Orsiro Mission 3/30/140Recalls & reported problems
Biotronik AG · Class 3 · Implant
All models of Orsiro Mission (49)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 07640130456204), received between Nov 1, 2023 and May 26, 2026 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Failure to Advance” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Advance9
- Material Deformation2
- Device Dislodged or Dislocated1
Patient problems
- No Clinical Signs, Symptoms or Conditions12
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 26, 2026Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION IN A SEGMENT OF THE MEDIAL TO DISTAL SUPERIOR MESENTERIC ARTERY (SMA). IT IS UNKNOWN IF A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION. THE PHYSICIAN WAS ABLE TO CROSS AND TREAT THE LESION USING A SELF-EXPANDABLE STENT-GRAFT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25273770 ↗MalfunctionMay 26, 2026Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A SEVERELY CALCIFIED LESION IN A SEVERELY TORTUOUS SEGMENT OF THE MEDIAL LAD. A PRE-DILATION OF THE LESION WAS PERFORMED INCLUDING SHOCKWAVE BALLOONS. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25273769 ↗MalfunctionMay 5, 2026Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION IN A SEGMENT OF THE MEDIAL TO DISTAL LAD. IT IS UNKNOWN IF A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25079824 ↗MalfunctionApr 24, 2026Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION IN A MODERATELY TO SEVERELY TORTUOUS SEGMENT OF THE MEDIAL LAD. IT IS UNKNOWN IF A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24977859 ↗MalfunctionApr 11, 2026Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A LESION IN A SEGMENT OF AN UNDEFINED VESSEL. IT IS UNKNOWN IF A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24851483 ↗MalfunctionMar 17, 2026Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION (90 % STENOSIS DEGREE) IN A MODERATELY TORTUOUS SEGMENT OF THE MEDIAL LAD. A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24616625 ↗MalfunctionMar 17, 2026Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A SEVERELY CALCIFIED LESION IN A MODERATELY TO SEVERELY TORTUOUS SEGMENT FROM THE PROXIMAL LCX TO THE OM2. A PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION "DUE TO THE SEVERE TORTUOSITY OF THE VESSEL".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24616621 ↗MalfunctionFeb 24, 2026Material Deformation
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A SEVERELY CALCIFIED LESION IN A SEVERELY TORTUOUS SEGMENT OF THE OSTIAL CX. WHILE INSERTING THE ORSIRO MISSION, THE STENT BUCKLED ON ITSELF, STILL ON THE BALLOON. AFTER REMOVAL FROM BODY, THE STENT IS STILL INTACT ON BALLOON. A STENT FROM ANOTHER MANUFACTURER WAS USED TO COMPLETE THE INTERVENTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24427059 ↗MalfunctionJan 9, 2026Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MILDLY CALCIFIED LESION (80 % STENOSIS DEGREE) IN A MILDLY TORTUOUS SEGMENT OF THE MEDIAL RCA. PRE-DILATION OF THE LESION WAS PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION. THE COMPLAINT DEVICE WAS DISCARDED AT THE HOSPITAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24017600 ↗MalfunctionNov 26, 2025Material Deformation; Material Deformation
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF THE MID TO DISTAL LAD. DURING THE REMOVAL OF THE STENT COVER OF THE ORSIRO MISSION THERE WAS RESISTANCE. THE STENT WAS DISCOVERED TO BE DAMAGED, AND THE BALLOON APPEARED TO BE INFLATED ON THE END OF THE STENT. ANOTHER ORSIRO MISSION WAS USED WITHOUT ANY ISSUES. THE PATIENT HAD NO COMPLICATIONS DURING OR POST PROCEDURE AND WAS DISCHARGED HOME FROM THE HOSPITAL THE NEXT DAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23653769 ↗MalfunctionNov 12, 2025Failure to Advance; Failure to Advance
THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION (96 % STENOSIS DEGREE) IN A SEVERELY TORTUOUS SEGMENT OF THE MEDIAL RCA. AFTER PRE-DILATION OF THE LESION, THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23530643 ↗MalfunctionNov 1, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A PREDILATED LESION. DURING INTERVENTION, A NEW HEMOSTATIC VALVE WAS USED. THE VALVE WAS NOT FULLY OPENED AND STRIPPED THE STENT OFF WHILE ENTERING THE HEMOSTATIC VALVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18048161 ↗
Other Coronary drug-eluting stent devices
- SYNERGY™ H7493926038300Boston Scientific Corporation · Coronary drug-eluting stent8 deaths124 reports
- SYNERGY™ H7493926016300Boston Scientific Corporation · Coronary drug-eluting stent7 deaths106 reports
- SYNERGY™ H7493926020300Boston Scientific Corporation · Coronary drug-eluting stent9 deaths90 reports
- SYNERGY™ H7493926016250Boston Scientific Corporation · Coronary drug-eluting stent6 deaths82 reports
- XIENCE Sierra™ 1550350-33Abbott Vascular Inc. · Coronary drug-eluting stent1 recall32 reports
- Resolute Onyx™ RONYX30022UXMedtronic, Inc. · Coronary drug-eluting stent6 deaths76 reports
- SYNERGY™ H7493926024300Boston Scientific Corporation · Coronary drug-eluting stent3 deaths75 reports
- Resolute Onyx™ RONYX30015UXMedtronic, Inc. · Coronary drug-eluting stent3 deaths73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page