Orsiro Mission 2.25/30/140Recalls & reported problems
Biotronik AG · Class 3 · Implant
All models of Orsiro Mission (49)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07640130456174), received between Nov 2, 2021 and May 5, 2026 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Failure to Advance” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Advance4
- Material Deformation2
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 5, 2026Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A LESION IN A SEGMENT OF THE MEDIAL LAD. IT IS UNKNOWN IF A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25079822 ↗MalfunctionMar 17, 2026Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A SEVERELY CALCIFIED LESION (95 % STENOSIS DEGREE) IN A MODERATELY TORTUOUS SEGMENT OF THE DISTAL LAD. A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24616623 ↗MalfunctionJan 27, 2026Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A SEVERELY CALCIFIED LESION (80 % STENOSIS DEGREE) IN A SEVERELY TORTUOUS SEGMENT OF THE PROXIMAL LCX. A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION. THE PHYSICIAN WAS ABLE TO CROSS THE LESION USING A 2.5/18 ORSIRO MISSION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24188024 ↗MalfunctionFeb 20, 2025Failure to Advance; Material Deformation; Failure to Advance; Material Deformation
THE ORSIRO MISSION STENT SYSTEM WAS SELECTED FOR TREATMENT OF A SEVERELY CALCIFIED LESION WITHOUT PRE-DILATATION. THE DEVICE WAS PULLED BACK OUT OF PATIENT, AND IT WAS DETECTED THAT ONE STRUT WAS BENT. SUBSEQUENTLY, PRE-DILATATION WAS PERFORMED, AND A 2ND STENT OF SAME SIZE WAS IMPLANTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21421993 ↗MalfunctionMar 23, 2022Material Deformation; Material Deformation
AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT. A STENT DEFORMATION WAS DETECTED RIGHT OUT OF THE PACKAGING. A STRUT WAS STICKING OUT AT MID STENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13871161 ↗MalfunctionNov 2, 2021Fracture; Fracture
AN ORSIRO DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A PRE-DILATED LESION. WHEN INTRODUCING THE DEVICE INTO THE BODY, THE BACK END OF THE SHAFT BENT. RESISTANCE WAS FELT WHEN THE DEVICE WAS ADVANCED DEEPER INTO THE VESSEL. BY PUSHING HARDER, THE KINK GOT WORSE. WHILE REMOVING THE STENT SYSTEM FROM THE GUIDE, THE BACK END OF THE SHAFT BROKE APART.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12741002 ↗
Other Coronary drug-eluting stent devices
- SYNERGY™ H7493926038300Boston Scientific Corporation · Coronary drug-eluting stent8 deaths124 reports
- SYNERGY™ H7493926016300Boston Scientific Corporation · Coronary drug-eluting stent7 deaths106 reports
- SYNERGY™ H7493926020300Boston Scientific Corporation · Coronary drug-eluting stent9 deaths90 reports
- SYNERGY™ H7493926016250Boston Scientific Corporation · Coronary drug-eluting stent6 deaths82 reports
- XIENCE Sierra™ 1550350-33Abbott Vascular Inc. · Coronary drug-eluting stent1 recall32 reports
- Resolute Onyx™ RONYX30022UXMedtronic, Inc. · Coronary drug-eluting stent6 deaths76 reports
- SYNERGY™ H7493926024300Boston Scientific Corporation · Coronary drug-eluting stent3 deaths75 reports
- Resolute Onyx™ RONYX30015UXMedtronic, Inc. · Coronary drug-eluting stent3 deaths73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026