Orsiro Mission 3.5/26/140Recalls & reported problems
Biotronik AG · Class 3 · Implant
All models of Orsiro Mission (49)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07640130456150), received between Sep 7, 2023 and Jun 10, 2026 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated2
- Failure to Advance1
- Patient Device Interaction Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Chest Pain1
- Thrombosis/Thrombus1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 10, 2026Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION IN A MODERATELY TORTUOUS SEGMENT OF A MEDIAL VESSEL. IT IS UNKNOWN IF A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION. THE INTERVENTION WAS COMPLETED BY USING ANOTHER STENT. THE COMPLAINT DEVICE WAS NOT RETURNED FROM THE HOSPITAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25453263 ↗MalfunctionDec 16, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT. THE STENT FELL OFF THE BALLOON PRIOR TO IMPLANTATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20939337 ↗InjuryOct 16, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF THE PROXIMAL LAD. IMMEDIATELY AFTER PROCEDURE AN IN-STENT THROMBOSIS OCCURRED. PATIENT DEVELOPED CHEST PAIN. RE-BALLOONING OF THE STENT WAS PERFORMED AND ADDITIONAL HEPARIN WAS GIVEN. PATIENT WAS TREATED FOR HEPARIN-INDUCED THROMBOCYTOPENIA (HIT). PHYSICIAN SHARED THAT DURING THE CASE THE PATIENTS ACT WOULD BE 182, GIVE HEPARIN WOULD GO TO 350 THEN WHEN RE-CHECKED AGAIN WOULD BE BACK DOWN TO 180.
Patient problems: Chest Pain; Thrombosis/Thrombus
MAUDE report 20468091 ↗MalfunctionSep 7, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT, BUT THE STENT DISLODGED FROM THE BALLOON WHEN THE PACKAGING STYLET WAS REMOVED FROM THE COMPLAINT INSTRUMENT. THIS OCCURRED OUTSIDE OF THE PATIENT DURING THE PREPARATIONS FOR THE INTERVENTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17698983 ↗
Other Coronary drug-eluting stent devices
- SYNERGY™ H7493926038300Boston Scientific Corporation · Coronary drug-eluting stent8 deaths124 reports
- SYNERGY™ H7493926016300Boston Scientific Corporation · Coronary drug-eluting stent7 deaths106 reports
- SYNERGY™ H7493926020300Boston Scientific Corporation · Coronary drug-eluting stent9 deaths90 reports
- SYNERGY™ H7493926016250Boston Scientific Corporation · Coronary drug-eluting stent6 deaths82 reports
- XIENCE Sierra™ 1550350-33Abbott Vascular Inc. · Coronary drug-eluting stent1 recall32 reports
- Resolute Onyx™ RONYX30022UXMedtronic, Inc. · Coronary drug-eluting stent6 deaths76 reports
- SYNERGY™ H7493926024300Boston Scientific Corporation · Coronary drug-eluting stent3 deaths75 reports
- Resolute Onyx™ RONYX30015UXMedtronic, Inc. · Coronary drug-eluting stent3 deaths73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page