OPN NC 350-015-004Recalls & reported problems
SIS Medical AG · OPN NC Ø3.5x15 PTCA catheters · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07640142570608), received between May 21, 2025 and Dec 30, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Insufficient Device Problem Information2
- Material Rupture2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Insufficient Information1
- Stenosis1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 30, 2025Material Rupture; Material Rupture
THE COMPLAINANT STATED THAT HE OPN BALLOON BROKE WHEN IT REACHED 22 ATMOSPHERES FOR APPROXIMATELY 15 SECONDS EVEN THOUGH IT IS MANUFACTURED TO WITHSTAND 34 ATMOSPHERES. THE COMPLAINANT REPORTS THAT THE PATIENT DIDN'T' HAVE A STENT, THAT THE EVENT DID NOT AFFECT THE PATIENT, AND THAT THE STENT WAS EVENTUALLY SUCCESSFULLY PLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23930997 ↗MalfunctionNov 5, 2025Break
A BALLON WAS INSERTED OVER THE WIRE DURING A CORONARY INTERVENTION. THE PORTION OF THE BALLOON SHAFT THAT WAS OUTSIDE THE PATIENT'S BODY BROKE. THE BALLOON WAS REMOVED AND NOT USED. ALL PORTIONS OF THE BALLOON WERE ACCOUNTED FOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23476430 ↗MalfunctionOct 27, 2025Break; Insufficient Device Problem Information; Break; Insufficient Device Problem Information
THE REPORTER STATES THAT, "THE DEVICE WAS BEING USED IN THE NORMAL FASHION, AND THE SHAFT BROKE DURING THE INSERTION IN THE MIDDLE OF THE SHAFT. THE OTHER HALF WAS RETRIEVED AND THE PARTS TRASHED. THERE WERE NO KINKS OBSERVED PRIOR TO USE. THERE WAS NO IMPACT TO THE PATIENT WHEN THE BROKEN SHAFT. THEY USED ANOTHER MODALITY TO FINISH THE PROCEDURE..."
Patient problems: Insufficient Information
MAUDE report 23394025 ↗MalfunctionMay 21, 2025Material Rupture; Insufficient Device Problem Information; Material Rupture; Insufficient Device Problem Information
THE CLINICIAN REPORTED VIA A DISTRIBUTOR THAT DURING DELIVERY THE SHAFT OF AN OPN NC BALLOON CATHETER "SNAPPED IN HALF" WHILE ADVANCING INTO THE PATIENT THROUGH THE GUIDING CATHETER. THERE WAS NO REPORTED IMPACT TO THE PATIENT AND THE PATIENT WAS DISCHARGED HOME.
Patient problems: Stenosis
MAUDE report 22061494 ↗
Other Catheters, Transluminal Coronary Angioplasty, Percutaneous devices
- MINI TREK™ 1012270-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous223 reports
- Maverick²™ H7493892815200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous1 death203 reports
- MINI TREK™ 1012270-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous162 reports
- Maverick²™ H7493892820200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous161 reports
- NC TREK 1012454-08Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 recalls59 reports
- TREK™ 1012272-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 deaths158 reports
- NC TREK 1012453-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall84 reports
- NC TREK 1012453-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall1 death83 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗