Skip to content

Orsiro Mission 2.75/40/140Recalls & reported problems

Biotronik AG · Class 3 · Implant

All models of Orsiro Mission (49)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07640130456310), received between Dec 15, 2022 and Aug 28, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1
  • Fracture1
  • Inflation Problem1
  • Material Deformation1
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 28, 2026Fracture; Material Deformation; Unintended Movement

    The Orsiro Mission drug-eluting stent system was selected for treatment of a mildly calcified lesion (65% stenosis degree) in a mildly tortuous segment of the proximal RCA. During the advancement to the target lesion, resistance was met, resulting in bending and deformation of the stent struts. The device was not deployed. Subsequently, the patient moved their arm, which led to dislodgement of the guiding catheter, guidewire, and stent into the aorta. An attempt was made to retrieve all devices through the radial sheath; however, the stent together with the distal tip could not be withdrawn. A radial artery cutdown procedure was subsequently required to remove the remaining part.

    Patient problems: Foreign Body In Patient

    MAUDE report 50494942 ↗
  • MalfunctionNov 21, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION IN THE MILDLY TORTUOUS PROXIMAL LAD. THE SYSTEM WAS ADVANCED INTO THE VESSEL AND DILATED. AFTER DEVICE WAS REMOVED FROM THE VESSEL, NO STENT WAS SEEN ON THE ANGIOGRAM. THE DEVICE WAS INSPECTED, AND THE STENT WAS FOUND ON THE STYLET ON THE TABLE. AN OTHER STENT WAS USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18188712 ↗
  • MalfunctionDec 15, 2022Inflation Problem; Inflation Problem

    AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT. AFTER PREDILATATION OF A CALCIFIED LESION IN THE RCA, THE DEVICE WAS INSERTED, BUT WHEN INFLATING UP TO NORMAL PRESSURE, NO EXPANSION OF THE STENT WAS VISIBLE. THE DEVICE WAS REMOVED AND INSPECTED AND WAS FOUND STILL CRIMPED ON THE BALLOON. OUTSIDE PATIENTS BODY THE STENT COULD BE EXPANDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15987799 ↗

Other Coronary drug-eluting stent devices

All Coronary drug-eluting stent devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page