Orsiro Mission 3.5/22/140Recalls & reported problems

Biotronik AG · Class 3 · Implant

All models of Orsiro Mission (49)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 07640130456099), received between Jan 27, 2022 and Jul 22, 2026 — classified as 1 injury, 11 malfunctions.

The most frequently reported specific device problem is “Failure to Advance” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 2 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Advance8
  • Device Dislodged or Dislocated2
  • Fluid/Blood Leak1
  • Inflation Problem1
  • Material Puncture/Hole1
  • Patient Device Interaction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11
  • Thrombosis/Thrombus1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 22, 2026Failure to Advance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25904112 ↗
  • MalfunctionJun 10, 2026Failure to Advance

    THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF AN UNSPECIFIED LESION IN A SEGMENT OF THE PROXIMAL LAD. IT IS UNKNOWN IF A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION. THE COMPLAINT DEVICE WAS NOT RETURNED FROM THE HOSPITAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25453264 ↗
  • MalfunctionJun 5, 2026Failure to Advance

    THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A LESION (60 % STENOSIS DEGREE) IN AN UNKNOWN SEGMENT OF A MEDIAL VESSEL. IT IS UNKNOWN IF A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25405791 ↗
  • MalfunctionApr 14, 2026Failure to Advance

    THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION (90 % STENOSIS DEGREE) IN A MODERATELY TORTUOUS SEGMENT OF AN UNSPECIFIED MEDIAL VESSEL. IT IS UNKNOWN IF A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24870951 ↗
  • MalfunctionFeb 26, 2026Device Dislodged or Dislocated

    THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION (60% STENOSIS DEGREE) IN A MILDLY TORTUOUS SECTION OF THE OSTIAL CX/LEFT MAIN. THE STENT CAME OFF THE BALLOON BEFORE REACHING THE PRE-DILATED TARGET LESION. THE UNDEPLOYED STENT WAS RETRIEVED WITH A SNARE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24452149 ↗
  • MalfunctionFeb 13, 2026Failure to Advance

    THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION IN A MODERATELY TORTUOUS SEGMENT OF THE MEDIAL LAD. IT IS UNKNOWN IF A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24332223 ↗
  • MalfunctionJan 9, 2026Failure to Advance

    THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION (89 % STENOSIS DEGREE) IN A SEVERELY TORTUOUS SEGMENT OF THE DISTAL RCA. PRE-DILATION OF THE LESION WAS PER-FORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION. THE PHYSICIAN WAS ABLE TO CROSS THE LESION USING ANOTHER MANUFACTURERS STENT. THE COMPLAINT DEVICE WAS DISCARDED AT THE HOSPITAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24019209 ↗
  • MalfunctionNov 17, 2025Failure to Advance; Failure to Advance

    THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A SEVERELY CALCIFIED LESION (90 % STENOSIS DEGREE) IN A MODERATELY TORTUOUS SEGMENT OF A MEDIAL CORONARY BIFURCATION. IT IS UNKNOWN IF A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23566641 ↗
  • MalfunctionNov 11, 2024Failure to Advance; Failure to Advance

    AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION (90 PERCENT STENOSIS DEGREE) IN A MILDLY TORTUOUS PART OF THE DISTAL RCA. AFTER PRE-DILATION OF THE LESION, THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENTS BODY BUT COULD NOT CROSS THE LESION. DURING THE ADVANCEMENT OF THE DEVICE, THE WIRE AND GUIDING CATHETER POSITION WERE LOST. A SMALL WIRE PERFORATION WAS NOTED. NO FURTHER ATTEMPTS TO TREAT THE LESION WERE PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20658861 ↗
  • MalfunctionApr 22, 2024Fluid/Blood Leak; Device Dislodged or Dislocated; Fluid/Blood Leak; Device Dislodged or Dislocated

    AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A LESION (STENOSIS DEGREE: 95 PERCENT) IN THE MILDLY TORTUOUS MID SVG TO DIAGONAL. A SHOCKWAVE PRE-DILATATION WAS DONE OVER A FILTER-WIRE PROTECTION DEVICE. THE ORSIRO MISSION WAS INTRODUCED BUT COULD NOT BE INFLATED MORE THAN 2 ATM DUE TO HOLE IN BALLOON. STENT AND BALLOON HAD TO BE REMOVED WITH GUIDE CATHETER AND SHEATH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19154468 ↗
  • InjuryJun 26, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MILDLY CALCIFIED LESION (95 PERCENT STENOSIS DEGREE) IN THE MILDLY TORTUOUS MID RCA. AN ACUTE STENT THROMBOSIS OCCURRED 1.5 H AFTER IMPLANTATION OF THE ORSIRO MISSION STENT. AFTER BALLOONING OF THE STENT THROMBOSIS, A TIMI 3 FLOW WAS RE-ESTABLISHED. PATIENT WAS LOADED ON ASPIRIN AND PLAVIX AFTER INITIAL PROCEDURE. IV CANGRELOR WAS GIVEN DURING REPEAT INTERVENTION TO OPEN THROMBOSED STENT. PATIENT WAS TRANSFERRED TO HEART CATH RECOVERY FOR DURATION OF CANGRELOR DRIP AND THEN TRANSFERRED TO ICU. ONE DAY AFTER PROCEDURE THE PATIENT WAS DOING GREAT AND WAS DISCHARGED HOME.

    Patient problems: Thrombosis/Thrombus

    MAUDE report 17206215 ↗
  • MalfunctionJan 27, 2022Inflation Problem; Material Puncture/Hole; Inflation Problem; Material Puncture/Hole

    AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT. THE DEVICE WAS PLACED DOWN IN A VEIN GRAFT AND INFLATED TO NOMINAL PRESSURE, BUT THE BALLOON WOULD NOT INFLATE DUE TO A PIN HOLE AT DISTAL END OF BALLOON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13367508 ↗

Other Coronary drug-eluting stent devices