AND OUTSIDE PATIENT. THE STENT WAS DELIBERATELY CUT AND 1MM REMAINED IN PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque; Hematoma

    MAUDE report 9758264 ↗

Other STENT, SUPERFICIAL FEMORAL ARTERY devices

All STENT, SUPERFICIAL FEMORAL ARTERY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page