Orsiro Mission 4/22/140Recalls & reported problems
Biotronik AG · Class 3 · Implant
All models of Orsiro Mission (49)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07640130456105), received between Dec 23, 2025 and Feb 4, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Failure to Advance” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Advance3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 4, 2026Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION (80 % STENOSIS DEGREE) IN A MODERATELY TORTUOUS SEGMENT OF THE DISTAL RCA. A PRE-DILATION OF THE LESION WAS PERFORMED VIA SHOCKWAVE BALLOON. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION. THE COMPLAINT DEVICE WAS NOT RETURNED FROM THE HOSPITAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24256793 ↗MalfunctionJan 12, 2026Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A SEVERELY CALCIFIED LESION (90 % STENOSIS DEGREE) IN A SEVERELY TORTUOUS SEGMENT OF THE DISTAL RCA-PLA. A PRE-DILATION OF THE LESION WAS PERFORMED, AND THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION. IT WAS POSSIBLE TO CROSS THE LESION USING A 2.25/9 ORSIRO MISSION, FOLLOWED BY ANOTHER 2.25/9 ORSIRO MISSION WITHOUT ISSUE AND GOOD RESULT. THE COMPLAINT DEVICE WAS DISCARDED AT THE HOSPITAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24035383 ↗MalfunctionDec 23, 2025Failure to Advance; Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION (90 % STENOSIS DEGREE) IN A SEVERELY TORTUOUS SEGMENT OF THE MEDIAL TO DISTAL RCA. A PRE-DILATION OF THE LESION WAS PERFORMED WITH. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION. THE PHYSICIAN WAS ABLE TO CROSS THE LESION USING A 3.5/22 ORSIRO MISSION. THE COMPLAINT DEVICE WAS DISCARDED AT THE HOSPITAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23874594 ↗
Other Coronary drug-eluting stent devices
- SYNERGY™ H7493926038300Boston Scientific Corporation · Coronary drug-eluting stent8 deaths124 reports
- SYNERGY™ H7493926016300Boston Scientific Corporation · Coronary drug-eluting stent7 deaths106 reports
- SYNERGY™ H7493926020300Boston Scientific Corporation · Coronary drug-eluting stent9 deaths90 reports
- SYNERGY™ H7493926016250Boston Scientific Corporation · Coronary drug-eluting stent6 deaths82 reports
- XIENCE Sierra™ 1550350-33Abbott Vascular Inc. · Coronary drug-eluting stent1 recall32 reports
- Resolute Onyx™ RONYX30022UXMedtronic, Inc. · Coronary drug-eluting stent6 deaths76 reports
- SYNERGY™ H7493926024300Boston Scientific Corporation · Coronary drug-eluting stent3 deaths75 reports
- Resolute Onyx™ RONYX30015UXMedtronic, Inc. · Coronary drug-eluting stent3 deaths73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page