Orsiro Mission 2.25/15/140Recalls & reported problems

Biotronik AG · Class 3 · Implant

All models of Orsiro Mission (49)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07640130455931), received between Jul 5, 2022 and Jun 4, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Failure to Advance” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Advance5
  • Device Dislodged or Dislocated2
  • Device Handling Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 4, 2026Device Handling Problem

    THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A SEVERELY CALCIFIED LESION (90% STENOSIS DEGREE) IN A MILDLY TORTUOUS SEGMENT OF THE PROXIMAL CX. THE STENT AS INSERTED TO THE PROXIMAL CIRCUMFLEX BUT WAS TOO LONG AND WAS REACHING INTO THE PROXIMAL LAD. THE STENT WAS REMOVED AND A SHORTER STENT WAS USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25387751 ↗
  • MalfunctionMay 5, 2026Failure to Advance

    THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION (80 % STENOSIS DEGREE) IN A SEVERELY TORTUOUS SEGMENT OF THE MEDIAL LCX. A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25080469 ↗
  • MalfunctionApr 14, 2026Failure to Advance

    THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION (80 % STENOSIS DEGREE) IN A MILDLY TORTUOUS SEGMENT OF THE OSTIAL OM. A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24870959 ↗
  • MalfunctionFeb 13, 2026Failure to Advance

    THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION IN A MILDLY TORTUOUS SEGMENT OF THE OSTIAL 1ST DIAGONAL ARTERY. A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24332225 ↗
  • MalfunctionJan 23, 2026Failure to Advance

    THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A SEVERELY CALCIFIED LESION IN A MODERATELY TORTUOUS SEGMENT OF THE MEDIAL LCX. A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION. THE TARGET LESION WAS FURTHER PRE-DILATED AND THE COMPLAINT DEVICE WAS RE-INSERTED BUT STILL WAS NOT ABLE TO CROSS THE LESION. THE PROCEDURE WAS STOPPED AND WILL BE RE-SCHEDULED FOR A ROTABLATOR PREPARATION OF THE LESION AND FOLLOWING STENTING WITH AN ORSIRO MISSION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24159224 ↗
  • MalfunctionSep 21, 2023Failure to Advance; Device Dislodged or Dislocated; Failure to Advance; Device Dislodged or Dislocated

    AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION IN THE MODERATELY TORTUOUS DIAGONAL. THE LESION COULD NOT BE CROSSED WITH THE DEVICE. UPON WITHDRAWAL IT DISLODGED FROM THE BALLOON AND STAYED INSIDE THE LESION. THE STENT WAS REMOVED FROM THE VESSEL WITH A NC BALLOON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17789914 ↗
  • MalfunctionJul 5, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT. THE STENT CAME LOOSE BEFORE EXPANDING THE BALLOON. (MEDWATCH MW5110033).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14919605 ↗

Other Coronary drug-eluting stent devices

All Coronary drug-eluting stent devices →