Skip to content

Orsiro Mission 3.5/26/140Recalls & reported problems

Biotronik AG · Class 3 · Implant

All models of Orsiro Mission (49)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07640130456150), received between Sep 7, 2023 and Jun 10, 2026 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated2
  • Failure to Advance1
  • Patient Device Interaction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Chest Pain1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 10, 2026Failure to Advance

    THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION IN A MODERATELY TORTUOUS SEGMENT OF A MEDIAL VESSEL. IT IS UNKNOWN IF A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION. THE INTERVENTION WAS COMPLETED BY USING ANOTHER STENT. THE COMPLAINT DEVICE WAS NOT RETURNED FROM THE HOSPITAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25453263 ↗
  • MalfunctionDec 16, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT. THE STENT FELL OFF THE BALLOON PRIOR TO IMPLANTATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20939337 ↗
  • InjuryOct 16, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF THE PROXIMAL LAD. IMMEDIATELY AFTER PROCEDURE AN IN-STENT THROMBOSIS OCCURRED. PATIENT DEVELOPED CHEST PAIN. RE-BALLOONING OF THE STENT WAS PERFORMED AND ADDITIONAL HEPARIN WAS GIVEN. PATIENT WAS TREATED FOR HEPARIN-INDUCED THROMBOCYTOPENIA (HIT). PHYSICIAN SHARED THAT DURING THE CASE THE PATIENTS ACT WOULD BE 182, GIVE HEPARIN WOULD GO TO 350 THEN WHEN RE-CHECKED AGAIN WOULD BE BACK DOWN TO 180.

    Patient problems: Chest Pain; Thrombosis/Thrombus

    MAUDE report 20468091 ↗
  • MalfunctionSep 7, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT, BUT THE STENT DISLODGED FROM THE BALLOON WHEN THE PACKAGING STYLET WAS REMOVED FROM THE COMPLAINT INSTRUMENT. THIS OCCURRED OUTSIDE OF THE PATIENT DURING THE PREPARATIONS FOR THE INTERVENTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17698983 ↗

Other Coronary drug-eluting stent devices

All Coronary drug-eluting stent devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page