Skip to content

SET Rotarex™S 8 F x 85 cm 80238Recalls & reported problems

Straub Medical AG · Rotational Excisional Atherectomy System. The Catheter Set includes one Rotarex™S catheter (gauge 8F; length 85 cm), one guidewire (0.018” diameter; 220cm in length), one drape and one collecting bag. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07640142811879), received between May 26, 2022 and Mar 10, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 5, 2025Class IOpen, Classified

    Atherectomy Catheter eIFU updated to clarify and emphasize procedural steps intended to reduce the likelihood of catheter breakage events. Catheter has an outer cylinder connected to a rotating helix, which could fracture and/or break, which would require retrieval, and helix facture/break could cause vessel injury and lead to severe bleeding.

    Z-1244-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Deformation2
  • Break1
  • Failure to Advance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 10, 2026Material Deformation

    A PATIENT UNDERWENT A THROMBECTOMY AND ATHERECTOMY PROCEDURE USING A ROTAREX DEVICE. DURING THE PROCEDURE, THE CATHETER WAS PLACED ON THE FIELD, BUT UPON INSPECTION, THE END OF THE DEVICE WAS FOUND TO BE BENT, PREVENTING INSERTION OF THE WIRE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24556100 ↗
  • MalfunctionOct 9, 2023Failure to Advance; Material Deformation; Failure to Advance; Material Deformation

    IT WAS REPORTED THAT DURING THE PREPARATION OF A RECANALIZATION PROCEDURE, THE TIP OF THE CATHETER ALLEGEDLY KINKED. IT WAS FURTHER REPORTED THAT THE WIRE ALLEGEDLY FAILED TO ADVANCE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17893882 ↗
  • MalfunctionMay 26, 2022Break; Break

    IT WAS REPORTED THAT DURING A RECANALIZATION PROCEDURE, THE CATHETERS SPLITED IN TO PIECES WHEN REMOVING THE DEVICE OUTSIDE THE BODY FROM THE SHEATH . THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14515785 ↗

Other Catheter, Peripheral, Atherectomy devices

All Catheter, Peripheral, Atherectomy devices →

Sources and updates