Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07640130456341), received between Jun 1, 2023 and May 26, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1
  • Failure to Advance1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 26, 2026Failure to Advance

    THE ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION (80 % STENOSIS DEGREE) IN A MODERATELY TORTUOUS SEGMENT OF AN OSTIAL VESSEL. A PRE-DILATION OF THE LESION WAS NOT PERFORMED. THE AFFECTED DEVICE WAS INTRODUCED INTO THE PATIENT'S BODY BUT COULD NOT CROSS THE LESION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25273768 ↗
  • MalfunctionJul 20, 2023Use of Device Problem; Use of Device Problem

    AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT. THE STENT WAS IMPLANTED WHICH WAS A WEEK PAST EXPIRATION DATE. THAT THE DEVICE WAS EXPIRED, WAS DETECTED AFTER IMPLANTATION. THE DEVICE CAME FROM TRUNK STOCK OF THE DISTRIBUTOR IN HAWAII. THE PHYSICIAN WAS HAPPY WITH RESULT AND IS NOT OVERLY CONCERNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17363930 ↗
  • MalfunctionJun 1, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    AN ORSIRO MISSION DRUG-ELUTING STENT SYSTEM WAS SELECTED FOR TREATMENT. HOWEVER, THE ORSIRO MISSION STENT SEPARATED FROM THE BALLOON PLATFORM. WHEN EXITING THE TIP OF THE FORTRESS WITH THE CATHETER, THE STENT ITSELF WENT (ON ITS OWN WITHOUT THE CATHETER) DOWN THE SFA AND LODGED IN THE POPLITEAL ARTERY. AN ATTEMPT WAS MADE TO RE-ENTER THE STENT WITH A BALLOON AND TO PULL BACK THE STENT INTO THE SHEATH. THE STENT WAS RETRIEVED BACK UP THE SFA, BUT THE BALLOON LOST ITS HOLD. THE STENT DISLODGED AGAIN. FINALLY, THE STENT WAS REMOVED WITH A SNARE. THE SCRUB ADMITTED THAT SHE PULLED AT THE PROTECTIVE SLEEVE INSTEAD OF THE BLACK TIP OF THE STYLET. THIS MIGHT HAVE CAUSED THE STENT TO DISLODGE FROM THE BALLOON PLATFORM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17039423 ↗

Other Coronary drug-eluting stent devices

All Coronary drug-eluting stent devices →