Orsiro Mission 2.5/22/140Recalls & reported problems
Biotronik AG · Class 3 · Implant
All models of Orsiro Mission (49)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 07640130456068), received between Aug 1, 2022 and Jun 3, 2026 — classified as 13 malfunctions.
The most frequently reported specific device problem is “Failure to Advance” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Advance7
- Material Deformation3
- Device Dislodged or Dislocated2
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions13
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 3, 2026Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF AN UNSPECIFIED LESION IN AN UNKNOWN VESSEL. IT IS UNKNOWN IF A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25373559 ↗MalfunctionMay 22, 2026Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION (90 % STENOSIS DEGREE) IN A MILDLY TORTUOUS SEGMENT OF THE MEDIAL RCA. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25254569 ↗MalfunctionMay 19, 2026Material Deformation
THE ORSIRO MISSION WAS SELECTED FOR TREATMENT OF A MODERATE CALCIFIED LESION, IN THE MID TO DISTAL LAD. THE STENT WAS DEFORMED. IT WAS DETECTED OUTSIDE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25213578 ↗MalfunctionApr 10, 2026Material Deformation
WHEN PULLING THE ORSIRO MISSION OUT OF THE HOOP AND PREPARING TO LOAD OVER THE GUIDEWIRE IT WAS NOTICED, THAT THE DISTAL PART OF THE STENT WAS FLARED UP OFF THE BALLOON. THE DEVICE WAS NEVER INSERTED INSIDE THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24843488 ↗MalfunctionApr 10, 2026Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION (90 % STENOSIS DEGREE) IN A SEVERELY TORTUOUS SEGMENT OF THE MEDIAL LAD. A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24838891 ↗MalfunctionMar 24, 2026Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A LESION IN A SEGMENT OF UNSPECIFIED VESSEL. IT IS UNKNOWN IF A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24672686 ↗MalfunctionMar 24, 2026Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A SEVERELY CALCIFIED LESION IN A MODERATELY TORTUOUS SEGMENT OF THE MEDIAL LAD. A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION. THE PHYSICIAN WAS NOT ABLE TO CROSS THE LESION WITH ANY OTHER DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24672231 ↗MalfunctionMar 17, 2026Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A LESION IN A SEGMENT OF THE MEDIAL LCX. A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24615205 ↗MalfunctionJan 27, 2026Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION (80 % STENOSIS DEGREE) IN A MODERATELY TO SEVERELY TORTUOUS SEGMENT OF THE MEDIAL LAD. A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24188023 ↗MalfunctionJun 3, 2024Use of Device Problem; Use of Device Problem
AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT AND WAS DEPLOYED IN THE PATIENT, ALTHOUGH THE PRODUCT HAD ALREADY EXPIRED. THE HOSPITAL FOUND THAT ITS PROTOCOL WAS NOT FOLLOWED. THE DEVICE WAS NOT CONFIRMED BEFORE IT WAS OPENED ON THE TABLE, AND THE HOSPITAL SYSTEM WAS DOWN FOR AN HOUR, SO THERE WAS NO WARNING THAT THE DEVICE HAD EXPIRED. NO PATIENT COMPLICATIONS WERE REPORTED. THE PATIENT WILL BE FOLLOWED UP FOR 90 DAYS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19449883 ↗MalfunctionAug 14, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT. THE STENT UNSEATED FROM DELIVER CATHETER BEFORE ENTERING THE PATIENTS BODY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17541726 ↗MalfunctionFeb 8, 2023Material Deformation; Material Deformation
AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION (99 PERCENT STENOSIS DEGREE) IN THE DISTAL LAD. A STENT STRUT WAS NOTED TO BE RAISED AT TIME OF STENT PREPARATION PRIOR TO ENTERING THE PATIENT. A NEW STENT WAS OPENED AND UTILIZED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16320813 ↗MalfunctionAug 1, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A SEVERELY CALCIFIED LESION IN THE MID LAD. THE STENT DISLODGED WHILE ATTEMPTING TO ADVANCE THE DEVICE TO THE LESION WHICH WAS HEAVILY CALCIFIED AND HAD LIKELY CONCURRENT INTIMAL DISRUPTION AT THE DISTAL EDGE OF A MORE PROXIMAL ORSIRO MISSION STENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15139280 ↗
Other Coronary drug-eluting stent devices
- SYNERGY™ H7493926038300Boston Scientific Corporation · Coronary drug-eluting stent8 deaths124 reports
- SYNERGY™ H7493926016300Boston Scientific Corporation · Coronary drug-eluting stent7 deaths106 reports
- SYNERGY™ H7493926020300Boston Scientific Corporation · Coronary drug-eluting stent9 deaths90 reports
- SYNERGY™ H7493926016250Boston Scientific Corporation · Coronary drug-eluting stent6 deaths82 reports
- XIENCE Sierra™ 1550350-33Abbott Vascular Inc. · Coronary drug-eluting stent1 recall32 reports
- Resolute Onyx™ RONYX30022UXMedtronic, Inc. · Coronary drug-eluting stent6 deaths76 reports
- SYNERGY™ H7493926024300Boston Scientific Corporation · Coronary drug-eluting stent3 deaths75 reports
- Resolute Onyx™ RONYX30015UXMedtronic, Inc. · Coronary drug-eluting stent3 deaths73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.