Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07640130456310), received between Dec 15, 2022 and Aug 28, 2026 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated1
- Fracture1
- Inflation Problem1
- Material Deformation1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Foreign Body In Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryAug 28, 2026Fracture; Material Deformation; Unintended Movement
The Orsiro Mission drug-eluting stent system was selected for treatment of a mildly calcified lesion (65% stenosis degree) in a mildly tortuous segment of the proximal RCA. During the advancement to the target lesion, resistance was met, resulting in bending and deformation of the stent struts. The device was not deployed. Subsequently, the patient moved their arm, which led to dislodgement of the guiding catheter, guidewire, and stent into the aorta. An attempt was made to retrieve all devices through the radial sheath; however, the stent together with the distal tip could not be withdrawn. A radial artery cutdown procedure was subsequently required to remove the remaining part.
Patient problems: Foreign Body In Patient
MAUDE report 50494942 ↗MalfunctionNov 21, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION IN THE MILDLY TORTUOUS PROXIMAL LAD. THE SYSTEM WAS ADVANCED INTO THE VESSEL AND DILATED. AFTER DEVICE WAS REMOVED FROM THE VESSEL, NO STENT WAS SEEN ON THE ANGIOGRAM. THE DEVICE WAS INSPECTED, AND THE STENT WAS FOUND ON THE STYLET ON THE TABLE. AN OTHER STENT WAS USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18188712 ↗MalfunctionDec 15, 2022Inflation Problem; Inflation Problem
AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT. AFTER PREDILATATION OF A CALCIFIED LESION IN THE RCA, THE DEVICE WAS INSERTED, BUT WHEN INFLATING UP TO NORMAL PRESSURE, NO EXPANSION OF THE STENT WAS VISIBLE. THE DEVICE WAS REMOVED AND INSPECTED AND WAS FOUND STILL CRIMPED ON THE BALLOON. OUTSIDE PATIENTS BODY THE STENT COULD BE EXPANDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15987799 ↗
Other Coronary drug-eluting stent devices
- SYNERGY™ H7493926038300Boston Scientific Corporation · Coronary drug-eluting stent8 deaths124 reports
- SYNERGY™ H7493926016300Boston Scientific Corporation · Coronary drug-eluting stent7 deaths106 reports
- SYNERGY™ H7493926020300Boston Scientific Corporation · Coronary drug-eluting stent9 deaths90 reports
- SYNERGY™ H7493926016250Boston Scientific Corporation · Coronary drug-eluting stent6 deaths82 reports
- XIENCE Sierra™ 1550350-33Abbott Vascular Inc. · Coronary drug-eluting stent1 recall32 reports
- Resolute Onyx™ RONYX30022UXMedtronic, Inc. · Coronary drug-eluting stent6 deaths76 reports
- SYNERGY™ H7493926024300Boston Scientific Corporation · Coronary drug-eluting stent3 deaths75 reports
- Resolute Onyx™ RONYX30015UXMedtronic, Inc. · Coronary drug-eluting stent3 deaths73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page