Orsiro Mission 2.75/18/140Recalls & reported problems
Biotronik AG · Class 3 · Implant
All models of Orsiro Mission (49)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07640130456013), received between Jan 12, 2022 and Mar 6, 2026 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated1
- Failure to Advance1
- Fluid/Blood Leak1
- Material Deformation1
- Material Rupture1
- Migration1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Vascular Dissection1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 6, 2026Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A LESION IN AN UNKNOWN TARGET VESSEL. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24529522 ↗MalfunctionMar 5, 2025Fluid/Blood Leak; Material Rupture; Fluid/Blood Leak; Material Rupture
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A PREDILATED LESION IN THE DISTAL RCA. THE STENT BALLOON RUPTURED PRIOR TO DEPLOYING THE STENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21527477 ↗MalfunctionFeb 15, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A DISTAL LAD LESION. THE STENT WAS DELIVERED BUT NEEDED TO CHANGE POSITION, SINCE IT MOVED. AFTER REPOSITIONING IT WAS NOTICED THAT THE STENT WAS DISLODGED. THE STENT DELIVERY WAS PULLED BACK, AND IT WAS CONFIRMED THAT THE STENT HAD DISLODGED. ANOTHER STENT WAS DEPLOYED TO CRUSH THE DISLODGED STENT AGAINST THE VESSEL WALL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18712744 ↗InjuryJan 12, 2022Material Deformation; Migration; Material Deformation; Migration
AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A LESION WITH 99 PERCENT STENOSIS DEGREE IN THE CX. AFTER POST-DILATATION OF THE ORSIRO MISSION STENT WITH AN NC BALLOON AND FURTHER PREPARATION OF ANOTHER DISTAL LESION, THE BALLOON GOT STUCK AT THE DISTAL END OF THE IMPLANTED STENT. DURING WITHDRAWAL OF THE BALLOON, THE PREVIOUSLY IMPLANTED STENT BECAME DEFORMED, MIGRATED AND WAS FINALLY ADAPTED TO THE VESSEL WALL. ADDITIONALLY, THERE WAS A NON-FLOW-LIMITING DISSECTION WHICH WAS TREATED WITH A DES. FINALLY, THE PATIENT WAS STABLE AND WAS DISCHARGED HOME.
Patient problems: Vascular Dissection
MAUDE report 13236397 ↗
Other Coronary drug-eluting stent devices
- SYNERGY™ H7493926038300Boston Scientific Corporation · Coronary drug-eluting stent8 deaths124 reports
- SYNERGY™ H7493926016300Boston Scientific Corporation · Coronary drug-eluting stent7 deaths106 reports
- SYNERGY™ H7493926020300Boston Scientific Corporation · Coronary drug-eluting stent9 deaths90 reports
- SYNERGY™ H7493926016250Boston Scientific Corporation · Coronary drug-eluting stent6 deaths82 reports
- XIENCE Sierra™ 1550350-33Abbott Vascular Inc. · Coronary drug-eluting stent1 recall32 reports
- Resolute Onyx™ RONYX30022UXMedtronic, Inc. · Coronary drug-eluting stent6 deaths76 reports
- SYNERGY™ H7493926024300Boston Scientific Corporation · Coronary drug-eluting stent3 deaths75 reports
- Resolute Onyx™ RONYX30015UXMedtronic, Inc. · Coronary drug-eluting stent3 deaths73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page